Positional dependence of activity determination in single photon emission computed tomography
Accurate image quantification requires accurate calibration of the detector and is vital if dosimetry is to be performed in molecular radiotherapy. A dependence on the position of calibration has been observed in single photon emission computed tomography images when attenuation correction (AC) and...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6635063/ https://www.ncbi.nlm.nih.gov/pubmed/31136536 http://dx.doi.org/10.1097/MNM.0000000000001034 |
Sumario: | Accurate image quantification requires accurate calibration of the detector and is vital if dosimetry is to be performed in molecular radiotherapy. A dependence on the position of calibration has been observed in single photon emission computed tomography images when attenuation correction (AC) and scatter correction are applied. This work investigates the origin of this dependence in single photon emission computed tomography scans of phantom inserts filled with (177)Lu solution. A 113 ml sphere and inserts representing a mathematical model of a spleen and an anatomical model of a patient spleen were imaged at the centre and edge of elliptical phantoms. For these inserts, the difference in calibration factor between the positions was around 10% for images reconstructed with AC and triple energy window scatter correction. A combination of experimental imaging and Monte Carlo simulation was used to isolate possible causes due to imaging or reconstruction in turn. Inconsistent application of AC between different reconstruction systems was identified as the origin of the positional dependence. |
---|