Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study

BACKGROUND: Accelerometers, often in conjunction with heart rate sensors, are extensively used to track physical activity (PA) in research. Research-grade instruments are often expensive and have limited battery capacity, limited storage, and high participant burden. Consumer-based activity trackers...

Descripción completa

Detalles Bibliográficos
Autores principales: Henriksen, André, Grimsgaard, Sameline, Horsch, Alexander, Hartvigsen, Gunnar, Hopstock, Laila
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6716339/
https://www.ncbi.nlm.nih.gov/pubmed/31420958
http://dx.doi.org/10.2196/14438
_version_ 1783447359302139904
author Henriksen, André
Grimsgaard, Sameline
Horsch, Alexander
Hartvigsen, Gunnar
Hopstock, Laila
author_facet Henriksen, André
Grimsgaard, Sameline
Horsch, Alexander
Hartvigsen, Gunnar
Hopstock, Laila
author_sort Henriksen, André
collection PubMed
description BACKGROUND: Accelerometers, often in conjunction with heart rate sensors, are extensively used to track physical activity (PA) in research. Research-grade instruments are often expensive and have limited battery capacity, limited storage, and high participant burden. Consumer-based activity trackers are equipped with similar technology and designed for long-term wear, and can therefore potentially be used in research. OBJECTIVE: We aimed to assess the criterion validity of the Polar M430 sport watch, compared with 2 research-grade instruments (ActiGraph and Actiheart), worn on 4 different locations using 1- and 3-axis accelerometers. METHODS: A total of 50 participants wore 2 ActiGraphs (wrist and hip), 2 Actihearts (upper and lower chest position), and 1 Polar M430 sport watch for 1 full day. We compared reported time (minutes) spent in sedentary behavior and in light, moderate, vigorous, and moderate to vigorous PA, step counts, activity energy expenditure, and total energy expenditure between devices. We used Pearson correlations, intraclass correlations, mean absolute percentage errors (MAPEs), and Bland-Altman plots to assess criterion validity. RESULTS: Pearson correlations between the Polar M430 and all research-grade instruments were moderate or stronger for vigorous PA (r range .59-.76), moderate to vigorous PA (r range .51-.75), steps (r range .85-.87), total energy expenditure (r range .88-.94), and activity energy expenditure (r range .74-.79). Bland-Altman plots showed higher agreement for higher intensities of PA. MAPE was high for most outcomes. Only total energy expenditure measured by the hip-worn ActiGraph and both Actiheart positions had acceptable or close to acceptable errors with MAPEs of 6.94% (ActiGraph, 3 axes), 8.26% (ActiGraph, 1 axis), 14.54% (Actiheart, upper position), and 14.37% (Actiheart, lower position). The wrist-worn ActiGraph had a MAPE of 15.94% for measuring steps. All other outcomes had a MAPE of 22% or higher. For most outcomes, the Polar M430 was most strongly correlated with the hip-worn triaxial ActiGraph, with a moderate or strong Pearson correlation for sedentary behavior (r=.52) and for light (r=.7), moderate (r=.57), vigorous (r=.76), and moderate to vigorous (r=.75) PA. In addition, correlations were strong or very strong for activity energy expenditure (r=.75), steps (r=.85), and total energy expenditure (r=.91). CONCLUSIONS: The Polar M430 can potentially be used as an addition to established research-grade instruments to collect some PA variables over a prolonged period. However, due to the high MAPE of most outcomes, only total energy expenditure can be trusted to provide close to valid results. Depending on the variable, the Polar M430 over- or underreported most metrics, and may therefore be better suited to report changes in PA over time for some outcomes, rather than as an accurate instrument for PA status in a population.
format Online
Article
Text
id pubmed-6716339
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-67163392019-09-19 Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study Henriksen, André Grimsgaard, Sameline Horsch, Alexander Hartvigsen, Gunnar Hopstock, Laila JMIR Form Res Original Paper BACKGROUND: Accelerometers, often in conjunction with heart rate sensors, are extensively used to track physical activity (PA) in research. Research-grade instruments are often expensive and have limited battery capacity, limited storage, and high participant burden. Consumer-based activity trackers are equipped with similar technology and designed for long-term wear, and can therefore potentially be used in research. OBJECTIVE: We aimed to assess the criterion validity of the Polar M430 sport watch, compared with 2 research-grade instruments (ActiGraph and Actiheart), worn on 4 different locations using 1- and 3-axis accelerometers. METHODS: A total of 50 participants wore 2 ActiGraphs (wrist and hip), 2 Actihearts (upper and lower chest position), and 1 Polar M430 sport watch for 1 full day. We compared reported time (minutes) spent in sedentary behavior and in light, moderate, vigorous, and moderate to vigorous PA, step counts, activity energy expenditure, and total energy expenditure between devices. We used Pearson correlations, intraclass correlations, mean absolute percentage errors (MAPEs), and Bland-Altman plots to assess criterion validity. RESULTS: Pearson correlations between the Polar M430 and all research-grade instruments were moderate or stronger for vigorous PA (r range .59-.76), moderate to vigorous PA (r range .51-.75), steps (r range .85-.87), total energy expenditure (r range .88-.94), and activity energy expenditure (r range .74-.79). Bland-Altman plots showed higher agreement for higher intensities of PA. MAPE was high for most outcomes. Only total energy expenditure measured by the hip-worn ActiGraph and both Actiheart positions had acceptable or close to acceptable errors with MAPEs of 6.94% (ActiGraph, 3 axes), 8.26% (ActiGraph, 1 axis), 14.54% (Actiheart, upper position), and 14.37% (Actiheart, lower position). The wrist-worn ActiGraph had a MAPE of 15.94% for measuring steps. All other outcomes had a MAPE of 22% or higher. For most outcomes, the Polar M430 was most strongly correlated with the hip-worn triaxial ActiGraph, with a moderate or strong Pearson correlation for sedentary behavior (r=.52) and for light (r=.7), moderate (r=.57), vigorous (r=.76), and moderate to vigorous (r=.75) PA. In addition, correlations were strong or very strong for activity energy expenditure (r=.75), steps (r=.85), and total energy expenditure (r=.91). CONCLUSIONS: The Polar M430 can potentially be used as an addition to established research-grade instruments to collect some PA variables over a prolonged period. However, due to the high MAPE of most outcomes, only total energy expenditure can be trusted to provide close to valid results. Depending on the variable, the Polar M430 over- or underreported most metrics, and may therefore be better suited to report changes in PA over time for some outcomes, rather than as an accurate instrument for PA status in a population. JMIR Publications 2019-08-16 /pmc/articles/PMC6716339/ /pubmed/31420958 http://dx.doi.org/10.2196/14438 Text en ©André Henriksen, Sameline Grimsgaard, Alexander Horsch, Gunnar Hartvigsen, Laila Hopstock. Originally published in JMIR Formative Research (http://formative.jmir.org), 16.08.2019. https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Formative Research, is properly cited. The complete bibliographic information, a link to the original publication on http://formative.jmir.org, as well as this copyright and license information must be included.
spellingShingle Original Paper
Henriksen, André
Grimsgaard, Sameline
Horsch, Alexander
Hartvigsen, Gunnar
Hopstock, Laila
Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
title Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
title_full Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
title_fullStr Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
title_full_unstemmed Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
title_short Validity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
title_sort validity of the polar m430 activity monitor in free-living conditions: validation study
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6716339/
https://www.ncbi.nlm.nih.gov/pubmed/31420958
http://dx.doi.org/10.2196/14438
work_keys_str_mv AT henriksenandre validityofthepolarm430activitymonitorinfreelivingconditionsvalidationstudy
AT grimsgaardsameline validityofthepolarm430activitymonitorinfreelivingconditionsvalidationstudy
AT horschalexander validityofthepolarm430activitymonitorinfreelivingconditionsvalidationstudy
AT hartvigsengunnar validityofthepolarm430activitymonitorinfreelivingconditionsvalidationstudy
AT hopstocklaila validityofthepolarm430activitymonitorinfreelivingconditionsvalidationstudy