An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone

BACKGROUND: There are no medical drill specifications capable of achieving bone drilling in a short time under low-thrust and low-speed drilling. Gekkou-drill® is an industrial drill that enables drilling with low cutting resistance by its characteristic point design. Our aims were to develop Gekkou...

Descripción completa

Detalles Bibliográficos
Autores principales: Enokida, Makoto, Kanaya, Haruhisa, Uehara, Kazutake, Ueki, Masaru, Nagashima, Hideki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6690560/
https://www.ncbi.nlm.nih.gov/pubmed/31417971
http://dx.doi.org/10.1016/j.heliyon.2019.e02189
_version_ 1783443208699641856
author Enokida, Makoto
Kanaya, Haruhisa
Uehara, Kazutake
Ueki, Masaru
Nagashima, Hideki
author_facet Enokida, Makoto
Kanaya, Haruhisa
Uehara, Kazutake
Ueki, Masaru
Nagashima, Hideki
author_sort Enokida, Makoto
collection PubMed
description BACKGROUND: There are no medical drill specifications capable of achieving bone drilling in a short time under low-thrust and low-speed drilling. Gekkou-drill® is an industrial drill that enables drilling with low cutting resistance by its characteristic point design. Our aims were to develop Gekkou-modified drills by processing to the points of currently available medical drills and to verify whether these modified drills enable less invasive drilling procedure for bone tissue in thermal exposure compared with unmodified medical drills. MATERIALS AND METHODS: Two commercially available 3.2-mm drills were compared before and after Gekkou modification. Drilling of pig tibias was performed at speeds of 300, 800, and 1,500 rpm and a uniform thrust force of 10 N. Temperature at the entry point for bone drilling was measured using a digital thermometer system. The feed rates were calculated using cortical thickness and monitoring data of the digital force gauge. RESULTS: Two unmodified drills could not penetrate the cortical bone on the near side at 300 rpm, even after 5 min of drilling. The maximum temperatures with modified drills were 54.6 °C and 46.2 °C at 300 rpm. At medium to high speeds, those were statistically significantly lower than with unmodified drills (58.5 °C vs. 90.5 °C at 800 rpm, 62.6 °C vs. 80.8 °C and 73.9 °C vs. 104.6 °C at 1,500 rpm). The feed rates for modified drills were 4.9–6.9 times as high as unmodified drills at 800 rpm, and 3.4 to 4.5 times at 1,500 rpm. On the other hand, the feed rates of modified drills at 300 rpm were equal to or higher than those of unmodified drills at 1500 rpm. CONCLUSION: Gekkou-modified drills clearly suppressed the temperature rise and increased the feed rate compared with conventional drills. Furthermore, it was notable that these modified drills had higher performance even at conditions of low thrust and low speed.
format Online
Article
Text
id pubmed-6690560
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-66905602019-08-15 An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone Enokida, Makoto Kanaya, Haruhisa Uehara, Kazutake Ueki, Masaru Nagashima, Hideki Heliyon Article BACKGROUND: There are no medical drill specifications capable of achieving bone drilling in a short time under low-thrust and low-speed drilling. Gekkou-drill® is an industrial drill that enables drilling with low cutting resistance by its characteristic point design. Our aims were to develop Gekkou-modified drills by processing to the points of currently available medical drills and to verify whether these modified drills enable less invasive drilling procedure for bone tissue in thermal exposure compared with unmodified medical drills. MATERIALS AND METHODS: Two commercially available 3.2-mm drills were compared before and after Gekkou modification. Drilling of pig tibias was performed at speeds of 300, 800, and 1,500 rpm and a uniform thrust force of 10 N. Temperature at the entry point for bone drilling was measured using a digital thermometer system. The feed rates were calculated using cortical thickness and monitoring data of the digital force gauge. RESULTS: Two unmodified drills could not penetrate the cortical bone on the near side at 300 rpm, even after 5 min of drilling. The maximum temperatures with modified drills were 54.6 °C and 46.2 °C at 300 rpm. At medium to high speeds, those were statistically significantly lower than with unmodified drills (58.5 °C vs. 90.5 °C at 800 rpm, 62.6 °C vs. 80.8 °C and 73.9 °C vs. 104.6 °C at 1,500 rpm). The feed rates for modified drills were 4.9–6.9 times as high as unmodified drills at 800 rpm, and 3.4 to 4.5 times at 1,500 rpm. On the other hand, the feed rates of modified drills at 300 rpm were equal to or higher than those of unmodified drills at 1500 rpm. CONCLUSION: Gekkou-modified drills clearly suppressed the temperature rise and increased the feed rate compared with conventional drills. Furthermore, it was notable that these modified drills had higher performance even at conditions of low thrust and low speed. Elsevier 2019-08-02 /pmc/articles/PMC6690560/ /pubmed/31417971 http://dx.doi.org/10.1016/j.heliyon.2019.e02189 Text en © 2019 The Author(s) http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Article
Enokida, Makoto
Kanaya, Haruhisa
Uehara, Kazutake
Ueki, Masaru
Nagashima, Hideki
An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
title An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
title_full An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
title_fullStr An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
title_full_unstemmed An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
title_short An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
title_sort experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6690560/
https://www.ncbi.nlm.nih.gov/pubmed/31417971
http://dx.doi.org/10.1016/j.heliyon.2019.e02189
work_keys_str_mv AT enokidamakoto anexperimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT kanayaharuhisa anexperimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT ueharakazutake anexperimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT uekimasaru anexperimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT nagashimahideki anexperimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT enokidamakoto experimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT kanayaharuhisa experimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT ueharakazutake experimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT uekimasaru experimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone
AT nagashimahideki experimentalcomparativestudyofdrillingefficiencyandtemperatureelevationwithunmodifiedandmodifiedmedicaldrillsinpigtibiabone