Cargando…
Comparison of higher order wavefront aberrations with four aberrometers
PURPOSE: To evaluate the agreement of selected higher order aberration measurements between aberrometers based on three different wavefront technologies. METHODS: Twenty-three eyes of 23 participants were compared between Zywave, OPD-Scan III, and iDesign aberrometers, for total ocular aberrations....
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6611280/ https://www.ncbi.nlm.nih.gov/pubmed/31238402 http://dx.doi.org/10.4103/ijo.IJO_1464_18 |
_version_ | 1783432663762206720 |
---|---|
author | Cook, William H McKelvie, James Wallace, Henry B Misra, Stuti L |
author_facet | Cook, William H McKelvie, James Wallace, Henry B Misra, Stuti L |
author_sort | Cook, William H |
collection | PubMed |
description | PURPOSE: To evaluate the agreement of selected higher order aberration measurements between aberrometers based on three different wavefront technologies. METHODS: Twenty-three eyes of 23 participants were compared between Zywave, OPD-Scan III, and iDesign aberrometers, for total ocular aberrations. Participants were between 19 and 69 years of age, and exclusion criteria were previous ocular surgery or trauma, contact lens wear within the preceding 2 weeks, and ocular or systemic disease. Corneal aberrations were compared between the OPD-Scan III and GALILEI™ G2 aberrometers. Zernike coefficients of vertical and oblique trefoil, vertical and horizontal coma, and spherical aberration were analyzed in R software. RESULTS: In all, 276 scans were captured in total, with a male-to-female ratio of 11:12. Total ocular vertical coma [mean difference (MD) = 0.026 μm, P < 0.005], vertical trefoil (MD = 0.033 μm, P < 0.05), and spherical aberration (MD = 0.022 μm, P < 0.05) differed significantly between the iDesign and OPD-Scan III. Differences in total vertical (MD = 0.072 μm, P < 0.05) and oblique trefoil (MD = 0.058 μm, P < 0.05) were demonstrated between the Zywave and OPD-Scan III, and spherical aberration (MD = 0.030 μm, P < 0.005) between iDesign and Zywave. iDesign corneal horizontal coma (MD = 0.025 μm, P < 0.05) and spherical aberration (MD = 0.043 μm, P < 0.005) measurements were significantly different between the GALILEI™ G2 and the OPD-Scan III. CONCLUSION: Zywave, iDesign, and OPD-Scan III, and GALILEI™ G2 and OPD-Scan III may be used interchangeably for their total ocular and corneal wavefront functions, respectively; however, care must be taken if using these devices for guiding ablation or monitoring corneal disease. |
format | Online Article Text |
id | pubmed-6611280 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-66112802019-07-22 Comparison of higher order wavefront aberrations with four aberrometers Cook, William H McKelvie, James Wallace, Henry B Misra, Stuti L Indian J Ophthalmol Original Article PURPOSE: To evaluate the agreement of selected higher order aberration measurements between aberrometers based on three different wavefront technologies. METHODS: Twenty-three eyes of 23 participants were compared between Zywave, OPD-Scan III, and iDesign aberrometers, for total ocular aberrations. Participants were between 19 and 69 years of age, and exclusion criteria were previous ocular surgery or trauma, contact lens wear within the preceding 2 weeks, and ocular or systemic disease. Corneal aberrations were compared between the OPD-Scan III and GALILEI™ G2 aberrometers. Zernike coefficients of vertical and oblique trefoil, vertical and horizontal coma, and spherical aberration were analyzed in R software. RESULTS: In all, 276 scans were captured in total, with a male-to-female ratio of 11:12. Total ocular vertical coma [mean difference (MD) = 0.026 μm, P < 0.005], vertical trefoil (MD = 0.033 μm, P < 0.05), and spherical aberration (MD = 0.022 μm, P < 0.05) differed significantly between the iDesign and OPD-Scan III. Differences in total vertical (MD = 0.072 μm, P < 0.05) and oblique trefoil (MD = 0.058 μm, P < 0.05) were demonstrated between the Zywave and OPD-Scan III, and spherical aberration (MD = 0.030 μm, P < 0.005) between iDesign and Zywave. iDesign corneal horizontal coma (MD = 0.025 μm, P < 0.05) and spherical aberration (MD = 0.043 μm, P < 0.005) measurements were significantly different between the GALILEI™ G2 and the OPD-Scan III. CONCLUSION: Zywave, iDesign, and OPD-Scan III, and GALILEI™ G2 and OPD-Scan III may be used interchangeably for their total ocular and corneal wavefront functions, respectively; however, care must be taken if using these devices for guiding ablation or monitoring corneal disease. Wolters Kluwer - Medknow 2019-07 /pmc/articles/PMC6611280/ /pubmed/31238402 http://dx.doi.org/10.4103/ijo.IJO_1464_18 Text en Copyright: © 2019 Indian Journal of Ophthalmology http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Cook, William H McKelvie, James Wallace, Henry B Misra, Stuti L Comparison of higher order wavefront aberrations with four aberrometers |
title | Comparison of higher order wavefront aberrations with four aberrometers |
title_full | Comparison of higher order wavefront aberrations with four aberrometers |
title_fullStr | Comparison of higher order wavefront aberrations with four aberrometers |
title_full_unstemmed | Comparison of higher order wavefront aberrations with four aberrometers |
title_short | Comparison of higher order wavefront aberrations with four aberrometers |
title_sort | comparison of higher order wavefront aberrations with four aberrometers |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6611280/ https://www.ncbi.nlm.nih.gov/pubmed/31238402 http://dx.doi.org/10.4103/ijo.IJO_1464_18 |
work_keys_str_mv | AT cookwilliamh comparisonofhigherorderwavefrontaberrationswithfouraberrometers AT mckelviejames comparisonofhigherorderwavefrontaberrationswithfouraberrometers AT wallacehenryb comparisonofhigherorderwavefrontaberrationswithfouraberrometers AT misrastutil comparisonofhigherorderwavefrontaberrationswithfouraberrometers |