Cargando…

Invasive Validation of Antares, a New Algorithm to Calculate Central Blood Pressure from Oscillometric Upper Arm Pulse Waves

Background: Antares is an algorithm for pulse wave analysis (PWA) by oscillometric blood pressure (BP) monitors in order to estimate central (aortic) blood pressure (cBP). Antares aims to enable brachial cuff-based BP monitors to be type II-devices, determining absolute cBP values independently of p...

Descripción completa

Detalles Bibliográficos
Autores principales: Dörr, Marcus, Richter, Stefan, Eckert, Siegfried, Ohlow, Marc-Alexander, Hammer, Fabian, Hummel, Astrid, Dornberger, Vivien, Genzel, Elisabeth, Baulmann, Johannes
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6678985/
https://www.ncbi.nlm.nih.gov/pubmed/31336569
http://dx.doi.org/10.3390/jcm8071073
Descripción
Sumario:Background: Antares is an algorithm for pulse wave analysis (PWA) by oscillometric blood pressure (BP) monitors in order to estimate central (aortic) blood pressure (cBP). Antares aims to enable brachial cuff-based BP monitors to be type II-devices, determining absolute cBP values independently of potential peripheral BP inaccuracies. The present study is an invasive validation of the Antares algorithm in the custo screen 400. Methods: We followed entirely the 2017 ARTERY protocol for validation of non-invasive cBP devices, the 2013 American National Standards Institute, Inc./Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2 and 2018 AAMI/European Society of Hypertension (ESH)/ISO validation standard protocols. In total, 191 patients undergoing cardiac catheterization were included, of which 145 patients entered analysis. Invasive cBP recordings were compared to simultaneous non-invasive cBP estimations using the Antares algorithm, integrated into an oscillometric BP monitor. Results: Mean difference between invasive and non-invasively estimated systolic cBP was 0.71 mmHg with standard deviation of 5.95 mmHg, fulfilling highest validation criteria. Conclusion: Antares is the first algorithm for estimation of cBP that entirely fulfills the 2017 ARTERY and AAMI/ESH/ISO validation protocols. The Antares algorithm turns the custo screen 400 BP monitor into a type II-device. Integration of Antares into commercially available BP monitors could make it possible to measure PWA parameters in virtually every practice in future.