A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS

In this paper, a dual-biomarker-based neural sensing and conditioning device is proposed for closing the feedback loop in deep brain stimulation devices. The device explores both local field potentials (LFPs) and action potentials (APs) as measured biomarkers. It includes two channels, each having f...

Descripción completa

Detalles Bibliográficos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: IEEE 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6752632/
https://www.ncbi.nlm.nih.gov/pubmed/31667027
http://dx.doi.org/10.1109/JTEHM.2019.2937776
_version_ 1783452766540136448
collection PubMed
description In this paper, a dual-biomarker-based neural sensing and conditioning device is proposed for closing the feedback loop in deep brain stimulation devices. The device explores both local field potentials (LFPs) and action potentials (APs) as measured biomarkers. It includes two channels, each having four main parts: (1) a pre-amplifier with built-in low-pass filter, (2) a ground shifting circuit, (3) an amplifier with low-pass function, and (4) a high-pass filter. The design specifications include miniature-size, light-weight, and 100 dB gain in the LFP and AP channels. This device has been validated through bench and in-vitro tests. The bench tests have been performed using different sinusoidal signals and pre-recorded neural signals. The in-vitro tests have been conducted in the saline solution that mimics the brain environment. The total weight of the device including a 3 V coin battery, and battery holder is 1.2 g. The diameter of the device is 11.2 mm. The device can be used to concurrently sense LFPs and APs for closing the feedback loop in closed-loop deep brain stimulation systems. It provides a tetherless head-mountable platform suitable for pre-clinical trials.
format Online
Article
Text
id pubmed-6752632
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher IEEE
record_format MEDLINE/PubMed
spelling pubmed-67526322019-10-30 A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS IEEE J Transl Eng Health Med Article In this paper, a dual-biomarker-based neural sensing and conditioning device is proposed for closing the feedback loop in deep brain stimulation devices. The device explores both local field potentials (LFPs) and action potentials (APs) as measured biomarkers. It includes two channels, each having four main parts: (1) a pre-amplifier with built-in low-pass filter, (2) a ground shifting circuit, (3) an amplifier with low-pass function, and (4) a high-pass filter. The design specifications include miniature-size, light-weight, and 100 dB gain in the LFP and AP channels. This device has been validated through bench and in-vitro tests. The bench tests have been performed using different sinusoidal signals and pre-recorded neural signals. The in-vitro tests have been conducted in the saline solution that mimics the brain environment. The total weight of the device including a 3 V coin battery, and battery holder is 1.2 g. The diameter of the device is 11.2 mm. The device can be used to concurrently sense LFPs and APs for closing the feedback loop in closed-loop deep brain stimulation systems. It provides a tetherless head-mountable platform suitable for pre-clinical trials. IEEE 2019-08-30 /pmc/articles/PMC6752632/ /pubmed/31667027 http://dx.doi.org/10.1109/JTEHM.2019.2937776 Text en This work is licensed under a Creative Commons Attribution 4.0 License. For more information, see http://creativecommons.org/licenses/by/4.0/.
spellingShingle Article
A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS
title A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS
title_full A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS
title_fullStr A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS
title_full_unstemmed A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS
title_short A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS
title_sort miniature dual-biomarker-based sensing and conditioning device for closed-loop dbs
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6752632/
https://www.ncbi.nlm.nih.gov/pubmed/31667027
http://dx.doi.org/10.1109/JTEHM.2019.2937776
work_keys_str_mv AT aminiaturedualbiomarkerbasedsensingandconditioningdeviceforclosedloopdbs
AT aminiaturedualbiomarkerbasedsensingandconditioningdeviceforclosedloopdbs
AT miniaturedualbiomarkerbasedsensingandconditioningdeviceforclosedloopdbs
AT miniaturedualbiomarkerbasedsensingandconditioningdeviceforclosedloopdbs