Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic
Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients’ self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and in...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6714735/ https://www.ncbi.nlm.nih.gov/pubmed/29374969 http://dx.doi.org/10.1177/1076029617752247 |
_version_ | 1783447111190183936 |
---|---|
author | Moiz, Bushra Rashid, Anila Hasan, Muhammad Jafri, Lena Raheem, Ahmed |
author_facet | Moiz, Bushra Rashid, Anila Hasan, Muhammad Jafri, Lena Raheem, Ahmed |
author_sort | Moiz, Bushra |
collection | PubMed |
description | Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients’ self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and investigated whether the results of POCT coagulometer can reliably be used without being confirmed by standard laboratory analyzer. A total of 200 INR measurements by POCT coagulometer (CoaguChek XS Pro) and laboratory analyzer (Sysmex CS2000i) were compared using Passing-Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurement was further analyzed in relation to dosing decision. The correlation of INR measurements between CoaguChek XS Pro and Sysmex CS2000i was excellent (correlation coefficient = 0.973). The overall mean difference was 0.21 INR ± 0.32 (range: 1.7-0.44). The mean difference was found to get increased as INR results increased and was 0.09 in the subtherapeutic range (≤1.9 INR), 0.29 INR in the therapeutic range (2.0-3.0 INR), while 0.4 INR in the supratherapeutic range (>3.0 INR). The overall agreement was excellent (κ = 0.916) and overall 11 (5.5%) of 200 INR measurements showed a difference in dosing decision between the 2 instruments. The positive bias of POC-INR is evident in the supratherapeutic range which could affect the dosing decision requiring confirmation with the laboratory INR measurement. |
format | Online Article Text |
id | pubmed-6714735 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-67147352019-09-04 Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic Moiz, Bushra Rashid, Anila Hasan, Muhammad Jafri, Lena Raheem, Ahmed Clin Appl Thromb Hemost Original Articles Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients’ self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and investigated whether the results of POCT coagulometer can reliably be used without being confirmed by standard laboratory analyzer. A total of 200 INR measurements by POCT coagulometer (CoaguChek XS Pro) and laboratory analyzer (Sysmex CS2000i) were compared using Passing-Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurement was further analyzed in relation to dosing decision. The correlation of INR measurements between CoaguChek XS Pro and Sysmex CS2000i was excellent (correlation coefficient = 0.973). The overall mean difference was 0.21 INR ± 0.32 (range: 1.7-0.44). The mean difference was found to get increased as INR results increased and was 0.09 in the subtherapeutic range (≤1.9 INR), 0.29 INR in the therapeutic range (2.0-3.0 INR), while 0.4 INR in the supratherapeutic range (>3.0 INR). The overall agreement was excellent (κ = 0.916) and overall 11 (5.5%) of 200 INR measurements showed a difference in dosing decision between the 2 instruments. The positive bias of POC-INR is evident in the supratherapeutic range which could affect the dosing decision requiring confirmation with the laboratory INR measurement. SAGE Publications 2018-01-28 2018-10 /pmc/articles/PMC6714735/ /pubmed/29374969 http://dx.doi.org/10.1177/1076029617752247 Text en © The Author(s) 2018 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Articles Moiz, Bushra Rashid, Anila Hasan, Muhammad Jafri, Lena Raheem, Ahmed Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic |
title | Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic |
title_full | Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic |
title_fullStr | Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic |
title_full_unstemmed | Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic |
title_short | Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic |
title_sort | prospective comparison of point-of-care device and standard analyzer for monitoring of international normalized ratio in outpatient oral anticoagulant clinic |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6714735/ https://www.ncbi.nlm.nih.gov/pubmed/29374969 http://dx.doi.org/10.1177/1076029617752247 |
work_keys_str_mv | AT moizbushra prospectivecomparisonofpointofcaredeviceandstandardanalyzerformonitoringofinternationalnormalizedratioinoutpatientoralanticoagulantclinic AT rashidanila prospectivecomparisonofpointofcaredeviceandstandardanalyzerformonitoringofinternationalnormalizedratioinoutpatientoralanticoagulantclinic AT hasanmuhammad prospectivecomparisonofpointofcaredeviceandstandardanalyzerformonitoringofinternationalnormalizedratioinoutpatientoralanticoagulantclinic AT jafrilena prospectivecomparisonofpointofcaredeviceandstandardanalyzerformonitoringofinternationalnormalizedratioinoutpatientoralanticoagulantclinic AT raheemahmed prospectivecomparisonofpointofcaredeviceandstandardanalyzerformonitoringofinternationalnormalizedratioinoutpatientoralanticoagulantclinic |