Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
STUDY DESIGN: Prospective pre-clinical and clinical cohort study. OBJECTIVES: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation e...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6686387/ https://www.ncbi.nlm.nih.gov/pubmed/31431874 http://dx.doi.org/10.1177/2192568218804556 |
_version_ | 1783442553729712128 |
---|---|
author | Guha, Daipayan Jakubovic, Raphael Gupta, Shaurya Fehlings, Michael G. Mainprize, Todd G. Yee, Albert Yang, Victor X. D. |
author_facet | Guha, Daipayan Jakubovic, Raphael Gupta, Shaurya Fehlings, Michael G. Mainprize, Todd G. Yee, Albert Yang, Victor X. D. |
author_sort | Guha, Daipayan |
collection | PubMed |
description | STUDY DESIGN: Prospective pre-clinical and clinical cohort study. OBJECTIVES: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation error as a function of distance from the registered vertebral level, and from intersegmental mobility due to surgical manipulation and patient respiration. METHODS: Navigation errors from working distant to the registered level, and from surgical manipulation, were quantified in 4 human cadavers. The 3-dimensional (3D) position of a tracked tool tip at 0 to 5 levels from the DRF, and during targeting of pedicle screw tracts, was captured in real-time by an optical navigation system. Respiration-induced vertebral motion was quantified from 10 clinical cases of open posterior instrumentation. The 3D position of a custom spinous-process clamp was tracked over 12 respiratory cycles. RESULTS: An increase in mean 3D navigation error of ≥2 mm was observed at ≥2 levels from the DRF in the cervical and lumbar spine. Mean ± SD displacement due to surgical manipulation was 1.55 ± 1.13 mm in 3D across all levels, ≥2 mm in 17.4%, 19.2%, and 38.5% of levels in the cervical, thoracic, and lumbar spine, respectively. Mean ± SD respiration-induced 3D motion was 1.96 ± 1.32 mm, greatest in the lower thoracic spine (P < .001). Tidal volume and positive end-expiratory pressure correlated positively with increased vertebral displacement. CONCLUSIONS: Vertebral motion is unaccounted for during image-guided surgery when performed at levels distant from the DRF. Navigating instrumentation within 2 levels of the DRF likely minimizes the risk of navigation error. |
format | Online Article Text |
id | pubmed-6686387 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-66863872019-08-20 Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study Guha, Daipayan Jakubovic, Raphael Gupta, Shaurya Fehlings, Michael G. Mainprize, Todd G. Yee, Albert Yang, Victor X. D. Global Spine J Original Articles STUDY DESIGN: Prospective pre-clinical and clinical cohort study. OBJECTIVES: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation error as a function of distance from the registered vertebral level, and from intersegmental mobility due to surgical manipulation and patient respiration. METHODS: Navigation errors from working distant to the registered level, and from surgical manipulation, were quantified in 4 human cadavers. The 3-dimensional (3D) position of a tracked tool tip at 0 to 5 levels from the DRF, and during targeting of pedicle screw tracts, was captured in real-time by an optical navigation system. Respiration-induced vertebral motion was quantified from 10 clinical cases of open posterior instrumentation. The 3D position of a custom spinous-process clamp was tracked over 12 respiratory cycles. RESULTS: An increase in mean 3D navigation error of ≥2 mm was observed at ≥2 levels from the DRF in the cervical and lumbar spine. Mean ± SD displacement due to surgical manipulation was 1.55 ± 1.13 mm in 3D across all levels, ≥2 mm in 17.4%, 19.2%, and 38.5% of levels in the cervical, thoracic, and lumbar spine, respectively. Mean ± SD respiration-induced 3D motion was 1.96 ± 1.32 mm, greatest in the lower thoracic spine (P < .001). Tidal volume and positive end-expiratory pressure correlated positively with increased vertebral displacement. CONCLUSIONS: Vertebral motion is unaccounted for during image-guided surgery when performed at levels distant from the DRF. Navigating instrumentation within 2 levels of the DRF likely minimizes the risk of navigation error. SAGE Publications 2018-10-09 2019-08 /pmc/articles/PMC6686387/ /pubmed/31431874 http://dx.doi.org/10.1177/2192568218804556 Text en © The Author(s) 2018 http://creativecommons.org/licenses/by-nc-nd/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 License (http://www.creativecommons.org/licenses/by-nc-nd/4.0/) which permits non-commercial use, reproduction and distribution of the work as published without adaptation or alteration, without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Articles Guha, Daipayan Jakubovic, Raphael Gupta, Shaurya Fehlings, Michael G. Mainprize, Todd G. Yee, Albert Yang, Victor X. D. Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study |
title | Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study |
title_full | Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study |
title_fullStr | Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study |
title_full_unstemmed | Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study |
title_short | Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study |
title_sort | intraoperative error propagation in 3-dimensional spinal navigation from nonsegmental registration: a prospective cadaveric and clinical study |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6686387/ https://www.ncbi.nlm.nih.gov/pubmed/31431874 http://dx.doi.org/10.1177/2192568218804556 |
work_keys_str_mv | AT guhadaipayan intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy AT jakubovicraphael intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy AT guptashaurya intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy AT fehlingsmichaelg intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy AT mainprizetoddg intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy AT yeealbert intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy AT yangvictorxd intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy |