Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

BACKGROUND: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated. OBJECTIVE: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. MATERIAL AND METHODS: The phantom, structure sets, VMAT and IMRT beam parameter s...

Descripción completa

Detalles Bibliográficos
Autores principales: A., Nainggolan, S A., Pawiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Journal of Biomedical Physics and Engineering 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6709355/
https://www.ncbi.nlm.nih.gov/pubmed/31531292
http://dx.doi.org/10.31661/jbpe.v0i0.839
_version_ 1783446191432794112
author A., Nainggolan
S A., Pawiro
author_facet A., Nainggolan
S A., Pawiro
author_sort A., Nainggolan
collection PubMed
description BACKGROUND: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated. OBJECTIVE: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. MATERIAL AND METHODS: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to create local treatment plans of VMAT and IMRT. The local planning results were compared with the results of TG 119. Point measurement at high and low dose regions were measured using three ionization chambers with different active volumes (CC01, CC13, FC65G). The composite dose was measured by a 2D detector array and analyzed for the percentage of points passing the gamma criteria of 3 % dose difference (DD) and 3 mm distance-to-agreement (DTA) and 2 % DD and 2 mm DTA. RESULTS: The local treatment plans of VMAT and IMRT capable to meet the dose goals criteria set by TG 119 except for C-shape hard. Three ionization chambers with various active volumes for point measurement showed an increase in the confidence limit (CL), the larger the active volume was found proportional to increase the value of CL. The results obtained from ion chambers CC01 and CC13 could met the dose criteria set by TG 119, but results obtained from ion chamber FC65G fail the criteria. All gamma evaluation results show more than 95% data points pass the criteria of 3% DD and 3 mm DTA and the gamma index CL results fall within the TG 119 criteria, which is below 12.4. CONCLUSION: TG 119 methodology and recommendations have successfully been used to evaluate commissioning accuracy of VMAT. The CL value of the study could be used as a reference and recommendation to evaluate the accuracy and integrity of treatment planning and treatment delivery systems of VMAT and IMRT.
format Online
Article
Text
id pubmed-6709355
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Journal of Biomedical Physics and Engineering
record_format MEDLINE/PubMed
spelling pubmed-67093552019-09-17 Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol A., Nainggolan S A., Pawiro J Biomed Phys Eng Original Article BACKGROUND: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated. OBJECTIVE: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. MATERIAL AND METHODS: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to create local treatment plans of VMAT and IMRT. The local planning results were compared with the results of TG 119. Point measurement at high and low dose regions were measured using three ionization chambers with different active volumes (CC01, CC13, FC65G). The composite dose was measured by a 2D detector array and analyzed for the percentage of points passing the gamma criteria of 3 % dose difference (DD) and 3 mm distance-to-agreement (DTA) and 2 % DD and 2 mm DTA. RESULTS: The local treatment plans of VMAT and IMRT capable to meet the dose goals criteria set by TG 119 except for C-shape hard. Three ionization chambers with various active volumes for point measurement showed an increase in the confidence limit (CL), the larger the active volume was found proportional to increase the value of CL. The results obtained from ion chambers CC01 and CC13 could met the dose criteria set by TG 119, but results obtained from ion chamber FC65G fail the criteria. All gamma evaluation results show more than 95% data points pass the criteria of 3% DD and 3 mm DTA and the gamma index CL results fall within the TG 119 criteria, which is below 12.4. CONCLUSION: TG 119 methodology and recommendations have successfully been used to evaluate commissioning accuracy of VMAT. The CL value of the study could be used as a reference and recommendation to evaluate the accuracy and integrity of treatment planning and treatment delivery systems of VMAT and IMRT. Journal of Biomedical Physics and Engineering 2019-08-01 /pmc/articles/PMC6709355/ /pubmed/31531292 http://dx.doi.org/10.31661/jbpe.v0i0.839 Text en Copyright: © Journal of Biomedical Physics and Engineering http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
A., Nainggolan
S A., Pawiro
Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
title Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
title_full Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
title_fullStr Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
title_full_unstemmed Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
title_short Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
title_sort dosimetric evaluation of volumetric modulated arc therapy (vmat) and intensity modulated radiotherapy (imrt) using aapm tg 119 protocol
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6709355/
https://www.ncbi.nlm.nih.gov/pubmed/31531292
http://dx.doi.org/10.31661/jbpe.v0i0.839
work_keys_str_mv AT anainggolan dosimetricevaluationofvolumetricmodulatedarctherapyvmatandintensitymodulatedradiotherapyimrtusingaapmtg119protocol
AT sapawiro dosimetricevaluationofvolumetricmodulatedarctherapyvmatandintensitymodulatedradiotherapyimrtusingaapmtg119protocol