Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography

OBJECTIVES: The frequency of abdominal computed tomography examinations is increasing, leading to a significant level of patient dose. This study aims to quantify and evaluate the effects of automatic tube current modulation (ATCM) technique on patient dose and image quality in contrast-enhanced bip...

Descripción completa

Detalles Bibliográficos
Autores principales: Yurt, Ayşegül, Özsoykal, İsmail, Obuz, Funda
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Galenos Publishing 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6746012/
https://www.ncbi.nlm.nih.gov/pubmed/31507141
http://dx.doi.org/10.4274/mirt.galenos.2019.83723
_version_ 1783451639185670144
author Yurt, Ayşegül
Özsoykal, İsmail
Obuz, Funda
author_facet Yurt, Ayşegül
Özsoykal, İsmail
Obuz, Funda
author_sort Yurt, Ayşegül
collection PubMed
description OBJECTIVES: The frequency of abdominal computed tomography examinations is increasing, leading to a significant level of patient dose. This study aims to quantify and evaluate the effects of automatic tube current modulation (ATCM) technique on patient dose and image quality in contrast-enhanced biphasic abdominal examinations. METHODS: Two different scan protocols, based on constant tube current and ATCM technique, were used on 64 patients who visited our radiology department periodically. For three patient groups with different patient size, results from two protocols were compared with respect to patient dose and image quality. Dosimetric evaluations were based on the Computed Tomography Dose Index, dose length product, and effective dose. For the comparison of image qualities between two protocols, Noise Index (NI) and Contrast to Noise Ratio (CNR) values were determined for each image. Additionally, the quality of each image was evaluated subjectively by an experienced radiologist, and the results were compared between the two protocols. RESULTS: Dose reductions of 31% and 21% were achieved by the ATCM protocol in the arterial and portal phases, respectively. On the other hand, NI exhibited an increase between 9% and 46% for liver, fat and aorta. CNR values were observed to decrease between 5% and 19%. All images were evaluated by a radiologist, and no obstacle limiting a reliable diagnostic evaluation was found in any image obtained by either technique. CONCLUSION: These results showed that the ATCM technique reduces patient dose significantly while maintaining a certain level of image quality.
format Online
Article
Text
id pubmed-6746012
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Galenos Publishing
record_format MEDLINE/PubMed
spelling pubmed-67460122019-10-01 Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography Yurt, Ayşegül Özsoykal, İsmail Obuz, Funda Mol Imaging Radionucl Ther Original Article OBJECTIVES: The frequency of abdominal computed tomography examinations is increasing, leading to a significant level of patient dose. This study aims to quantify and evaluate the effects of automatic tube current modulation (ATCM) technique on patient dose and image quality in contrast-enhanced biphasic abdominal examinations. METHODS: Two different scan protocols, based on constant tube current and ATCM technique, were used on 64 patients who visited our radiology department periodically. For three patient groups with different patient size, results from two protocols were compared with respect to patient dose and image quality. Dosimetric evaluations were based on the Computed Tomography Dose Index, dose length product, and effective dose. For the comparison of image qualities between two protocols, Noise Index (NI) and Contrast to Noise Ratio (CNR) values were determined for each image. Additionally, the quality of each image was evaluated subjectively by an experienced radiologist, and the results were compared between the two protocols. RESULTS: Dose reductions of 31% and 21% were achieved by the ATCM protocol in the arterial and portal phases, respectively. On the other hand, NI exhibited an increase between 9% and 46% for liver, fat and aorta. CNR values were observed to decrease between 5% and 19%. All images were evaluated by a radiologist, and no obstacle limiting a reliable diagnostic evaluation was found in any image obtained by either technique. CONCLUSION: These results showed that the ATCM technique reduces patient dose significantly while maintaining a certain level of image quality. Galenos Publishing 2019-10 2019-09-06 /pmc/articles/PMC6746012/ /pubmed/31507141 http://dx.doi.org/10.4274/mirt.galenos.2019.83723 Text en ©Copyright 2019 by Turkish Society of Nuclear Medicine | Molecular Imaging and Radionuclide Therapy published by Galenos Yayınevi. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Yurt, Ayşegül
Özsoykal, İsmail
Obuz, Funda
Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
title Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
title_full Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
title_fullStr Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
title_full_unstemmed Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
title_short Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography
title_sort effects of the use of automatic tube current modulation on patient dose and image quality in computed tomography
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6746012/
https://www.ncbi.nlm.nih.gov/pubmed/31507141
http://dx.doi.org/10.4274/mirt.galenos.2019.83723
work_keys_str_mv AT yurtaysegul effectsoftheuseofautomatictubecurrentmodulationonpatientdoseandimagequalityincomputedtomography
AT ozsoykalismail effectsoftheuseofautomatictubecurrentmodulationonpatientdoseandimagequalityincomputedtomography
AT obuzfunda effectsoftheuseofautomatictubecurrentmodulationonpatientdoseandimagequalityincomputedtomography