Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)

This method comparison study used radiographs of 20 mice and 20 budgerigars to investigate comparability between computed radiography (CR) and high-resolution screen-film systems and study the effects of reduced radiation doses on image quality of digital radiographs of small patients. Exposure sett...

Descripción completa

Detalles Bibliográficos
Autores principales: Tebrün, Wiebke, Ludewig, Eberhard, Köhler, Claudia, Böhme, Julia, Pees, Michael
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6624299/
https://www.ncbi.nlm.nih.gov/pubmed/31296929
http://dx.doi.org/10.1038/s41598-019-46546-5
_version_ 1783434242217213952
author Tebrün, Wiebke
Ludewig, Eberhard
Köhler, Claudia
Böhme, Julia
Pees, Michael
author_facet Tebrün, Wiebke
Ludewig, Eberhard
Köhler, Claudia
Böhme, Julia
Pees, Michael
author_sort Tebrün, Wiebke
collection PubMed
description This method comparison study used radiographs of 20 mice and 20 budgerigars to investigate comparability between computed radiography (CR) and high-resolution screen-film systems and study the effects of reduced radiation doses on image quality of digital radiographs of small patients. Exposure settings used with the mammography screen-film system (SF) were taken as baseline settings. A powder-based storage-phosphor system (CR(P)) and a needle-based storage-phosphor system (CR(N)) were used with the same settings (D/100%) and half the detector dose (D/50%). Using a scoring system four reviewers assessed five criteria per species covering soft tissue and bone structures. Results were evaluated for differences between reviewers (interobserver variability), systems and settings (intersystem variability, using visual grading characteristic analysis). Correlations were significant (p ≤ 0.05) for interobserver variability in 86.7% of the cases. Correlation coefficients ranged from 0.206 to 0.772. For mice and budgerigars, the CR(N) system was rated as superior to the SF and CR(P) system for most criteria, being significant in two cases each. Comparing the SF and CR(P) system, the conventional method scored higher for all criteria, in one case significantly. For both species and both digital systems, dose reduction to 50% resulted in significantly worse scores for most criteria. In summary, the needle-based storage-phosphor technique proved to be superior compared to the conventional storage-phosphor and mammography screen-film system. Needle-based detector systems are suitable substitutes for high-resolution screen–film systems when performing diagnostic imaging of small patients. Dose reduction to 50% of the corresponding dose needed in high-resolution film-screen systems cannot be recommended.
format Online
Article
Text
id pubmed-6624299
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-66242992019-07-19 Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice) Tebrün, Wiebke Ludewig, Eberhard Köhler, Claudia Böhme, Julia Pees, Michael Sci Rep Article This method comparison study used radiographs of 20 mice and 20 budgerigars to investigate comparability between computed radiography (CR) and high-resolution screen-film systems and study the effects of reduced radiation doses on image quality of digital radiographs of small patients. Exposure settings used with the mammography screen-film system (SF) were taken as baseline settings. A powder-based storage-phosphor system (CR(P)) and a needle-based storage-phosphor system (CR(N)) were used with the same settings (D/100%) and half the detector dose (D/50%). Using a scoring system four reviewers assessed five criteria per species covering soft tissue and bone structures. Results were evaluated for differences between reviewers (interobserver variability), systems and settings (intersystem variability, using visual grading characteristic analysis). Correlations were significant (p ≤ 0.05) for interobserver variability in 86.7% of the cases. Correlation coefficients ranged from 0.206 to 0.772. For mice and budgerigars, the CR(N) system was rated as superior to the SF and CR(P) system for most criteria, being significant in two cases each. Comparing the SF and CR(P) system, the conventional method scored higher for all criteria, in one case significantly. For both species and both digital systems, dose reduction to 50% resulted in significantly worse scores for most criteria. In summary, the needle-based storage-phosphor technique proved to be superior compared to the conventional storage-phosphor and mammography screen-film system. Needle-based detector systems are suitable substitutes for high-resolution screen–film systems when performing diagnostic imaging of small patients. Dose reduction to 50% of the corresponding dose needed in high-resolution film-screen systems cannot be recommended. Nature Publishing Group UK 2019-07-11 /pmc/articles/PMC6624299/ /pubmed/31296929 http://dx.doi.org/10.1038/s41598-019-46546-5 Text en © The Author(s) 2019 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Article
Tebrün, Wiebke
Ludewig, Eberhard
Köhler, Claudia
Böhme, Julia
Pees, Michael
Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
title Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
title_full Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
title_fullStr Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
title_full_unstemmed Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
title_short Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
title_sort needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6624299/
https://www.ncbi.nlm.nih.gov/pubmed/31296929
http://dx.doi.org/10.1038/s41598-019-46546-5
work_keys_str_mv AT tebrunwiebke needlebasedstoragephosphordetectorradiographyissuperiortoaconventionalpowderbasedstoragephosphordetectorandahighresolutionscreenfilmsysteminsmallpatientsbudgerigarsandmice
AT ludewigeberhard needlebasedstoragephosphordetectorradiographyissuperiortoaconventionalpowderbasedstoragephosphordetectorandahighresolutionscreenfilmsysteminsmallpatientsbudgerigarsandmice
AT kohlerclaudia needlebasedstoragephosphordetectorradiographyissuperiortoaconventionalpowderbasedstoragephosphordetectorandahighresolutionscreenfilmsysteminsmallpatientsbudgerigarsandmice
AT bohmejulia needlebasedstoragephosphordetectorradiographyissuperiortoaconventionalpowderbasedstoragephosphordetectorandahighresolutionscreenfilmsysteminsmallpatientsbudgerigarsandmice
AT peesmichael needlebasedstoragephosphordetectorradiographyissuperiortoaconventionalpowderbasedstoragephosphordetectorandahighresolutionscreenfilmsysteminsmallpatientsbudgerigarsandmice