Overranging and overbeaming measurement in area detector computed tomography: A method for simultaneous measurement in volume helical acquisition

PURPOSE: We propose a novel method to assess overbeaming and overranging, as well as the effect of reducing longitudinal exposure range, by using a dynamic z‐collimator in area detector computed tomography. METHODS AND MATERIALS: A 500‐mm diameter cylindrical imaging plate was exposed by helical sca...

Descripción completa

Detalles Bibliográficos
Autores principales: Urikura, Atsushi, Hara, Takanori, Yoshida, Tsukasa, Nishimaru, Eiji, Hoshino, Takashi, Nakaya, Yoshihiro, Endo, Masahiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6612690/
https://www.ncbi.nlm.nih.gov/pubmed/31165567
http://dx.doi.org/10.1002/acm2.12650
Descripción
Sumario:PURPOSE: We propose a novel method to assess overbeaming and overranging, as well as the effect of reducing longitudinal exposure range, by using a dynamic z‐collimator in area detector computed tomography. METHODS AND MATERIALS: A 500‐mm diameter cylindrical imaging plate was exposed by helical scanning in a dark room. The beam collimation of the helical acquisitions was set at 32 and 80 mm. Overbeaming and overranging with the dynamic z‐collimator were measured. RESULTS: The actual beam widths were approximately 39 and 88 mm at 32 and 80 mm collimation, respectively, and were relatively reduced owing to increased beam collimation. Overranging was 27.0 and 48.2 mm with a pitch of 0.83 and 1.49 at 32 mm collimation and 72.5 and 83.1 mm with a pitch of 0.87 and 0.99 at 80 mm collimation. The dynamic z‐collimator relatively reduced the overranging by 17.3% and 17.1% for the 32 and 80 mm collimation, respectively. CONCLUSION: We devised a method to simultaneously measure overbeaming and overranging with only one helical acquisition. Although the dynamic z‐collimator reduced the overranging by approximately 17%, wider collimation widths and higher pitch settings would increase the exposure dose outside the scan range.