Cargando…

Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve

Background and purpose — Little is known about the role of retained trunnions in revision hip arthroplasties, i.e., when only the femoral head is substituted. Wear (fretting corrosion) and ceramic head fractures are 2 poorly understood concerns related to use, and the role of adapter sleeves has not...

Descripción completa

Detalles Bibliográficos
Autores principales: Affatato, Saverio, Cosentino, Monica, Castagnini, Francesco, Bordini, Barbara
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Taylor & Francis 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6746291/
https://www.ncbi.nlm.nih.gov/pubmed/31210073
http://dx.doi.org/10.1080/17453674.2019.1618649
_version_ 1783451684981178368
author Affatato, Saverio
Cosentino, Monica
Castagnini, Francesco
Bordini, Barbara
author_facet Affatato, Saverio
Cosentino, Monica
Castagnini, Francesco
Bordini, Barbara
author_sort Affatato, Saverio
collection PubMed
description Background and purpose — Little is known about the role of retained trunnions in revision hip arthroplasties, i.e., when only the femoral head is substituted. Wear (fretting corrosion) and ceramic head fractures are 2 poorly understood concerns related to use, and the role of adapter sleeves has not been defined. In this registry study we assessed the influence of sleeve interposition on re-revision rates in revision hip arthroplasties with retained stems. Confounding factors (demographics, implant-related features) and failures were also analyzed. Patients and methods — We conducted a registry study on 1,127 revised implants (retained trunnion and head exchange). In 26% of implants an adapter sleeve was interposed; in 74% no adapter sleeve was implanted. Demographic and implant-related features were investigated including a descriptive analysis of failures. Results — The mean follow-up of revised implants with and without the use of an adapter sleeve was 3.3 and 5.1 years, respectively. The implant survival without an adapter sleeve was significantly higher, 98.4% (95% CI 96.9–99.8) vs. 95.2% (CI 93.2–96.6) with an adapter sleeve at 5 years. No re-revisions due to adverse local tissue reactions or ceramic head fractures were reported. In order to overcome the different distribution of head materials and head sizes in the two cohorts, only Delta balls were investigated. Interpretation — Adapter sleeve interposition had a minor influence on the revision rates. No adverse local tissue reactions or head fractures occurred.
format Online
Article
Text
id pubmed-6746291
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Taylor & Francis
record_format MEDLINE/PubMed
spelling pubmed-67462912019-10-01 Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve Affatato, Saverio Cosentino, Monica Castagnini, Francesco Bordini, Barbara Acta Orthop Article Background and purpose — Little is known about the role of retained trunnions in revision hip arthroplasties, i.e., when only the femoral head is substituted. Wear (fretting corrosion) and ceramic head fractures are 2 poorly understood concerns related to use, and the role of adapter sleeves has not been defined. In this registry study we assessed the influence of sleeve interposition on re-revision rates in revision hip arthroplasties with retained stems. Confounding factors (demographics, implant-related features) and failures were also analyzed. Patients and methods — We conducted a registry study on 1,127 revised implants (retained trunnion and head exchange). In 26% of implants an adapter sleeve was interposed; in 74% no adapter sleeve was implanted. Demographic and implant-related features were investigated including a descriptive analysis of failures. Results — The mean follow-up of revised implants with and without the use of an adapter sleeve was 3.3 and 5.1 years, respectively. The implant survival without an adapter sleeve was significantly higher, 98.4% (95% CI 96.9–99.8) vs. 95.2% (CI 93.2–96.6) with an adapter sleeve at 5 years. No re-revisions due to adverse local tissue reactions or ceramic head fractures were reported. In order to overcome the different distribution of head materials and head sizes in the two cohorts, only Delta balls were investigated. Interpretation — Adapter sleeve interposition had a minor influence on the revision rates. No adverse local tissue reactions or head fractures occurred. Taylor & Francis 2019-10 2019-06-18 /pmc/articles/PMC6746291/ /pubmed/31210073 http://dx.doi.org/10.1080/17453674.2019.1618649 Text en © 2019 The Author(s). Published by Taylor & Francis on behalf of the Nordic Orthopedic Federation http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Article
Affatato, Saverio
Cosentino, Monica
Castagnini, Francesco
Bordini, Barbara
Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
title Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
title_full Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
title_fullStr Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
title_full_unstemmed Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
title_short Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
title_sort registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6746291/
https://www.ncbi.nlm.nih.gov/pubmed/31210073
http://dx.doi.org/10.1080/17453674.2019.1618649
work_keys_str_mv AT affatatosaverio registrystudyonfailureincidencein1127revisedhipimplantswithstemtrunnionreuseafter10yearsoffollowuplimitedinfluenceofanadaptersleeve
AT cosentinomonica registrystudyonfailureincidencein1127revisedhipimplantswithstemtrunnionreuseafter10yearsoffollowuplimitedinfluenceofanadaptersleeve
AT castagninifrancesco registrystudyonfailureincidencein1127revisedhipimplantswithstemtrunnionreuseafter10yearsoffollowuplimitedinfluenceofanadaptersleeve
AT bordinibarbara registrystudyonfailureincidencein1127revisedhipimplantswithstemtrunnionreuseafter10yearsoffollowuplimitedinfluenceofanadaptersleeve