Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
PURPOSES: The newly released Protura 6 degrees-of-freedom couch (CIVCO) has limited quality assurance protocols and pertinent publications. Herein, we report our experiences of the Protura system acceptance, commissioning, and quality assurance. METHODS: The Protura system integration was tested wit...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6704415/ https://www.ncbi.nlm.nih.gov/pubmed/31434547 http://dx.doi.org/10.1177/1533033819870778 |
_version_ | 1783445502182817792 |
---|---|
author | Chan, Maria F. Lim, Seng-Boh Li, Xiang Tang, Xiaoli Zhang, Peng Shi, Chengyu |
author_facet | Chan, Maria F. Lim, Seng-Boh Li, Xiang Tang, Xiaoli Zhang, Peng Shi, Chengyu |
author_sort | Chan, Maria F. |
collection | PubMed |
description | PURPOSES: The newly released Protura 6 degrees-of-freedom couch (CIVCO) has limited quality assurance protocols and pertinent publications. Herein, we report our experiences of the Protura system acceptance, commissioning, and quality assurance. METHODS: The Protura system integration was tested with peripheral equipment on the following items: couch movement range limit, 6 degrees-of-freedom movement accuracy, weight test and couch sagging, system connection with Linac, isocentricity of couch and rotation alignment, kV and cone-beam computed tomography imaging of HexaCHECK with MIMI phantom (Standard Imaging), and an in-house custom 6 degrees-of-freedom quality assurance phantom. A couch transmission measurement was also performed. RESULTS: The vertical, longitudinal, and lateral ranges of the 6 degrees-of-freedom couch pedestal are 43.9 to 0.0 cm, 24.6 to 149.5 cm, −20.6 to 20.7 cm, respectively. The couch movement accuracy was within 1 mm in all directions. The couch sagging with a 200 lbs (∼91 kg) evenly distributed object is 1.0 cm and 0.4° pitch in the distal end of the couch. The isocentricity of the couch was about 0.5 mm in diameter of all crosshair projections on the couch isocenter level, and the largest couch rotation alignment observed was (0.3°) at the couch angle of 90°. The deviation from the reference position (zero position) of the HexaCHECK phantom, measured by matching the cone-beam computed tomography with the reference planning computed tomography, was found to be below 0.2 mm in the anterior–posterior and right–left dimensions, 0.4 mm in superior–inferior dimension, and 0.1° in roll, pitch, and yaw directions. CONCLUSIONS: A 6 degrees-of-freedom quality assurance phantom is helpful for the commissioning and routine quality assurance tests. Due to the third-party integration with Linac, the system is prone to “double-correction” errors. A rigorous quality assurance program is the key to a successful clinical implementation of the Protura system. |
format | Online Article Text |
id | pubmed-6704415 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-67044152019-08-29 Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy Chan, Maria F. Lim, Seng-Boh Li, Xiang Tang, Xiaoli Zhang, Peng Shi, Chengyu Technol Cancer Res Treat Original Article PURPOSES: The newly released Protura 6 degrees-of-freedom couch (CIVCO) has limited quality assurance protocols and pertinent publications. Herein, we report our experiences of the Protura system acceptance, commissioning, and quality assurance. METHODS: The Protura system integration was tested with peripheral equipment on the following items: couch movement range limit, 6 degrees-of-freedom movement accuracy, weight test and couch sagging, system connection with Linac, isocentricity of couch and rotation alignment, kV and cone-beam computed tomography imaging of HexaCHECK with MIMI phantom (Standard Imaging), and an in-house custom 6 degrees-of-freedom quality assurance phantom. A couch transmission measurement was also performed. RESULTS: The vertical, longitudinal, and lateral ranges of the 6 degrees-of-freedom couch pedestal are 43.9 to 0.0 cm, 24.6 to 149.5 cm, −20.6 to 20.7 cm, respectively. The couch movement accuracy was within 1 mm in all directions. The couch sagging with a 200 lbs (∼91 kg) evenly distributed object is 1.0 cm and 0.4° pitch in the distal end of the couch. The isocentricity of the couch was about 0.5 mm in diameter of all crosshair projections on the couch isocenter level, and the largest couch rotation alignment observed was (0.3°) at the couch angle of 90°. The deviation from the reference position (zero position) of the HexaCHECK phantom, measured by matching the cone-beam computed tomography with the reference planning computed tomography, was found to be below 0.2 mm in the anterior–posterior and right–left dimensions, 0.4 mm in superior–inferior dimension, and 0.1° in roll, pitch, and yaw directions. CONCLUSIONS: A 6 degrees-of-freedom quality assurance phantom is helpful for the commissioning and routine quality assurance tests. Due to the third-party integration with Linac, the system is prone to “double-correction” errors. A rigorous quality assurance program is the key to a successful clinical implementation of the Protura system. SAGE Publications 2019-08-21 /pmc/articles/PMC6704415/ /pubmed/31434547 http://dx.doi.org/10.1177/1533033819870778 Text en © The Author(s) 2019 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Article Chan, Maria F. Lim, Seng-Boh Li, Xiang Tang, Xiaoli Zhang, Peng Shi, Chengyu Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy |
title | Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy |
title_full | Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy |
title_fullStr | Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy |
title_full_unstemmed | Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy |
title_short | Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy |
title_sort | commissioning and evaluation of a third-party 6 degrees-of-freedom couch used in radiotherapy |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6704415/ https://www.ncbi.nlm.nih.gov/pubmed/31434547 http://dx.doi.org/10.1177/1533033819870778 |
work_keys_str_mv | AT chanmariaf commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy AT limsengboh commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy AT lixiang commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy AT tangxiaoli commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy AT zhangpeng commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy AT shichengyu commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy |