Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife
PURPOSE: To compare plan quality and delivery efficiency of noncoplanar volumetric modulated arc therapy with coplanar volumetric modulated arc therapy, intensity-modulated radiation therapy, and CyberKnife for multiple brain metastases. METHODS: For 15 patients with multiple brain metastases, nonco...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6710677/ https://www.ncbi.nlm.nih.gov/pubmed/31451059 http://dx.doi.org/10.1177/1533033819871621 |
_version_ | 1783446384283746304 |
---|---|
author | Zhang, Shuming Yang, Ruijie Shi, Chengyu Li, Jiaqi Zhuang, Hongqing Tian, Suqing Wang, Junjie |
author_facet | Zhang, Shuming Yang, Ruijie Shi, Chengyu Li, Jiaqi Zhuang, Hongqing Tian, Suqing Wang, Junjie |
author_sort | Zhang, Shuming |
collection | PubMed |
description | PURPOSE: To compare plan quality and delivery efficiency of noncoplanar volumetric modulated arc therapy with coplanar volumetric modulated arc therapy, intensity-modulated radiation therapy, and CyberKnife for multiple brain metastases. METHODS: For 15 patients with multiple brain metastases, noncoplanar volumetric modulated arc therapy, coplanar volumetric modulated arc therapy, intensity-modulated radiation therapy, and CyberKnife plans with a prescription dose of 30 Gy in 3 fractions were generated. Noncoplanar volumetric modulated arc therapy and coplanar volumetric modulated arc therapy plans consisted of 4 noncoplanar arcs and 2 full coplanar arcs, respectively. Intensity-modulated radiation therapy plans consisted of 7 coplanar fields. CyberKnife plans used skull tracking to ensure accurate position. All plans were generated to cover 95% target volume with prescription dose. Gradient index, conformity index, normal brain tissue volume (V (3Gy) − V (24Gy)), monitor units, and beam on time were evaluated. RESULTS: Gradient index was the lowest for CyberKnife (3.49 ± 0.65), followed by noncoplanar volumetric modulated arc therapy (4.21 ± 1.38), coplanar volumetric modulated arc therapy (4.87 ± 1.35), and intensity-modulated radiation therapy (5.36 ± 1.98). Conformity index was the largest for noncoplanar volumetric modulated arc therapy (0.87 ± 0.03), followed by coplanar volumetric modulated arc therapy (0.86 ± 0.04), CyberKnife (0.86 ± 0.07), and intensity-modulated radiation therapy (0.85 ± 0.05). Normal brain tissue volume at high-to-moderate dose spreads (V (24Gy) − V (9Gy)) was significantly reduced in noncoplanar volumetric modulated arc therapy over that of intensity-modulated radiation therapy and coplanar volumetric modulated arc therapy. Normal brain tissue volume for noncoplanar volumetric modulated arc therapy was comparable with noncoplanar volumetric modulated arc therapy at high-dose level (V (24Gy) − V (15Gy)) and larger than CyberKnife at moderate-to-low dose level (V (12Gy) − V (3Gy)). Monitor units was highest for CyberKnife (28 733.59 ± 7197.85), followed by intensity-modulated radiation therapy (4128.40 ± 1185.38), noncoplanar volumetric modulated arc therapy (3105.20 ± 371.23), and coplanar volumetric modulated arc therapy (2997.27 ± 446.84). Beam on time was longest for CyberKnife (30.25 ± 7.32 minutes), followed by intensity-modulated radiation therapy (2.95 ± 0.85 minutes), noncoplanar volumetric modulated arc therapy (2.61 ± 0.07 minutes), and coplanar volumetric modulated arc therapy (2.30 ± 0.23 minutes). CONCLUSION: For brain metastases far away from organs-at-risk, noncoplanar volumetric modulated arc therapy generated more rapid dose falloff and higher conformity compared to intensity-modulated radiation therapy and coplanar volumetric modulated arc therapy. Noncoplanar volumetric modulated arc therapy provided a comparable dose falloff with CyberKnife at high-dose level and a slower dose falloff than CyberKnife at moderate-to-low dose level. Noncoplanar volumetric modulated arc therapy plans had less monitor units and shorter beam on time than CyberKnife plans. |
format | Online Article Text |
id | pubmed-6710677 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-67106772019-09-05 Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife Zhang, Shuming Yang, Ruijie Shi, Chengyu Li, Jiaqi Zhuang, Hongqing Tian, Suqing Wang, Junjie Technol Cancer Res Treat Original Article PURPOSE: To compare plan quality and delivery efficiency of noncoplanar volumetric modulated arc therapy with coplanar volumetric modulated arc therapy, intensity-modulated radiation therapy, and CyberKnife for multiple brain metastases. METHODS: For 15 patients with multiple brain metastases, noncoplanar volumetric modulated arc therapy, coplanar volumetric modulated arc therapy, intensity-modulated radiation therapy, and CyberKnife plans with a prescription dose of 30 Gy in 3 fractions were generated. Noncoplanar volumetric modulated arc therapy and coplanar volumetric modulated arc therapy plans consisted of 4 noncoplanar arcs and 2 full coplanar arcs, respectively. Intensity-modulated radiation therapy plans consisted of 7 coplanar fields. CyberKnife plans used skull tracking to ensure accurate position. All plans were generated to cover 95% target volume with prescription dose. Gradient index, conformity index, normal brain tissue volume (V (3Gy) − V (24Gy)), monitor units, and beam on time were evaluated. RESULTS: Gradient index was the lowest for CyberKnife (3.49 ± 0.65), followed by noncoplanar volumetric modulated arc therapy (4.21 ± 1.38), coplanar volumetric modulated arc therapy (4.87 ± 1.35), and intensity-modulated radiation therapy (5.36 ± 1.98). Conformity index was the largest for noncoplanar volumetric modulated arc therapy (0.87 ± 0.03), followed by coplanar volumetric modulated arc therapy (0.86 ± 0.04), CyberKnife (0.86 ± 0.07), and intensity-modulated radiation therapy (0.85 ± 0.05). Normal brain tissue volume at high-to-moderate dose spreads (V (24Gy) − V (9Gy)) was significantly reduced in noncoplanar volumetric modulated arc therapy over that of intensity-modulated radiation therapy and coplanar volumetric modulated arc therapy. Normal brain tissue volume for noncoplanar volumetric modulated arc therapy was comparable with noncoplanar volumetric modulated arc therapy at high-dose level (V (24Gy) − V (15Gy)) and larger than CyberKnife at moderate-to-low dose level (V (12Gy) − V (3Gy)). Monitor units was highest for CyberKnife (28 733.59 ± 7197.85), followed by intensity-modulated radiation therapy (4128.40 ± 1185.38), noncoplanar volumetric modulated arc therapy (3105.20 ± 371.23), and coplanar volumetric modulated arc therapy (2997.27 ± 446.84). Beam on time was longest for CyberKnife (30.25 ± 7.32 minutes), followed by intensity-modulated radiation therapy (2.95 ± 0.85 minutes), noncoplanar volumetric modulated arc therapy (2.61 ± 0.07 minutes), and coplanar volumetric modulated arc therapy (2.30 ± 0.23 minutes). CONCLUSION: For brain metastases far away from organs-at-risk, noncoplanar volumetric modulated arc therapy generated more rapid dose falloff and higher conformity compared to intensity-modulated radiation therapy and coplanar volumetric modulated arc therapy. Noncoplanar volumetric modulated arc therapy provided a comparable dose falloff with CyberKnife at high-dose level and a slower dose falloff than CyberKnife at moderate-to-low dose level. Noncoplanar volumetric modulated arc therapy plans had less monitor units and shorter beam on time than CyberKnife plans. SAGE Publications 2019-08-26 /pmc/articles/PMC6710677/ /pubmed/31451059 http://dx.doi.org/10.1177/1533033819871621 Text en © The Author(s) 2019 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Article Zhang, Shuming Yang, Ruijie Shi, Chengyu Li, Jiaqi Zhuang, Hongqing Tian, Suqing Wang, Junjie Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife |
title | Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife |
title_full | Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife |
title_fullStr | Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife |
title_full_unstemmed | Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife |
title_short | Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife |
title_sort | noncoplanar vmat for brain metastases: a plan quality and delivery efficiency comparison with coplanar vmat, imrt, and cyberknife |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6710677/ https://www.ncbi.nlm.nih.gov/pubmed/31451059 http://dx.doi.org/10.1177/1533033819871621 |
work_keys_str_mv | AT zhangshuming noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife AT yangruijie noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife AT shichengyu noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife AT lijiaqi noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife AT zhuanghongqing noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife AT tiansuqing noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife AT wangjunjie noncoplanarvmatforbrainmetastasesaplanqualityanddeliveryefficiencycomparisonwithcoplanarvmatimrtandcyberknife |