Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study

BACKGROUND: We evaluated a magnetic resonance (MR)-conditional high-power microwave ablation system. METHODS: An ex vivo 1.5-T evaluation was conducted by varying the sequence (T1-weighted volume interpolated breath-hold examination, T1w-VIBE; T1-weighted fast low-angle shot, T1w-FLASH; T2-weighted...

Descripción completa

Detalles Bibliográficos
Autores principales: Grimm, Antonia, Winkelmann, Moritz, Weiß, Jakob, Gohla, Georg, Blumenstock, Gunnar, Nikolaou, Konstantin, Clasen, Stephan, Hoffmann, Rüdiger
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6757083/
https://www.ncbi.nlm.nih.gov/pubmed/31549338
http://dx.doi.org/10.1186/s41747-019-0115-4
_version_ 1783453512502345728
author Grimm, Antonia
Winkelmann, Moritz
Weiß, Jakob
Gohla, Georg
Blumenstock, Gunnar
Nikolaou, Konstantin
Clasen, Stephan
Hoffmann, Rüdiger
author_facet Grimm, Antonia
Winkelmann, Moritz
Weiß, Jakob
Gohla, Georg
Blumenstock, Gunnar
Nikolaou, Konstantin
Clasen, Stephan
Hoffmann, Rüdiger
author_sort Grimm, Antonia
collection PubMed
description BACKGROUND: We evaluated a magnetic resonance (MR)-conditional high-power microwave ablation system. METHODS: An ex vivo 1.5-T evaluation was conducted by varying the sequence (T1-weighted volume interpolated breath-hold examination, T1w-VIBE; T1-weighted fast low-angle shot, T1w-FLASH; T2-weighted turbo spin-echo, T2w-TSE), applicator angulation to B(0) (A-to-B(0)), slice orientation, and encoding direction. Tip location error (TLE) and artefact diameters were measured, and influence of imaging parameters was assessed with analysis of variance and post hoc testing. Twenty-four ex vivo ablations were conducted in three bovine livers at 80 W and 120 W. Ablation durations were 5, 10, and 15 min. Ablation zones were compared for short-axis diameter (SAD), volume, and sphericity index (SI) with unpaired t test. RESULTS: The artefact pattern was similar for all sequences. The shaft artefact (4.4 ± 2.9 mm, mean ± standard deviation) was dependent on the sequence (p = 0.012) and the A-to-B(0) (p < 0.001); the largest shaft diameter was measured with T1w-FLASH (6.3 ± 3.4 mm) and with perpendicular A-to-B(0) (6.7 ± 2.4 mm). The tip artefact (1.6 ± 0.7 mm) was dependent on A-to-B(0) (p = 0.001); TLE was -2.6 ± 1.0 mm. Ablation results at the maximum setting (15 min, 120 W) were SAD = 42.0 ± 1.41 mm; volume = 56.78 ± 3.08 cm(3), SI = 0.68 ± 0.05. In all ablations, SI ranged 0.68–0.75 with the smallest SI at 15 min and 120 W (p = 0.048). CONCLUSION: The system produced sufficiently large ablation zones and the artefact was appropriate for MR-guided interventions.
format Online
Article
Text
id pubmed-6757083
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-67570832019-10-07 Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study Grimm, Antonia Winkelmann, Moritz Weiß, Jakob Gohla, Georg Blumenstock, Gunnar Nikolaou, Konstantin Clasen, Stephan Hoffmann, Rüdiger Eur Radiol Exp Original Article BACKGROUND: We evaluated a magnetic resonance (MR)-conditional high-power microwave ablation system. METHODS: An ex vivo 1.5-T evaluation was conducted by varying the sequence (T1-weighted volume interpolated breath-hold examination, T1w-VIBE; T1-weighted fast low-angle shot, T1w-FLASH; T2-weighted turbo spin-echo, T2w-TSE), applicator angulation to B(0) (A-to-B(0)), slice orientation, and encoding direction. Tip location error (TLE) and artefact diameters were measured, and influence of imaging parameters was assessed with analysis of variance and post hoc testing. Twenty-four ex vivo ablations were conducted in three bovine livers at 80 W and 120 W. Ablation durations were 5, 10, and 15 min. Ablation zones were compared for short-axis diameter (SAD), volume, and sphericity index (SI) with unpaired t test. RESULTS: The artefact pattern was similar for all sequences. The shaft artefact (4.4 ± 2.9 mm, mean ± standard deviation) was dependent on the sequence (p = 0.012) and the A-to-B(0) (p < 0.001); the largest shaft diameter was measured with T1w-FLASH (6.3 ± 3.4 mm) and with perpendicular A-to-B(0) (6.7 ± 2.4 mm). The tip artefact (1.6 ± 0.7 mm) was dependent on A-to-B(0) (p = 0.001); TLE was -2.6 ± 1.0 mm. Ablation results at the maximum setting (15 min, 120 W) were SAD = 42.0 ± 1.41 mm; volume = 56.78 ± 3.08 cm(3), SI = 0.68 ± 0.05. In all ablations, SI ranged 0.68–0.75 with the smallest SI at 15 min and 120 W (p = 0.048). CONCLUSION: The system produced sufficiently large ablation zones and the artefact was appropriate for MR-guided interventions. Springer International Publishing 2019-09-23 /pmc/articles/PMC6757083/ /pubmed/31549338 http://dx.doi.org/10.1186/s41747-019-0115-4 Text en © The Author(s) 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Article
Grimm, Antonia
Winkelmann, Moritz
Weiß, Jakob
Gohla, Georg
Blumenstock, Gunnar
Nikolaou, Konstantin
Clasen, Stephan
Hoffmann, Rüdiger
Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study
title Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study
title_full Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study
title_fullStr Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study
title_full_unstemmed Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study
title_short Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an exvivo study
title_sort artefact and ablation performance of an mr-conditional high-power microwave system in bovine livers: an exvivo study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6757083/
https://www.ncbi.nlm.nih.gov/pubmed/31549338
http://dx.doi.org/10.1186/s41747-019-0115-4
work_keys_str_mv AT grimmantonia artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT winkelmannmoritz artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT weißjakob artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT gohlageorg artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT blumenstockgunnar artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT nikolaoukonstantin artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT clasenstephan artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy
AT hoffmannrudiger artefactandablationperformanceofanmrconditionalhighpowermicrowavesysteminbovineliversanexvivostudy