Cargando…
Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up
AIMS: Riata(®) implantable cardioverter-defibrillator (ICD) leads from St. Jude Medical are prone to malfunction. This study aimed to describe the rate of this lead's malfunction in a very long-term follow-up. METHODS: This single-centre observational study included 50 patients who received a R...
Autores principales: | , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6697461/ https://www.ncbi.nlm.nih.gov/pubmed/30794927 http://dx.doi.org/10.1016/j.ipej.2019.02.005 |
_version_ | 1783444389956157440 |
---|---|
author | San Antonio, Rodolfo Guasch, Eduard Chipa-Ccasani, Fredy Apolo, José Pujol-López, Margarida Fernández, Hael Trotta, Omar Niebla, Mireia Borràs, Roger Trucco, Emilce Arbelo, Elena Roca-Luque, Ivo Brugada, Josep Mont, Lluís Tolosana, José María |
author_facet | San Antonio, Rodolfo Guasch, Eduard Chipa-Ccasani, Fredy Apolo, José Pujol-López, Margarida Fernández, Hael Trotta, Omar Niebla, Mireia Borràs, Roger Trucco, Emilce Arbelo, Elena Roca-Luque, Ivo Brugada, Josep Mont, Lluís Tolosana, José María |
author_sort | San Antonio, Rodolfo |
collection | PubMed |
description | AIMS: Riata(®) implantable cardioverter-defibrillator (ICD) leads from St. Jude Medical are prone to malfunction. This study aimed to describe the rate of this lead's malfunction in a very long-term follow-up. METHODS: This single-centre observational study included 50 patients who received a Riata 7Fr dual-coil lead between 2003 and 2008. Follow-up was conducted both in person and remotely, and analysed at 8-month intervals. We evaluated the rates of cable externalization (CE), electrical failure (EF), and the interaction of these two complications. Structural lead failure was defined as radiographic CE. Oversensing of non-cardiac signal or sudden changes in impedance, sensing, or pacing thresholds constituted EF. RESULTS: During a mean follow-up of 10.2 ± 2.9 years, 16 patients (32%) died. We observed lead malfunction in 13 patients (26%): three (23%) due to CE, six (46%) to EF and four (31%) to both complications. Of the malfunctioning leads, 77% failed after seven years of follow-up. The incidence rate (IR) of overall malfunction per 100 patients per year was 0.9 during the first seven years post-implantation, increased to 7.0 after the 7th year and more than doubled (to 16.7) after 10 years. Beyond seven years post-implantation, IR per 100 patient-years increased in both EF and CE (from 0.6 to 5.6 vs. 0.3 to 4.2, respectively). Presence of CE was associated with a 4-fold increase in the proportion of EF. CONCLUSION: The incidence of Riata ICD lead malfunction, both for EF and CE, increased dramatically after seven years and then more than doubled after 10 years post-implantation. |
format | Online Article Text |
id | pubmed-6697461 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-66974612019-08-19 Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up San Antonio, Rodolfo Guasch, Eduard Chipa-Ccasani, Fredy Apolo, José Pujol-López, Margarida Fernández, Hael Trotta, Omar Niebla, Mireia Borràs, Roger Trucco, Emilce Arbelo, Elena Roca-Luque, Ivo Brugada, Josep Mont, Lluís Tolosana, José María Indian Pacing Electrophysiol J Original Article AIMS: Riata(®) implantable cardioverter-defibrillator (ICD) leads from St. Jude Medical are prone to malfunction. This study aimed to describe the rate of this lead's malfunction in a very long-term follow-up. METHODS: This single-centre observational study included 50 patients who received a Riata 7Fr dual-coil lead between 2003 and 2008. Follow-up was conducted both in person and remotely, and analysed at 8-month intervals. We evaluated the rates of cable externalization (CE), electrical failure (EF), and the interaction of these two complications. Structural lead failure was defined as radiographic CE. Oversensing of non-cardiac signal or sudden changes in impedance, sensing, or pacing thresholds constituted EF. RESULTS: During a mean follow-up of 10.2 ± 2.9 years, 16 patients (32%) died. We observed lead malfunction in 13 patients (26%): three (23%) due to CE, six (46%) to EF and four (31%) to both complications. Of the malfunctioning leads, 77% failed after seven years of follow-up. The incidence rate (IR) of overall malfunction per 100 patients per year was 0.9 during the first seven years post-implantation, increased to 7.0 after the 7th year and more than doubled (to 16.7) after 10 years. Beyond seven years post-implantation, IR per 100 patient-years increased in both EF and CE (from 0.6 to 5.6 vs. 0.3 to 4.2, respectively). Presence of CE was associated with a 4-fold increase in the proportion of EF. CONCLUSION: The incidence of Riata ICD lead malfunction, both for EF and CE, increased dramatically after seven years and then more than doubled after 10 years post-implantation. Elsevier 2019-02-19 /pmc/articles/PMC6697461/ /pubmed/30794927 http://dx.doi.org/10.1016/j.ipej.2019.02.005 Text en © 2019 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Original Article San Antonio, Rodolfo Guasch, Eduard Chipa-Ccasani, Fredy Apolo, José Pujol-López, Margarida Fernández, Hael Trotta, Omar Niebla, Mireia Borràs, Roger Trucco, Emilce Arbelo, Elena Roca-Luque, Ivo Brugada, Josep Mont, Lluís Tolosana, José María Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
title | Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
title_full | Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
title_fullStr | Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
title_full_unstemmed | Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
title_short | Failure-free survival of the Riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
title_sort | failure-free survival of the riata implantable cardioverter-defibrillator lead after a very long-term follow-up |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6697461/ https://www.ncbi.nlm.nih.gov/pubmed/30794927 http://dx.doi.org/10.1016/j.ipej.2019.02.005 |
work_keys_str_mv | AT sanantoniorodolfo failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT guascheduard failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT chipaccasanifredy failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT apolojose failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT pujollopezmargarida failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT fernandezhael failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT trottaomar failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT nieblamireia failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT borrasroger failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT truccoemilce failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT arbeloelena failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT rocaluqueivo failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT brugadajosep failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT montlluis failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup AT tolosanajosemaria failurefreesurvivaloftheriataimplantablecardioverterdefibrillatorleadafteraverylongtermfollowup |