Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends

Spectral measurement techniques, such as the near-infrared (NIR) and Raman spectroscopy, have been intensively researched. Nevertheless, even today, these techniques are still sparsely applied in industry due to their unpredictable and unstable measurements. This paper put forward two data fusion st...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhu, Shichao, Song, Zhuoming, Shi, Shengyu, Wang, Mengmeng, Jin, Gang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6720423/
https://www.ncbi.nlm.nih.gov/pubmed/31398890
http://dx.doi.org/10.3390/s19163463
_version_ 1783448123156201472
author Zhu, Shichao
Song, Zhuoming
Shi, Shengyu
Wang, Mengmeng
Jin, Gang
author_facet Zhu, Shichao
Song, Zhuoming
Shi, Shengyu
Wang, Mengmeng
Jin, Gang
author_sort Zhu, Shichao
collection PubMed
description Spectral measurement techniques, such as the near-infrared (NIR) and Raman spectroscopy, have been intensively researched. Nevertheless, even today, these techniques are still sparsely applied in industry due to their unpredictable and unstable measurements. This paper put forward two data fusion strategies (low-level and mid-level fusion) for combining the NIR and Raman spectra to generate fusion spectra or fusion characteristics in order to improve the in-line measurement precision of component content of molten polymer blends. Subsequently, the fusion value was applied to modeling. For evaluating the response of different models to data fusion strategy, partial least squares (PLS) regression, artificial neural network (ANN), and extreme learning machine (ELM) were applied to the modeling of four kinds of spectral data (NIR, Raman, low-level fused data, and mid-level fused data). A system simultaneously acquiring in-line NIR and Raman spectra was built, and the polypropylene/polystyrene (PP/PS) blends, which had different grades and covered different compounding percentages of PP, were prepared for use as a case study. The results show that data fusion strategies improve the ANN and ELM model. In particular, mid-level fusion enables the in-line measurement of component content of molten polymer blends to become more accurate and robust.
format Online
Article
Text
id pubmed-6720423
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-67204232019-09-10 Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends Zhu, Shichao Song, Zhuoming Shi, Shengyu Wang, Mengmeng Jin, Gang Sensors (Basel) Article Spectral measurement techniques, such as the near-infrared (NIR) and Raman spectroscopy, have been intensively researched. Nevertheless, even today, these techniques are still sparsely applied in industry due to their unpredictable and unstable measurements. This paper put forward two data fusion strategies (low-level and mid-level fusion) for combining the NIR and Raman spectra to generate fusion spectra or fusion characteristics in order to improve the in-line measurement precision of component content of molten polymer blends. Subsequently, the fusion value was applied to modeling. For evaluating the response of different models to data fusion strategy, partial least squares (PLS) regression, artificial neural network (ANN), and extreme learning machine (ELM) were applied to the modeling of four kinds of spectral data (NIR, Raman, low-level fused data, and mid-level fused data). A system simultaneously acquiring in-line NIR and Raman spectra was built, and the polypropylene/polystyrene (PP/PS) blends, which had different grades and covered different compounding percentages of PP, were prepared for use as a case study. The results show that data fusion strategies improve the ANN and ELM model. In particular, mid-level fusion enables the in-line measurement of component content of molten polymer blends to become more accurate and robust. MDPI 2019-08-08 /pmc/articles/PMC6720423/ /pubmed/31398890 http://dx.doi.org/10.3390/s19163463 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Zhu, Shichao
Song, Zhuoming
Shi, Shengyu
Wang, Mengmeng
Jin, Gang
Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends
title Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends
title_full Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends
title_fullStr Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends
title_full_unstemmed Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends
title_short Fusion of Near-Infrared and Raman Spectroscopy for In-Line Measurement of Component Content of Molten Polymer Blends
title_sort fusion of near-infrared and raman spectroscopy for in-line measurement of component content of molten polymer blends
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6720423/
https://www.ncbi.nlm.nih.gov/pubmed/31398890
http://dx.doi.org/10.3390/s19163463
work_keys_str_mv AT zhushichao fusionofnearinfraredandramanspectroscopyforinlinemeasurementofcomponentcontentofmoltenpolymerblends
AT songzhuoming fusionofnearinfraredandramanspectroscopyforinlinemeasurementofcomponentcontentofmoltenpolymerblends
AT shishengyu fusionofnearinfraredandramanspectroscopyforinlinemeasurementofcomponentcontentofmoltenpolymerblends
AT wangmengmeng fusionofnearinfraredandramanspectroscopyforinlinemeasurementofcomponentcontentofmoltenpolymerblends
AT jingang fusionofnearinfraredandramanspectroscopyforinlinemeasurementofcomponentcontentofmoltenpolymerblends