Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users

The aim of this study was to model the absorption in the head of an electromagnetic field (EMF) emitted by a radiofrequency identification reader operating at a frequency of 13.56 MHz (recognized as an RFID HF reader), with respect to the direct biophysical effects evaluated by the specific absorpti...

Descripción completa

Detalles Bibliográficos
Autores principales: Zradziński, Patryk, Karpowicz, Jolanta, Gryz, Krzysztof
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6749309/
https://www.ncbi.nlm.nih.gov/pubmed/31466315
http://dx.doi.org/10.3390/s19173724
_version_ 1783452248092704768
author Zradziński, Patryk
Karpowicz, Jolanta
Gryz, Krzysztof
author_facet Zradziński, Patryk
Karpowicz, Jolanta
Gryz, Krzysztof
author_sort Zradziński, Patryk
collection PubMed
description The aim of this study was to model the absorption in the head of an electromagnetic field (EMF) emitted by a radiofrequency identification reader operating at a frequency of 13.56 MHz (recognized as an RFID HF reader), with respect to the direct biophysical effects evaluated by the specific absorption rate (SAR), averaged over the entire head or locally, over any 10 g of tissues. The exposure effects were compared between the head of a user of a hearing implant with an acoustic sensor and a person without such an implant, used as a referenced case. The RFID HF reader, such as is used in shops or libraries, was modeled as a loop antenna (35 × 35 cm). SAR was calculated in a multi-layer ellipsoidal model of the head—with or without models of hearing implants of two types: Bonebridge (MED-EL, Austria) or bone anchored hearing aid attract (BAHA) (Cochlear, Sweden). Relative SAR values were calculated as the ratio between the SAR in the head of the implant user and the non-user. It was found that the use of BAHA hearing implants increased the effects of 13.56 MHz EMF exposure in the head in comparison to non-user—up to 2.1 times higher localized SAR in the worst case exposure scenario, and it is statistically significant higher than when Bonebridge implants are used (Kruscal–Wallis test with Bonferroni correction, p < 0.017). The evaluation of EMF exposure from an RFID reader with respect to limits established for the implant non-user population may be insufficient to protect an implant user when exposure approaches these limits, but the significant difference between exposure effects in users of various types of implants need to be considered.
format Online
Article
Text
id pubmed-6749309
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-67493092019-09-27 Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users Zradziński, Patryk Karpowicz, Jolanta Gryz, Krzysztof Sensors (Basel) Article The aim of this study was to model the absorption in the head of an electromagnetic field (EMF) emitted by a radiofrequency identification reader operating at a frequency of 13.56 MHz (recognized as an RFID HF reader), with respect to the direct biophysical effects evaluated by the specific absorption rate (SAR), averaged over the entire head or locally, over any 10 g of tissues. The exposure effects were compared between the head of a user of a hearing implant with an acoustic sensor and a person without such an implant, used as a referenced case. The RFID HF reader, such as is used in shops or libraries, was modeled as a loop antenna (35 × 35 cm). SAR was calculated in a multi-layer ellipsoidal model of the head—with or without models of hearing implants of two types: Bonebridge (MED-EL, Austria) or bone anchored hearing aid attract (BAHA) (Cochlear, Sweden). Relative SAR values were calculated as the ratio between the SAR in the head of the implant user and the non-user. It was found that the use of BAHA hearing implants increased the effects of 13.56 MHz EMF exposure in the head in comparison to non-user—up to 2.1 times higher localized SAR in the worst case exposure scenario, and it is statistically significant higher than when Bonebridge implants are used (Kruscal–Wallis test with Bonferroni correction, p < 0.017). The evaluation of EMF exposure from an RFID reader with respect to limits established for the implant non-user population may be insufficient to protect an implant user when exposure approaches these limits, but the significant difference between exposure effects in users of various types of implants need to be considered. MDPI 2019-08-28 /pmc/articles/PMC6749309/ /pubmed/31466315 http://dx.doi.org/10.3390/s19173724 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Zradziński, Patryk
Karpowicz, Jolanta
Gryz, Krzysztof
Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users
title Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users
title_full Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users
title_fullStr Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users
title_full_unstemmed Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users
title_short Electromagnetic Energy Absorption in a Head Approaching a Radiofrequency Identification (RFID) Reader Operating at 13.56 MHz in Users of Hearing Implants Versus Non-Users
title_sort electromagnetic energy absorption in a head approaching a radiofrequency identification (rfid) reader operating at 13.56 mhz in users of hearing implants versus non-users
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6749309/
https://www.ncbi.nlm.nih.gov/pubmed/31466315
http://dx.doi.org/10.3390/s19173724
work_keys_str_mv AT zradzinskipatryk electromagneticenergyabsorptioninaheadapproachingaradiofrequencyidentificationrfidreaderoperatingat1356mhzinusersofhearingimplantsversusnonusers
AT karpowiczjolanta electromagneticenergyabsorptioninaheadapproachingaradiofrequencyidentificationrfidreaderoperatingat1356mhzinusersofhearingimplantsversusnonusers
AT gryzkrzysztof electromagneticenergyabsorptioninaheadapproachingaradiofrequencyidentificationrfidreaderoperatingat1356mhzinusersofhearingimplantsversusnonusers