Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus
PURPOSE: To establish which reference body offers the greatest sensitivity in keratoconus (KC) diagnosis, obtain normative data for the myopic population with toric ellipsoid reference bodies, and determine the cutoff points for a population with KC. METHODS: A retrospective, observational study of...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6668563/ https://www.ncbi.nlm.nih.gov/pubmed/31396409 http://dx.doi.org/10.1155/2019/2593404 |
_version_ | 1783440242355732480 |
---|---|
author | Garcerant, Daniel Jiménez-Alfaro, Ignacio Alejandre, Nicolás |
author_facet | Garcerant, Daniel Jiménez-Alfaro, Ignacio Alejandre, Nicolás |
author_sort | Garcerant, Daniel |
collection | PubMed |
description | PURPOSE: To establish which reference body offers the greatest sensitivity in keratoconus (KC) diagnosis, obtain normative data for the myopic population with toric ellipsoid reference bodies, and determine the cutoff points for a population with KC. METHODS: A retrospective, observational study of the entire Scheimpflug tomographer database of the Fundación Jiménez Díaz in Madrid was conducted to identify a normal myopic and a KC myopic population. Three different reference bodies were tested on all patients: best fit sphere (BFS), best fit toric ellipsoid with fixed eccentricity (BFTEFE), and best fit toric ellipsoid (BFTE). Anterior and posterior elevation measurements at the apex and thinnest point were recorded, as well as the root mean square of posterior elevations (RMS-P). Normative data were extracted, and receiver operating characteristic (ROC) curves were generated to obtain cutoff points between the normal and KC population. RESULTS: A total of 301 eyes were included, comprising 219 normal myopic and 82 myopic KC eyes. BFS and BFTEFE produced the best results when measuring posterior elevation at the thinnest point. BFTE had better sensitivity with the RMS-P. From all measurements, best sensitivity (100%) was achieved with a cutoff point of 8 μm of posterior elevation at the thinnest point using the BFTEFE. BFTE was found to hide the cone in certain patients. CONCLUSIONS: Posterior elevation measured at the thinnest point with a BFTEFE is the best-performing parameter and, therefore, is recommended to discriminate between normal and KC patients within a myopic population. |
format | Online Article Text |
id | pubmed-6668563 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-66685632019-08-08 Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus Garcerant, Daniel Jiménez-Alfaro, Ignacio Alejandre, Nicolás J Ophthalmol Research Article PURPOSE: To establish which reference body offers the greatest sensitivity in keratoconus (KC) diagnosis, obtain normative data for the myopic population with toric ellipsoid reference bodies, and determine the cutoff points for a population with KC. METHODS: A retrospective, observational study of the entire Scheimpflug tomographer database of the Fundación Jiménez Díaz in Madrid was conducted to identify a normal myopic and a KC myopic population. Three different reference bodies were tested on all patients: best fit sphere (BFS), best fit toric ellipsoid with fixed eccentricity (BFTEFE), and best fit toric ellipsoid (BFTE). Anterior and posterior elevation measurements at the apex and thinnest point were recorded, as well as the root mean square of posterior elevations (RMS-P). Normative data were extracted, and receiver operating characteristic (ROC) curves were generated to obtain cutoff points between the normal and KC population. RESULTS: A total of 301 eyes were included, comprising 219 normal myopic and 82 myopic KC eyes. BFS and BFTEFE produced the best results when measuring posterior elevation at the thinnest point. BFTE had better sensitivity with the RMS-P. From all measurements, best sensitivity (100%) was achieved with a cutoff point of 8 μm of posterior elevation at the thinnest point using the BFTEFE. BFTE was found to hide the cone in certain patients. CONCLUSIONS: Posterior elevation measured at the thinnest point with a BFTEFE is the best-performing parameter and, therefore, is recommended to discriminate between normal and KC patients within a myopic population. Hindawi 2019-07-18 /pmc/articles/PMC6668563/ /pubmed/31396409 http://dx.doi.org/10.1155/2019/2593404 Text en Copyright © 2019 Daniel Garcerant et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Garcerant, Daniel Jiménez-Alfaro, Ignacio Alejandre, Nicolás Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus |
title | Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus |
title_full | Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus |
title_fullStr | Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus |
title_full_unstemmed | Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus |
title_short | Diagnostic Sensitivity of Different Reference Bodies When Using Scheimpflug Tomography in a Myopic Population with Keratoconus |
title_sort | diagnostic sensitivity of different reference bodies when using scheimpflug tomography in a myopic population with keratoconus |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6668563/ https://www.ncbi.nlm.nih.gov/pubmed/31396409 http://dx.doi.org/10.1155/2019/2593404 |
work_keys_str_mv | AT garcerantdaniel diagnosticsensitivityofdifferentreferencebodieswhenusingscheimpflugtomographyinamyopicpopulationwithkeratoconus AT jimenezalfaroignacio diagnosticsensitivityofdifferentreferencebodieswhenusingscheimpflugtomographyinamyopicpopulationwithkeratoconus AT alejandrenicolas diagnosticsensitivityofdifferentreferencebodieswhenusingscheimpflugtomographyinamyopicpopulationwithkeratoconus |