Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study

STUDY DESIGN: Prospective pre-clinical and clinical cohort study. OBJECTIVES: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation e...

Descripción completa

Detalles Bibliográficos
Autores principales: Guha, Daipayan, Jakubovic, Raphael, Gupta, Shaurya, Fehlings, Michael G., Mainprize, Todd G., Yee, Albert, Yang, Victor X. D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6686387/
https://www.ncbi.nlm.nih.gov/pubmed/31431874
http://dx.doi.org/10.1177/2192568218804556
_version_ 1783442553729712128
author Guha, Daipayan
Jakubovic, Raphael
Gupta, Shaurya
Fehlings, Michael G.
Mainprize, Todd G.
Yee, Albert
Yang, Victor X. D.
author_facet Guha, Daipayan
Jakubovic, Raphael
Gupta, Shaurya
Fehlings, Michael G.
Mainprize, Todd G.
Yee, Albert
Yang, Victor X. D.
author_sort Guha, Daipayan
collection PubMed
description STUDY DESIGN: Prospective pre-clinical and clinical cohort study. OBJECTIVES: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation error as a function of distance from the registered vertebral level, and from intersegmental mobility due to surgical manipulation and patient respiration. METHODS: Navigation errors from working distant to the registered level, and from surgical manipulation, were quantified in 4 human cadavers. The 3-dimensional (3D) position of a tracked tool tip at 0 to 5 levels from the DRF, and during targeting of pedicle screw tracts, was captured in real-time by an optical navigation system. Respiration-induced vertebral motion was quantified from 10 clinical cases of open posterior instrumentation. The 3D position of a custom spinous-process clamp was tracked over 12 respiratory cycles. RESULTS: An increase in mean 3D navigation error of ≥2 mm was observed at ≥2 levels from the DRF in the cervical and lumbar spine. Mean ± SD displacement due to surgical manipulation was 1.55 ± 1.13 mm in 3D across all levels, ≥2 mm in 17.4%, 19.2%, and 38.5% of levels in the cervical, thoracic, and lumbar spine, respectively. Mean ± SD respiration-induced 3D motion was 1.96 ± 1.32 mm, greatest in the lower thoracic spine (P < .001). Tidal volume and positive end-expiratory pressure correlated positively with increased vertebral displacement. CONCLUSIONS: Vertebral motion is unaccounted for during image-guided surgery when performed at levels distant from the DRF. Navigating instrumentation within 2 levels of the DRF likely minimizes the risk of navigation error.
format Online
Article
Text
id pubmed-6686387
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-66863872019-08-20 Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study Guha, Daipayan Jakubovic, Raphael Gupta, Shaurya Fehlings, Michael G. Mainprize, Todd G. Yee, Albert Yang, Victor X. D. Global Spine J Original Articles STUDY DESIGN: Prospective pre-clinical and clinical cohort study. OBJECTIVES: Current spinal navigation systems rely on a dynamic reference frame (DRF) for image-to-patient registration and tool tracking. Working distant to a DRF may generate inaccuracy. Here we quantitate predictors of navigation error as a function of distance from the registered vertebral level, and from intersegmental mobility due to surgical manipulation and patient respiration. METHODS: Navigation errors from working distant to the registered level, and from surgical manipulation, were quantified in 4 human cadavers. The 3-dimensional (3D) position of a tracked tool tip at 0 to 5 levels from the DRF, and during targeting of pedicle screw tracts, was captured in real-time by an optical navigation system. Respiration-induced vertebral motion was quantified from 10 clinical cases of open posterior instrumentation. The 3D position of a custom spinous-process clamp was tracked over 12 respiratory cycles. RESULTS: An increase in mean 3D navigation error of ≥2 mm was observed at ≥2 levels from the DRF in the cervical and lumbar spine. Mean ± SD displacement due to surgical manipulation was 1.55 ± 1.13 mm in 3D across all levels, ≥2 mm in 17.4%, 19.2%, and 38.5% of levels in the cervical, thoracic, and lumbar spine, respectively. Mean ± SD respiration-induced 3D motion was 1.96 ± 1.32 mm, greatest in the lower thoracic spine (P < .001). Tidal volume and positive end-expiratory pressure correlated positively with increased vertebral displacement. CONCLUSIONS: Vertebral motion is unaccounted for during image-guided surgery when performed at levels distant from the DRF. Navigating instrumentation within 2 levels of the DRF likely minimizes the risk of navigation error. SAGE Publications 2018-10-09 2019-08 /pmc/articles/PMC6686387/ /pubmed/31431874 http://dx.doi.org/10.1177/2192568218804556 Text en © The Author(s) 2018 http://creativecommons.org/licenses/by-nc-nd/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 License (http://www.creativecommons.org/licenses/by-nc-nd/4.0/) which permits non-commercial use, reproduction and distribution of the work as published without adaptation or alteration, without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Articles
Guha, Daipayan
Jakubovic, Raphael
Gupta, Shaurya
Fehlings, Michael G.
Mainprize, Todd G.
Yee, Albert
Yang, Victor X. D.
Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
title Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
title_full Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
title_fullStr Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
title_full_unstemmed Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
title_short Intraoperative Error Propagation in 3-Dimensional Spinal Navigation From Nonsegmental Registration: A Prospective Cadaveric and Clinical Study
title_sort intraoperative error propagation in 3-dimensional spinal navigation from nonsegmental registration: a prospective cadaveric and clinical study
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6686387/
https://www.ncbi.nlm.nih.gov/pubmed/31431874
http://dx.doi.org/10.1177/2192568218804556
work_keys_str_mv AT guhadaipayan intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy
AT jakubovicraphael intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy
AT guptashaurya intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy
AT fehlingsmichaelg intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy
AT mainprizetoddg intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy
AT yeealbert intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy
AT yangvictorxd intraoperativeerrorpropagationin3dimensionalspinalnavigationfromnonsegmentalregistrationaprospectivecadavericandclinicalstudy