Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients

AIMS: Accurate blood pressure (BP) measurement in continuous‐flow ventricular assist device (CF‐VAD) patients is imperative to reduce stroke risk. This study assesses the accuracy of the Doppler opening pressure method compared with the gold standard arterial line method in CF‐VAD patients. METHODS...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Song, Beckman, Jennifer A., Welch, Nathan G., Bjelkengren, Jason, Masri, Sofia Carolina, Minami, Elina, Stempien‐Otero, April, Levy, Wayne C., O'Brien, Kevin D., Lin, Shin, Farris, Stephen D., Cheng, Richard K., Wood, Gregory, Koomalsingh, Kevin, Kirkpatrick, James, McCabe, James, Leary, Peter J., Chassagne, Fanette, Chivukula, Venkat Keshav, Aliseda, Alberto, Mahr, Claudius
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6676287/
https://www.ncbi.nlm.nih.gov/pubmed/31099483
http://dx.doi.org/10.1002/ehf2.12456
_version_ 1783440742460424192
author Li, Song
Beckman, Jennifer A.
Welch, Nathan G.
Bjelkengren, Jason
Masri, Sofia Carolina
Minami, Elina
Stempien‐Otero, April
Levy, Wayne C.
O'Brien, Kevin D.
Lin, Shin
Farris, Stephen D.
Cheng, Richard K.
Wood, Gregory
Koomalsingh, Kevin
Kirkpatrick, James
McCabe, James
Leary, Peter J.
Chassagne, Fanette
Chivukula, Venkat Keshav
Aliseda, Alberto
Mahr, Claudius
author_facet Li, Song
Beckman, Jennifer A.
Welch, Nathan G.
Bjelkengren, Jason
Masri, Sofia Carolina
Minami, Elina
Stempien‐Otero, April
Levy, Wayne C.
O'Brien, Kevin D.
Lin, Shin
Farris, Stephen D.
Cheng, Richard K.
Wood, Gregory
Koomalsingh, Kevin
Kirkpatrick, James
McCabe, James
Leary, Peter J.
Chassagne, Fanette
Chivukula, Venkat Keshav
Aliseda, Alberto
Mahr, Claudius
author_sort Li, Song
collection PubMed
description AIMS: Accurate blood pressure (BP) measurement in continuous‐flow ventricular assist device (CF‐VAD) patients is imperative to reduce stroke risk. This study assesses the accuracy of the Doppler opening pressure method compared with the gold standard arterial line method in CF‐VAD patients. METHODS AND RESULTS: In a longitudinal cohort of HeartMate II and HVAD patients, arterial line BP and simultaneously measured Doppler opening pressure were obtained. Overall correlation, agreement between Doppler opening pressure and arterial line mean vs. systolic pressure, and the effect of arterial pulsatility on the accuracy of Doppler opening pressure were analysed. A total of 1933 pairs of Doppler opening pressure and arterial line pressure readings within 1 min of each other were identified in 154 patients (20% women, mean age 55 ± 15, 50% HeartMate II and 50% HVAD). Doppler opening pressure had good correlation with invasive mean arterial pressure (r = 0.742, P < 0.0001) and more closely approximated mean than systolic BP (mean error 2.4 vs. −8.4 mmHg). Arterial pulsatility did not have a clinically significant effect on the accuracy of the Doppler opening pressure method. CONCLUSIONS: Doppler opening pressure should be the standard non‐invasive method of BP measurement in CF‐VAD patients.
format Online
Article
Text
id pubmed-6676287
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-66762872019-08-06 Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients Li, Song Beckman, Jennifer A. Welch, Nathan G. Bjelkengren, Jason Masri, Sofia Carolina Minami, Elina Stempien‐Otero, April Levy, Wayne C. O'Brien, Kevin D. Lin, Shin Farris, Stephen D. Cheng, Richard K. Wood, Gregory Koomalsingh, Kevin Kirkpatrick, James McCabe, James Leary, Peter J. Chassagne, Fanette Chivukula, Venkat Keshav Aliseda, Alberto Mahr, Claudius ESC Heart Fail Original Research Articles AIMS: Accurate blood pressure (BP) measurement in continuous‐flow ventricular assist device (CF‐VAD) patients is imperative to reduce stroke risk. This study assesses the accuracy of the Doppler opening pressure method compared with the gold standard arterial line method in CF‐VAD patients. METHODS AND RESULTS: In a longitudinal cohort of HeartMate II and HVAD patients, arterial line BP and simultaneously measured Doppler opening pressure were obtained. Overall correlation, agreement between Doppler opening pressure and arterial line mean vs. systolic pressure, and the effect of arterial pulsatility on the accuracy of Doppler opening pressure were analysed. A total of 1933 pairs of Doppler opening pressure and arterial line pressure readings within 1 min of each other were identified in 154 patients (20% women, mean age 55 ± 15, 50% HeartMate II and 50% HVAD). Doppler opening pressure had good correlation with invasive mean arterial pressure (r = 0.742, P < 0.0001) and more closely approximated mean than systolic BP (mean error 2.4 vs. −8.4 mmHg). Arterial pulsatility did not have a clinically significant effect on the accuracy of the Doppler opening pressure method. CONCLUSIONS: Doppler opening pressure should be the standard non‐invasive method of BP measurement in CF‐VAD patients. John Wiley and Sons Inc. 2019-05-17 /pmc/articles/PMC6676287/ /pubmed/31099483 http://dx.doi.org/10.1002/ehf2.12456 Text en © 2019 The Authors. ESC Heart Failure published by John Wiley & Sons Ltd on behalf of the European Society of Cardiology. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle Original Research Articles
Li, Song
Beckman, Jennifer A.
Welch, Nathan G.
Bjelkengren, Jason
Masri, Sofia Carolina
Minami, Elina
Stempien‐Otero, April
Levy, Wayne C.
O'Brien, Kevin D.
Lin, Shin
Farris, Stephen D.
Cheng, Richard K.
Wood, Gregory
Koomalsingh, Kevin
Kirkpatrick, James
McCabe, James
Leary, Peter J.
Chassagne, Fanette
Chivukula, Venkat Keshav
Aliseda, Alberto
Mahr, Claudius
Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
title Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
title_full Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
title_fullStr Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
title_full_unstemmed Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
title_short Accuracy of Doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
title_sort accuracy of doppler blood pressure measurement in continuous‐flow left ventricular assist device patients
topic Original Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6676287/
https://www.ncbi.nlm.nih.gov/pubmed/31099483
http://dx.doi.org/10.1002/ehf2.12456
work_keys_str_mv AT lisong accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT beckmanjennifera accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT welchnathang accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT bjelkengrenjason accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT masrisofiacarolina accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT minamielina accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT stempienoteroapril accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT levywaynec accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT obrienkevind accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT linshin accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT farrisstephend accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT chengrichardk accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT woodgregory accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT koomalsinghkevin accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT kirkpatrickjames accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT mccabejames accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT learypeterj accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT chassagnefanette accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT chivukulavenkatkeshav accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT alisedaalberto accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients
AT mahrclaudius accuracyofdopplerbloodpressuremeasurementincontinuousflowleftventricularassistdevicepatients