Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability

BACKGROUND: Since inadequate heparin anticoagulation and insufficient reversal can result in complications during cardiopulmonary bypass (CPB) surgery, heparin anticoagulation monitoring by point-of-care (POC) activated clotting time (ACT) measurements is essential for CPB initiation, maintainance,...

Descripción completa

Detalles Bibliográficos
Autores principales: Dirkmann, Daniel, Nagy, Elisabeth, Britten, Martin W., Peters, Jürgen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6728977/
https://www.ncbi.nlm.nih.gov/pubmed/31492108
http://dx.doi.org/10.1186/s12871-019-0846-z
_version_ 1783449517019889664
author Dirkmann, Daniel
Nagy, Elisabeth
Britten, Martin W.
Peters, Jürgen
author_facet Dirkmann, Daniel
Nagy, Elisabeth
Britten, Martin W.
Peters, Jürgen
author_sort Dirkmann, Daniel
collection PubMed
description BACKGROUND: Since inadequate heparin anticoagulation and insufficient reversal can result in complications during cardiopulmonary bypass (CPB) surgery, heparin anticoagulation monitoring by point-of-care (POC) activated clotting time (ACT) measurements is essential for CPB initiation, maintainance, and anticoagulant reversal. However, concerns exist regarding reproducibility of ACT assays and comparability of devices. METHODS: We evaluated the agreement of ACT assays using four parallel measurements performed on two commonly used devices each (i.e., two Hemochron Signature Elite (Hemochron) and two Abbott i-STAT (i-STAT) devices, respectively). Blood samples from 30 patients undergoing cardiac surgery on CPB were assayed at specified steps (baseline, after heparin administration, after protamine administration) with four parallel measurements (two of each device type) using commercial Kaolin activated assays provided by the respective manufactures. Measurements were compared between identical and different device types using linear regression, Bland-Altman analyses, and calculation of Cohen’s kappa coefficient. RESULTS: Parallel i-STAT ACTs demonstrated a good linear correlation (r = 0.985). Bias, as determined by Bland-Altman analysis, was low (− 3.8 s; 95% limits of agreement (LOA): − 77.8 -70.2 s), and Cohen’s Kappa demonstrated good agreement (kappa = 0.809). Hemochron derived ACTs demonstrated worse linear correlation (r = 0.782), larger bias with considerably broader LOA (− 13.14 s; 95%LOA:-316.3–290 s), and lesser concordance between parallel assays (kappa = 0.554). Although demonstrating a fair linear correlation (r = 0.815), parallel measurements on different ACT-devices showed large bias (−20s; 95% LOA: − 290-250 s) and little concordance (kappa = 0.368). Overall, disconcordant results according to clinically predefined target values were more frequent with the Hemochron than i-STAT. Furthermore, while discrepancies in ACT between two parallel iSTAT assays showed little or no clinical relevance, deviations from parallel Hemochron assays and iSTAT versus Hemochron measurements revealed marked and sometimes clinically critical deviations. CONCLUSION: Currently used ACT point-of-care devices cannot be used interchangeably. Furthermore, our data question the reliability of the Hemochron in assessing adequacy of heparin anticoagulation monitoring for CPB.
format Online
Article
Text
id pubmed-6728977
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-67289772019-09-12 Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability Dirkmann, Daniel Nagy, Elisabeth Britten, Martin W. Peters, Jürgen BMC Anesthesiol Research Article BACKGROUND: Since inadequate heparin anticoagulation and insufficient reversal can result in complications during cardiopulmonary bypass (CPB) surgery, heparin anticoagulation monitoring by point-of-care (POC) activated clotting time (ACT) measurements is essential for CPB initiation, maintainance, and anticoagulant reversal. However, concerns exist regarding reproducibility of ACT assays and comparability of devices. METHODS: We evaluated the agreement of ACT assays using four parallel measurements performed on two commonly used devices each (i.e., two Hemochron Signature Elite (Hemochron) and two Abbott i-STAT (i-STAT) devices, respectively). Blood samples from 30 patients undergoing cardiac surgery on CPB were assayed at specified steps (baseline, after heparin administration, after protamine administration) with four parallel measurements (two of each device type) using commercial Kaolin activated assays provided by the respective manufactures. Measurements were compared between identical and different device types using linear regression, Bland-Altman analyses, and calculation of Cohen’s kappa coefficient. RESULTS: Parallel i-STAT ACTs demonstrated a good linear correlation (r = 0.985). Bias, as determined by Bland-Altman analysis, was low (− 3.8 s; 95% limits of agreement (LOA): − 77.8 -70.2 s), and Cohen’s Kappa demonstrated good agreement (kappa = 0.809). Hemochron derived ACTs demonstrated worse linear correlation (r = 0.782), larger bias with considerably broader LOA (− 13.14 s; 95%LOA:-316.3–290 s), and lesser concordance between parallel assays (kappa = 0.554). Although demonstrating a fair linear correlation (r = 0.815), parallel measurements on different ACT-devices showed large bias (−20s; 95% LOA: − 290-250 s) and little concordance (kappa = 0.368). Overall, disconcordant results according to clinically predefined target values were more frequent with the Hemochron than i-STAT. Furthermore, while discrepancies in ACT between two parallel iSTAT assays showed little or no clinical relevance, deviations from parallel Hemochron assays and iSTAT versus Hemochron measurements revealed marked and sometimes clinically critical deviations. CONCLUSION: Currently used ACT point-of-care devices cannot be used interchangeably. Furthermore, our data question the reliability of the Hemochron in assessing adequacy of heparin anticoagulation monitoring for CPB. BioMed Central 2019-09-06 /pmc/articles/PMC6728977/ /pubmed/31492108 http://dx.doi.org/10.1186/s12871-019-0846-z Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Dirkmann, Daniel
Nagy, Elisabeth
Britten, Martin W.
Peters, Jürgen
Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
title Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
title_full Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
title_fullStr Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
title_full_unstemmed Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
title_short Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
title_sort point-of-care measurement of activated clotting time for cardiac surgery as measured by the hemochron signature elite and the abbott i-stat: agreement, concordance, and clinical reliability
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6728977/
https://www.ncbi.nlm.nih.gov/pubmed/31492108
http://dx.doi.org/10.1186/s12871-019-0846-z
work_keys_str_mv AT dirkmanndaniel pointofcaremeasurementofactivatedclottingtimeforcardiacsurgeryasmeasuredbythehemochronsignatureeliteandtheabbottistatagreementconcordanceandclinicalreliability
AT nagyelisabeth pointofcaremeasurementofactivatedclottingtimeforcardiacsurgeryasmeasuredbythehemochronsignatureeliteandtheabbottistatagreementconcordanceandclinicalreliability
AT brittenmartinw pointofcaremeasurementofactivatedclottingtimeforcardiacsurgeryasmeasuredbythehemochronsignatureeliteandtheabbottistatagreementconcordanceandclinicalreliability
AT petersjurgen pointofcaremeasurementofactivatedclottingtimeforcardiacsurgeryasmeasuredbythehemochronsignatureeliteandtheabbottistatagreementconcordanceandclinicalreliability