Cargando…
Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac
PURPOSE: To investigate the plan quality and doses to the heart, contralateral breast (CB), ipsilateral lung (IL), and contralateral lung (CL) in tangential breast treatments using the Halcyon linac with megavoltage setup fields. METHODS: Radiotherapy treatment plans with tangential beams from 25 br...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6612683/ https://www.ncbi.nlm.nih.gov/pubmed/31183967 http://dx.doi.org/10.1002/acm2.12655 |
_version_ | 1783432915330269184 |
---|---|
author | Flores‐Martinez, Everardo Kim, Gwe‐Ya Yashar, Catheryn M. Cerviño, Laura I. |
author_facet | Flores‐Martinez, Everardo Kim, Gwe‐Ya Yashar, Catheryn M. Cerviño, Laura I. |
author_sort | Flores‐Martinez, Everardo |
collection | PubMed |
description | PURPOSE: To investigate the plan quality and doses to the heart, contralateral breast (CB), ipsilateral lung (IL), and contralateral lung (CL) in tangential breast treatments using the Halcyon linac with megavoltage setup fields. METHODS: Radiotherapy treatment plans with tangential beams from 25 breast cancer patients previously treated on a C‐arm linac were replanned for Halcyon. Thirteen corresponded to right‐sided breasts and 12 to left‐sided breasts, all with a dose prescription of 50 Gy in 25 fractions. Plans were created with the following setup imaging techniques: low‐dose (LD) MVCBCT, high‐quality (HQ) MVCBCT, LD‐MV and HQ‐MV pairs and the imaging dose was included in the plans. Plan quality metric values for the lumpectomy cavity, whole‐breast and doses to the organs at risk (OARs) were measured and compared with those from the original plans. RESULTS: No significant differences in plan quality were observed between the original and Halcyon plans. An increase in the mean dose (Mean) for all the organs was observed for the Halcyon plans. For right‐sided plans, the accumulated Mean over the 25 fractions in the C‐arm plans was 0.4 ± 0.3, 0.2 ± 0.2, 5.4 ± 1.3, and 0.1 ± 0.1 Gy for the heart, CB, IL, and CL, respectively, while values in the MVCBCT‐LD Halcyon plans were 1.2 ± 0.2, 0.6 ± 0.1, 6.5 ± 1.4, and 0.4 ± 0.1 Gy, respectively. For left‐sided treatments, Mean in the original plans was 0.9 ± 0.2, 0.1 ± 0.0, 4.2 ± 1.2, and 0.0 ± 0.0 Gy, while for the MVCBCT‐LD Halcyon plans values were 1.9 ± 0.2, 0.6 ± 0.2, 5.1 ± 1.2, and 0.5 ± 0.2 Gy, respectively. CONCLUSIONS: Plan quality for breast treatments using Halcyon is similar to the quality for a 6 MV, C‐arm plan. For treatments using megavoltage setup fields, the dose contribution to OARs from the imaging fields can be equal or higher than the dose from treatment fields. |
format | Online Article Text |
id | pubmed-6612683 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-66126832019-07-16 Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac Flores‐Martinez, Everardo Kim, Gwe‐Ya Yashar, Catheryn M. Cerviño, Laura I. J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: To investigate the plan quality and doses to the heart, contralateral breast (CB), ipsilateral lung (IL), and contralateral lung (CL) in tangential breast treatments using the Halcyon linac with megavoltage setup fields. METHODS: Radiotherapy treatment plans with tangential beams from 25 breast cancer patients previously treated on a C‐arm linac were replanned for Halcyon. Thirteen corresponded to right‐sided breasts and 12 to left‐sided breasts, all with a dose prescription of 50 Gy in 25 fractions. Plans were created with the following setup imaging techniques: low‐dose (LD) MVCBCT, high‐quality (HQ) MVCBCT, LD‐MV and HQ‐MV pairs and the imaging dose was included in the plans. Plan quality metric values for the lumpectomy cavity, whole‐breast and doses to the organs at risk (OARs) were measured and compared with those from the original plans. RESULTS: No significant differences in plan quality were observed between the original and Halcyon plans. An increase in the mean dose (Mean) for all the organs was observed for the Halcyon plans. For right‐sided plans, the accumulated Mean over the 25 fractions in the C‐arm plans was 0.4 ± 0.3, 0.2 ± 0.2, 5.4 ± 1.3, and 0.1 ± 0.1 Gy for the heart, CB, IL, and CL, respectively, while values in the MVCBCT‐LD Halcyon plans were 1.2 ± 0.2, 0.6 ± 0.1, 6.5 ± 1.4, and 0.4 ± 0.1 Gy, respectively. For left‐sided treatments, Mean in the original plans was 0.9 ± 0.2, 0.1 ± 0.0, 4.2 ± 1.2, and 0.0 ± 0.0 Gy, while for the MVCBCT‐LD Halcyon plans values were 1.9 ± 0.2, 0.6 ± 0.2, 5.1 ± 1.2, and 0.5 ± 0.2 Gy, respectively. CONCLUSIONS: Plan quality for breast treatments using Halcyon is similar to the quality for a 6 MV, C‐arm plan. For treatments using megavoltage setup fields, the dose contribution to OARs from the imaging fields can be equal or higher than the dose from treatment fields. John Wiley and Sons Inc. 2019-06-11 /pmc/articles/PMC6612683/ /pubmed/31183967 http://dx.doi.org/10.1002/acm2.12655 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Flores‐Martinez, Everardo Kim, Gwe‐Ya Yashar, Catheryn M. Cerviño, Laura I. Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac |
title | Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac |
title_full | Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac |
title_fullStr | Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac |
title_full_unstemmed | Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac |
title_short | Dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the Halcyon linac |
title_sort | dosimetric study of the plan quality and dose to organs at risk on tangential breast treatments using the halcyon linac |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6612683/ https://www.ncbi.nlm.nih.gov/pubmed/31183967 http://dx.doi.org/10.1002/acm2.12655 |
work_keys_str_mv | AT floresmartinezeverardo dosimetricstudyoftheplanqualityanddosetoorgansatriskontangentialbreasttreatmentsusingthehalcyonlinac AT kimgweya dosimetricstudyoftheplanqualityanddosetoorgansatriskontangentialbreasttreatmentsusingthehalcyonlinac AT yasharcatherynm dosimetricstudyoftheplanqualityanddosetoorgansatriskontangentialbreasttreatmentsusingthehalcyonlinac AT cervinolaurai dosimetricstudyoftheplanqualityanddosetoorgansatriskontangentialbreasttreatmentsusingthehalcyonlinac |