Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes

OBJECTIVE: To compare the difference between an automated oscillometric ABI measurement as compared to standard hand-held doppler ABI in patients with Type-2 diabetes. METHODS: This prospective study was conducted at foot clinic of Baqai Institute of Diabetology and Endocrinology (BIDE), Baqai Medic...

Descripción completa

Detalles Bibliográficos
Autores principales: Khan, Shair Zaman, Awn-Bin-Zafar, Waris, Nazish, Miyan, Zahid, Ulhaque, Muhammad Saif, Fawwad, Asher
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Professional Medical Publications 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6659060/
https://www.ncbi.nlm.nih.gov/pubmed/31372162
http://dx.doi.org/10.12669/pjms.35.4.30
_version_ 1783439061781839872
author Khan, Shair Zaman
Awn-Bin-Zafar,
Waris, Nazish
Miyan, Zahid
Ulhaque, Muhammad Saif
Fawwad, Asher
author_facet Khan, Shair Zaman
Awn-Bin-Zafar,
Waris, Nazish
Miyan, Zahid
Ulhaque, Muhammad Saif
Fawwad, Asher
author_sort Khan, Shair Zaman
collection PubMed
description OBJECTIVE: To compare the difference between an automated oscillometric ABI measurement as compared to standard hand-held doppler ABI in patients with Type-2 diabetes. METHODS: This prospective study was conducted at foot clinic of Baqai Institute of Diabetology and Endocrinology (BIDE), Baqai Medical University (BMU), a tertiary care unit, Karachi-Pakistan. The duration of study was February 2018 to March 2018. Patients with Type-2 diabetes attending the outpatient department (OPD) of foot clinic, irrespective of their symptoms were included. Baseline demographic, anthropometric measurements and biochemical parameters were recorded. The ABI was calculated with both devices by an automated oscillometric machine and standard hand-held doppler with the same investigator. RESULTS: Total of 93 patients with Type-2 diabetes, 18 (19.4%) females and 75(80.6%) males were recruited. Mean age was 54.67±9.59 years and mean systolic/diastolic blood pressure was 131.38±20.2/ 80.36±10.23mmHg. Most of the patients had poor glycemic control at presentation with a mean HbA1c of 9.56±2.44%. Mean standard handheld doppler ABI and automated oscillometric ABI was 1.28±1.08 and 1.07±0.23 for right foot (mean difference = 0.21; P= 0.075), and 1.14±0.45 and 1.1±0.25 for left foot (mean difference =0.04; P=0.434), respectively. Similarly, sensitivity and specificity between two modalities was observed 60% and 93.90% for right foot, meanwhile, 60% and 97.40% for left foot, respectively. CONCLUSION: An automated oscillometric method is comparable with standard handheld-doppler method. It is cost effective, convenient and less time consuming, can be widely used to measure ABI without special training.
format Online
Article
Text
id pubmed-6659060
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Professional Medical Publications
record_format MEDLINE/PubMed
spelling pubmed-66590602019-08-01 Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes Khan, Shair Zaman Awn-Bin-Zafar, Waris, Nazish Miyan, Zahid Ulhaque, Muhammad Saif Fawwad, Asher Pak J Med Sci Original Article OBJECTIVE: To compare the difference between an automated oscillometric ABI measurement as compared to standard hand-held doppler ABI in patients with Type-2 diabetes. METHODS: This prospective study was conducted at foot clinic of Baqai Institute of Diabetology and Endocrinology (BIDE), Baqai Medical University (BMU), a tertiary care unit, Karachi-Pakistan. The duration of study was February 2018 to March 2018. Patients with Type-2 diabetes attending the outpatient department (OPD) of foot clinic, irrespective of their symptoms were included. Baseline demographic, anthropometric measurements and biochemical parameters were recorded. The ABI was calculated with both devices by an automated oscillometric machine and standard hand-held doppler with the same investigator. RESULTS: Total of 93 patients with Type-2 diabetes, 18 (19.4%) females and 75(80.6%) males were recruited. Mean age was 54.67±9.59 years and mean systolic/diastolic blood pressure was 131.38±20.2/ 80.36±10.23mmHg. Most of the patients had poor glycemic control at presentation with a mean HbA1c of 9.56±2.44%. Mean standard handheld doppler ABI and automated oscillometric ABI was 1.28±1.08 and 1.07±0.23 for right foot (mean difference = 0.21; P= 0.075), and 1.14±0.45 and 1.1±0.25 for left foot (mean difference =0.04; P=0.434), respectively. Similarly, sensitivity and specificity between two modalities was observed 60% and 93.90% for right foot, meanwhile, 60% and 97.40% for left foot, respectively. CONCLUSION: An automated oscillometric method is comparable with standard handheld-doppler method. It is cost effective, convenient and less time consuming, can be widely used to measure ABI without special training. Professional Medical Publications 2019 /pmc/articles/PMC6659060/ /pubmed/31372162 http://dx.doi.org/10.12669/pjms.35.4.30 Text en Copyright: © Pakistan Journal of Medical Sciences http://creativecommons.org/licenses/by/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Khan, Shair Zaman
Awn-Bin-Zafar,
Waris, Nazish
Miyan, Zahid
Ulhaque, Muhammad Saif
Fawwad, Asher
Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes
title Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes
title_full Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes
title_fullStr Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes
title_full_unstemmed Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes
title_short Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes
title_sort comparison of ankle-brachial index (abi) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with type-2 diabetes
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6659060/
https://www.ncbi.nlm.nih.gov/pubmed/31372162
http://dx.doi.org/10.12669/pjms.35.4.30
work_keys_str_mv AT khanshairzaman comparisonofanklebrachialindexabimeasuredbyanautomatedoscillometricapparatuswiththatbystandardhandhelddopplerinpatientswithtype2diabetes
AT awnbinzafar comparisonofanklebrachialindexabimeasuredbyanautomatedoscillometricapparatuswiththatbystandardhandhelddopplerinpatientswithtype2diabetes
AT warisnazish comparisonofanklebrachialindexabimeasuredbyanautomatedoscillometricapparatuswiththatbystandardhandhelddopplerinpatientswithtype2diabetes
AT miyanzahid comparisonofanklebrachialindexabimeasuredbyanautomatedoscillometricapparatuswiththatbystandardhandhelddopplerinpatientswithtype2diabetes
AT ulhaquemuhammadsaif comparisonofanklebrachialindexabimeasuredbyanautomatedoscillometricapparatuswiththatbystandardhandhelddopplerinpatientswithtype2diabetes
AT fawwadasher comparisonofanklebrachialindexabimeasuredbyanautomatedoscillometricapparatuswiththatbystandardhandhelddopplerinpatientswithtype2diabetes