High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector

Here, the development and testing of a portable double peak pulse laser induced plasma spectroscopy (DPP-LIPS) based on passively Q-switched Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) laser excitation is reported. The latter delivered structured laser pulses at a repetition rate of up to 20 Hz...

Descripción completa

Detalles Bibliográficos
Autores principales: Agresti, Juri, Mencaglia, Andrea Azelio, Siano, Salvatore
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6749410/
https://www.ncbi.nlm.nih.gov/pubmed/31438514
http://dx.doi.org/10.3390/s19173634
_version_ 1783452272253992960
author Agresti, Juri
Mencaglia, Andrea Azelio
Siano, Salvatore
author_facet Agresti, Juri
Mencaglia, Andrea Azelio
Siano, Salvatore
author_sort Agresti, Juri
collection PubMed
description Here, the development and testing of a portable double peak pulse laser induced plasma spectroscopy (DPP-LIPS) based on passively Q-switched Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) laser excitation is reported. The latter delivered structured laser pulses at a repetition rate of up to 20 Hz, including two energy peaks of about 100 mJ each with a relative temporal spacing of about 80 µs. Plasma spectra were collected using a low-cost Czerny–Turner spectrometer equipped with a non-intensified CCD (Charge-Coupled Device) array. Such a DPP-LIPS setup is technologically simpler and cheaper than the usual ones. Despite the relatively large temporal separation between the mentioned laser peaks, significant spectral intensity enhancements with respect to the usual single peak pulse configuration were observed. The amplification factor measured ranged between 2 and 10, depending on the specific emission peaks and the Q-switched configuration, and a consequent significant improvement of the detection limit of trace elements was observed. The instrument was calibrated for the quantitative analysis of copper alloy through systematic measurements carried out on reference samples and was then tested in an example archaeometric characterization of a statuette from the Egyptian Museum of Florence.
format Online
Article
Text
id pubmed-6749410
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-67494102019-09-27 High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector Agresti, Juri Mencaglia, Andrea Azelio Siano, Salvatore Sensors (Basel) Article Here, the development and testing of a portable double peak pulse laser induced plasma spectroscopy (DPP-LIPS) based on passively Q-switched Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) laser excitation is reported. The latter delivered structured laser pulses at a repetition rate of up to 20 Hz, including two energy peaks of about 100 mJ each with a relative temporal spacing of about 80 µs. Plasma spectra were collected using a low-cost Czerny–Turner spectrometer equipped with a non-intensified CCD (Charge-Coupled Device) array. Such a DPP-LIPS setup is technologically simpler and cheaper than the usual ones. Despite the relatively large temporal separation between the mentioned laser peaks, significant spectral intensity enhancements with respect to the usual single peak pulse configuration were observed. The amplification factor measured ranged between 2 and 10, depending on the specific emission peaks and the Q-switched configuration, and a consequent significant improvement of the detection limit of trace elements was observed. The instrument was calibrated for the quantitative analysis of copper alloy through systematic measurements carried out on reference samples and was then tested in an example archaeometric characterization of a statuette from the Egyptian Museum of Florence. MDPI 2019-08-21 /pmc/articles/PMC6749410/ /pubmed/31438514 http://dx.doi.org/10.3390/s19173634 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Agresti, Juri
Mencaglia, Andrea Azelio
Siano, Salvatore
High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector
title High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector
title_full High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector
title_fullStr High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector
title_full_unstemmed High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector
title_short High Energy Double Peak Pulse Laser Induced Plasma Spectroscopy for Metal Characterization Using a Passively Q-Switched Laser Source and CCD Detector
title_sort high energy double peak pulse laser induced plasma spectroscopy for metal characterization using a passively q-switched laser source and ccd detector
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6749410/
https://www.ncbi.nlm.nih.gov/pubmed/31438514
http://dx.doi.org/10.3390/s19173634
work_keys_str_mv AT agrestijuri highenergydoublepeakpulselaserinducedplasmaspectroscopyformetalcharacterizationusingapassivelyqswitchedlasersourceandccddetector
AT mencagliaandreaazelio highenergydoublepeakpulselaserinducedplasmaspectroscopyformetalcharacterizationusingapassivelyqswitchedlasersourceandccddetector
AT sianosalvatore highenergydoublepeakpulselaserinducedplasmaspectroscopyformetalcharacterizationusingapassivelyqswitchedlasersourceandccddetector