Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom

PURPOSE: A contrast‐detail phantom such as CDRAD is frequently used for quality assurance, optimization of image quality, and several other purposes. However, it is often used without considering the uncertainty of the results. The aim of this study was to assess two figure of merits (FOM) originati...

Descripción completa

Detalles Bibliográficos
Autores principales: Konst, Bente, Weedon‐Fekjær, Harald, Båth, Magnus
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6612684/
https://www.ncbi.nlm.nih.gov/pubmed/31152576
http://dx.doi.org/10.1002/acm2.12649
_version_ 1783432915554664448
author Konst, Bente
Weedon‐Fekjær, Harald
Båth, Magnus
author_facet Konst, Bente
Weedon‐Fekjær, Harald
Båth, Magnus
author_sort Konst, Bente
collection PubMed
description PURPOSE: A contrast‐detail phantom such as CDRAD is frequently used for quality assurance, optimization of image quality, and several other purposes. However, it is often used without considering the uncertainty of the results. The aim of this study was to assess two figure of merits (FOM) originating from CDRAD regarding the variations of the FOMs by dose utilized to create the x‐ray image. The probability of overlapping (assessing an image acquired at a lower dose as better than an image acquired at a higher dose) was determined. METHODS: The CDRAD phantom located underneath 12, 20, and 26 cm PMMA was imaged 16 times at five dose levels using an x‐ray system with a flat‐panel detector. All images were analyzed by CDRAD Analyser, version 1.1, which calculated the FOM inverse image quality figure (IQF(inv)) and gave contrast detail curves for each image. Inherent properties of the CDRAD phantom were used to derive a new FOM h, which describes the size of the hole with the same diameter and depth that is just visible. Data were analyzed using heteroscedastic regression of mean and variance by dose. To ease interpretation, probabilities for overlaps were calculated assuming normal distribution, with associated bootstrap confidence intervals. RESULTS: The proportion of total variability in IQF(inv), explained by the dose (R(2)), was 91%, 85%, and 93% for 12, 20, and 26 cm PMMA. Corresponding results for h were 91%, 89%, and 95%. The overlap probability for different mAs levels was 1% for 0.8 vs 1.2 mAs, 5% for 1.2 vs 1.6 mAs, 10% for 1.6 vs 2.0 mAs, and 10% for 2.0 mAs vs 2.5 mAs for 12 cm PMMA. For 20 cm PMMA, it was 0.5% for 10 vs 16 mAs, 13% for 16 vs 20 mAs, 14% for 20 vs 25 mAs, and 14% for 25 vs 32 mAs. For 26 cm PMMA, the probability varied from 0% to 6% for various mAs levels. Even though the estimated probability for overlap was small, the 95% confidence interval (CI) showed relatively large uncertainties. For 12 cm PMMA, the associated CI for 0.8 vs 1.2 mAs was 0.1–3.2%, and the CI for 1.2 vs 1.6 mAs was 2.1–7.8%. CONCLUSIONS: Inverse image quality figure and h are about equally related to dose level. The FOM h, which describes the size of a hole that should be seen in the image, may be a more intuitive FOM than IQF(inv). However, considering the probabilities for overlap and their confidence intervals, the FOMs deduced from the CDRAD phantom are not sensitive to dose. Hence, CDRAD may not be an optimal phantom to differentiate between images acquired at different dose levels.
format Online
Article
Text
id pubmed-6612684
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-66126842019-07-16 Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom Konst, Bente Weedon‐Fekjær, Harald Båth, Magnus J Appl Clin Med Phys Medical Imaging PURPOSE: A contrast‐detail phantom such as CDRAD is frequently used for quality assurance, optimization of image quality, and several other purposes. However, it is often used without considering the uncertainty of the results. The aim of this study was to assess two figure of merits (FOM) originating from CDRAD regarding the variations of the FOMs by dose utilized to create the x‐ray image. The probability of overlapping (assessing an image acquired at a lower dose as better than an image acquired at a higher dose) was determined. METHODS: The CDRAD phantom located underneath 12, 20, and 26 cm PMMA was imaged 16 times at five dose levels using an x‐ray system with a flat‐panel detector. All images were analyzed by CDRAD Analyser, version 1.1, which calculated the FOM inverse image quality figure (IQF(inv)) and gave contrast detail curves for each image. Inherent properties of the CDRAD phantom were used to derive a new FOM h, which describes the size of the hole with the same diameter and depth that is just visible. Data were analyzed using heteroscedastic regression of mean and variance by dose. To ease interpretation, probabilities for overlaps were calculated assuming normal distribution, with associated bootstrap confidence intervals. RESULTS: The proportion of total variability in IQF(inv), explained by the dose (R(2)), was 91%, 85%, and 93% for 12, 20, and 26 cm PMMA. Corresponding results for h were 91%, 89%, and 95%. The overlap probability for different mAs levels was 1% for 0.8 vs 1.2 mAs, 5% for 1.2 vs 1.6 mAs, 10% for 1.6 vs 2.0 mAs, and 10% for 2.0 mAs vs 2.5 mAs for 12 cm PMMA. For 20 cm PMMA, it was 0.5% for 10 vs 16 mAs, 13% for 16 vs 20 mAs, 14% for 20 vs 25 mAs, and 14% for 25 vs 32 mAs. For 26 cm PMMA, the probability varied from 0% to 6% for various mAs levels. Even though the estimated probability for overlap was small, the 95% confidence interval (CI) showed relatively large uncertainties. For 12 cm PMMA, the associated CI for 0.8 vs 1.2 mAs was 0.1–3.2%, and the CI for 1.2 vs 1.6 mAs was 2.1–7.8%. CONCLUSIONS: Inverse image quality figure and h are about equally related to dose level. The FOM h, which describes the size of a hole that should be seen in the image, may be a more intuitive FOM than IQF(inv). However, considering the probabilities for overlap and their confidence intervals, the FOMs deduced from the CDRAD phantom are not sensitive to dose. Hence, CDRAD may not be an optimal phantom to differentiate between images acquired at different dose levels. John Wiley and Sons Inc. 2019-06-01 /pmc/articles/PMC6612684/ /pubmed/31152576 http://dx.doi.org/10.1002/acm2.12649 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Konst, Bente
Weedon‐Fekjær, Harald
Båth, Magnus
Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom
title Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom
title_full Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom
title_fullStr Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom
title_full_unstemmed Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom
title_short Image quality and radiation dose in planar imaging — Image quality figure of merits from the CDRAD phantom
title_sort image quality and radiation dose in planar imaging — image quality figure of merits from the cdrad phantom
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6612684/
https://www.ncbi.nlm.nih.gov/pubmed/31152576
http://dx.doi.org/10.1002/acm2.12649
work_keys_str_mv AT konstbente imagequalityandradiationdoseinplanarimagingimagequalityfigureofmeritsfromthecdradphantom
AT weedonfekjærharald imagequalityandradiationdoseinplanarimagingimagequalityfigureofmeritsfromthecdradphantom
AT bathmagnus imagequalityandradiationdoseinplanarimagingimagequalityfigureofmeritsfromthecdradphantom