Impact of device programming on the success of the first anti-tachycardia pacing therapy: An anonymized large-scale study
BACKGROUND: Antitachycardia pacing (ATP) is an effective treatment for ventricular tachycardia (VT). We evaluated the efficacy of different ATP programs based on a large remote monitoring data set from patients with implantable cardioverter-defibrillators (ICDs). METHODS: A dataset from 18,679 ICD p...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6687124/ https://www.ncbi.nlm.nih.gov/pubmed/31393871 http://dx.doi.org/10.1371/journal.pone.0219533 |
Sumario: | BACKGROUND: Antitachycardia pacing (ATP) is an effective treatment for ventricular tachycardia (VT). We evaluated the efficacy of different ATP programs based on a large remote monitoring data set from patients with implantable cardioverter-defibrillators (ICDs). METHODS: A dataset from 18,679 ICD patients was used to evaluate the first delivered ATP treatment. We considered all device programs that were used for at least 50 patients, leaving us with 7 different programs and a total of 32,045 episodes. We used the two-proportions z-test (α = 0.01) to compare the probability of success and the probability for acceleration in each group with the corresponding values of the default setting. RESULTS: Overall, the first ATP treatment terminated in 78.4%–97.5% of episodes with slow VT and 81.5%–91.1% of episodes with fast VT. The default setting of the ATP programs with the number of sequences S = 3 was applied to treat 30.1% of the slow and 36.6% of the fast episodes. Reducing the maximum number of sequences to S = 2 decreased the success rate for slow VT (P < 0.0001, h = 0.38), while the setting S = 4 resulted in the highest success rate of 97.5% (P < 0.0001, h = 0.27). CONCLUSION: While the default programs performed well, we found that increasing the number of sequences from 3 to 4 was a promising option to improve the overall ATP performance. |
---|