Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy

PURPOSES: The newly released Protura 6 degrees-of-freedom couch (CIVCO) has limited quality assurance protocols and pertinent publications. Herein, we report our experiences of the Protura system acceptance, commissioning, and quality assurance. METHODS: The Protura system integration was tested wit...

Descripción completa

Detalles Bibliográficos
Autores principales: Chan, Maria F., Lim, Seng-Boh, Li, Xiang, Tang, Xiaoli, Zhang, Peng, Shi, Chengyu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6704415/
https://www.ncbi.nlm.nih.gov/pubmed/31434547
http://dx.doi.org/10.1177/1533033819870778
_version_ 1783445502182817792
author Chan, Maria F.
Lim, Seng-Boh
Li, Xiang
Tang, Xiaoli
Zhang, Peng
Shi, Chengyu
author_facet Chan, Maria F.
Lim, Seng-Boh
Li, Xiang
Tang, Xiaoli
Zhang, Peng
Shi, Chengyu
author_sort Chan, Maria F.
collection PubMed
description PURPOSES: The newly released Protura 6 degrees-of-freedom couch (CIVCO) has limited quality assurance protocols and pertinent publications. Herein, we report our experiences of the Protura system acceptance, commissioning, and quality assurance. METHODS: The Protura system integration was tested with peripheral equipment on the following items: couch movement range limit, 6 degrees-of-freedom movement accuracy, weight test and couch sagging, system connection with Linac, isocentricity of couch and rotation alignment, kV and cone-beam computed tomography imaging of HexaCHECK with MIMI phantom (Standard Imaging), and an in-house custom 6 degrees-of-freedom quality assurance phantom. A couch transmission measurement was also performed. RESULTS: The vertical, longitudinal, and lateral ranges of the 6 degrees-of-freedom couch pedestal are 43.9 to 0.0 cm, 24.6 to 149.5 cm, −20.6 to 20.7 cm, respectively. The couch movement accuracy was within 1 mm in all directions. The couch sagging with a 200 lbs (∼91 kg) evenly distributed object is 1.0 cm and 0.4° pitch in the distal end of the couch. The isocentricity of the couch was about 0.5 mm in diameter of all crosshair projections on the couch isocenter level, and the largest couch rotation alignment observed was (0.3°) at the couch angle of 90°. The deviation from the reference position (zero position) of the HexaCHECK phantom, measured by matching the cone-beam computed tomography with the reference planning computed tomography, was found to be below 0.2 mm in the anterior–posterior and right–left dimensions, 0.4 mm in superior–inferior dimension, and 0.1° in roll, pitch, and yaw directions. CONCLUSIONS: A 6 degrees-of-freedom quality assurance phantom is helpful for the commissioning and routine quality assurance tests. Due to the third-party integration with Linac, the system is prone to “double-correction” errors. A rigorous quality assurance program is the key to a successful clinical implementation of the Protura system.
format Online
Article
Text
id pubmed-6704415
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-67044152019-08-29 Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy Chan, Maria F. Lim, Seng-Boh Li, Xiang Tang, Xiaoli Zhang, Peng Shi, Chengyu Technol Cancer Res Treat Original Article PURPOSES: The newly released Protura 6 degrees-of-freedom couch (CIVCO) has limited quality assurance protocols and pertinent publications. Herein, we report our experiences of the Protura system acceptance, commissioning, and quality assurance. METHODS: The Protura system integration was tested with peripheral equipment on the following items: couch movement range limit, 6 degrees-of-freedom movement accuracy, weight test and couch sagging, system connection with Linac, isocentricity of couch and rotation alignment, kV and cone-beam computed tomography imaging of HexaCHECK with MIMI phantom (Standard Imaging), and an in-house custom 6 degrees-of-freedom quality assurance phantom. A couch transmission measurement was also performed. RESULTS: The vertical, longitudinal, and lateral ranges of the 6 degrees-of-freedom couch pedestal are 43.9 to 0.0 cm, 24.6 to 149.5 cm, −20.6 to 20.7 cm, respectively. The couch movement accuracy was within 1 mm in all directions. The couch sagging with a 200 lbs (∼91 kg) evenly distributed object is 1.0 cm and 0.4° pitch in the distal end of the couch. The isocentricity of the couch was about 0.5 mm in diameter of all crosshair projections on the couch isocenter level, and the largest couch rotation alignment observed was (0.3°) at the couch angle of 90°. The deviation from the reference position (zero position) of the HexaCHECK phantom, measured by matching the cone-beam computed tomography with the reference planning computed tomography, was found to be below 0.2 mm in the anterior–posterior and right–left dimensions, 0.4 mm in superior–inferior dimension, and 0.1° in roll, pitch, and yaw directions. CONCLUSIONS: A 6 degrees-of-freedom quality assurance phantom is helpful for the commissioning and routine quality assurance tests. Due to the third-party integration with Linac, the system is prone to “double-correction” errors. A rigorous quality assurance program is the key to a successful clinical implementation of the Protura system. SAGE Publications 2019-08-21 /pmc/articles/PMC6704415/ /pubmed/31434547 http://dx.doi.org/10.1177/1533033819870778 Text en © The Author(s) 2019 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Article
Chan, Maria F.
Lim, Seng-Boh
Li, Xiang
Tang, Xiaoli
Zhang, Peng
Shi, Chengyu
Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
title Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
title_full Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
title_fullStr Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
title_full_unstemmed Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
title_short Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy
title_sort commissioning and evaluation of a third-party 6 degrees-of-freedom couch used in radiotherapy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6704415/
https://www.ncbi.nlm.nih.gov/pubmed/31434547
http://dx.doi.org/10.1177/1533033819870778
work_keys_str_mv AT chanmariaf commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy
AT limsengboh commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy
AT lixiang commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy
AT tangxiaoli commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy
AT zhangpeng commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy
AT shichengyu commissioningandevaluationofathirdparty6degreesoffreedomcouchusedinradiotherapy