Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air
A 57‐year‐old man was admitted to our hospital due to syncopal attack. He was diagnosed with Brugada syndrome due to which a subcutaneous implantable cardiac defibrillator (S‐ICD) was inserted using the standard technique. Two hours after the implantation, he experienced inappropriate shock while co...
Autores principales: | , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6686347/ https://www.ncbi.nlm.nih.gov/pubmed/31410242 http://dx.doi.org/10.1002/joa3.12210 |
_version_ | 1783442544330276864 |
---|---|
author | Nishinarita, Ryo Kishihara, Jun Matsuura, Gen Arakawa, Yuki Kobayashi, Shuhei Shirakawa, Yuki Horiguchi, Ai Nakamura, Hironori Ishizue, Naruya Oikawa, Jun Fukaya, Hidehira Niwano, Shinichi Ako, Junya |
author_facet | Nishinarita, Ryo Kishihara, Jun Matsuura, Gen Arakawa, Yuki Kobayashi, Shuhei Shirakawa, Yuki Horiguchi, Ai Nakamura, Hironori Ishizue, Naruya Oikawa, Jun Fukaya, Hidehira Niwano, Shinichi Ako, Junya |
author_sort | Nishinarita, Ryo |
collection | PubMed |
description | A 57‐year‐old man was admitted to our hospital due to syncopal attack. He was diagnosed with Brugada syndrome due to which a subcutaneous implantable cardiac defibrillator (S‐ICD) was inserted using the standard technique. Two hours after the implantation, he experienced inappropriate shock while conscious. Device interrogation revealed a contentious baseline shift and frequent oversensing of low‐amplitude signals, which was followed by a shock. Lateral chest X‐ray revealed subcutaneous air surrounding the proximal electrode. Another inappropriate shock could be avoided by changing the sensing vector. The subcutaneous air was completely resolved 7 days after implantation. |
format | Online Article Text |
id | pubmed-6686347 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-66863472019-08-13 Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air Nishinarita, Ryo Kishihara, Jun Matsuura, Gen Arakawa, Yuki Kobayashi, Shuhei Shirakawa, Yuki Horiguchi, Ai Nakamura, Hironori Ishizue, Naruya Oikawa, Jun Fukaya, Hidehira Niwano, Shinichi Ako, Junya J Arrhythm Case Reports A 57‐year‐old man was admitted to our hospital due to syncopal attack. He was diagnosed with Brugada syndrome due to which a subcutaneous implantable cardiac defibrillator (S‐ICD) was inserted using the standard technique. Two hours after the implantation, he experienced inappropriate shock while conscious. Device interrogation revealed a contentious baseline shift and frequent oversensing of low‐amplitude signals, which was followed by a shock. Lateral chest X‐ray revealed subcutaneous air surrounding the proximal electrode. Another inappropriate shock could be avoided by changing the sensing vector. The subcutaneous air was completely resolved 7 days after implantation. John Wiley and Sons Inc. 2019-06-17 /pmc/articles/PMC6686347/ /pubmed/31410242 http://dx.doi.org/10.1002/joa3.12210 Text en © 2019 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes. |
spellingShingle | Case Reports Nishinarita, Ryo Kishihara, Jun Matsuura, Gen Arakawa, Yuki Kobayashi, Shuhei Shirakawa, Yuki Horiguchi, Ai Nakamura, Hironori Ishizue, Naruya Oikawa, Jun Fukaya, Hidehira Niwano, Shinichi Ako, Junya Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
title | Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
title_full | Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
title_fullStr | Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
title_full_unstemmed | Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
title_short | Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
title_sort | early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air |
topic | Case Reports |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6686347/ https://www.ncbi.nlm.nih.gov/pubmed/31410242 http://dx.doi.org/10.1002/joa3.12210 |
work_keys_str_mv | AT nishinaritaryo earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT kishiharajun earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT matsuuragen earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT arakawayuki earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT kobayashishuhei earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT shirakawayuki earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT horiguchiai earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT nakamurahironori earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT ishizuenaruya earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT oikawajun earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT fukayahidehira earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT niwanoshinichi earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair AT akojunya earlyinappropriateshockinasubcutaneouscardiacdefibrillatorduetosubcutaneousair |