Cargando…
A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk
We present an integral smartphone-based whole-cell biosensor, LumiCellSense (LCS), which incorporates a 16-well biochip with an oxygen permeable coating, harboring bioluminescent Escherichia coli bioreporter cells, a macro lens, a lens barrel, a metal heater tray, and a temperature controller, enclo...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6767005/ https://www.ncbi.nlm.nih.gov/pubmed/31505815 http://dx.doi.org/10.3390/s19183882 |
_version_ | 1783454816779894784 |
---|---|
author | Lu, Mei-Yi Kao, Wei-Chen Belkin, Shimshon Cheng, Ji-Yen |
author_facet | Lu, Mei-Yi Kao, Wei-Chen Belkin, Shimshon Cheng, Ji-Yen |
author_sort | Lu, Mei-Yi |
collection | PubMed |
description | We present an integral smartphone-based whole-cell biosensor, LumiCellSense (LCS), which incorporates a 16-well biochip with an oxygen permeable coating, harboring bioluminescent Escherichia coli bioreporter cells, a macro lens, a lens barrel, a metal heater tray, and a temperature controller, enclosed in a light-impermeable case. The luminescence emitted by the bioreporter cells in response to the presence of the target chemicals is imaged by the phone’s camera, and a dedicated phone-embedded application, LCS_Logger, is employed to calculate photon emission intensity and plot it in real time on the device’s screen. An alert is automatically given when light intensity increases above the baseline, indicating the presence of the target. We demonstrate the efficacy of this system by the detection of residues of an antibiotic, ciprofloxacin (CIP), in whole milk, with a detection threshold of 7.2 ng/mL. This value is below the allowed maximum as defined by European Union regulations. |
format | Online Article Text |
id | pubmed-6767005 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-67670052019-10-02 A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk Lu, Mei-Yi Kao, Wei-Chen Belkin, Shimshon Cheng, Ji-Yen Sensors (Basel) Article We present an integral smartphone-based whole-cell biosensor, LumiCellSense (LCS), which incorporates a 16-well biochip with an oxygen permeable coating, harboring bioluminescent Escherichia coli bioreporter cells, a macro lens, a lens barrel, a metal heater tray, and a temperature controller, enclosed in a light-impermeable case. The luminescence emitted by the bioreporter cells in response to the presence of the target chemicals is imaged by the phone’s camera, and a dedicated phone-embedded application, LCS_Logger, is employed to calculate photon emission intensity and plot it in real time on the device’s screen. An alert is automatically given when light intensity increases above the baseline, indicating the presence of the target. We demonstrate the efficacy of this system by the detection of residues of an antibiotic, ciprofloxacin (CIP), in whole milk, with a detection threshold of 7.2 ng/mL. This value is below the allowed maximum as defined by European Union regulations. MDPI 2019-09-09 /pmc/articles/PMC6767005/ /pubmed/31505815 http://dx.doi.org/10.3390/s19183882 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Lu, Mei-Yi Kao, Wei-Chen Belkin, Shimshon Cheng, Ji-Yen A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk |
title | A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk |
title_full | A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk |
title_fullStr | A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk |
title_full_unstemmed | A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk |
title_short | A Smartphone-Based Whole-Cell Array Sensor for Detection of Antibiotics in Milk |
title_sort | smartphone-based whole-cell array sensor for detection of antibiotics in milk |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6767005/ https://www.ncbi.nlm.nih.gov/pubmed/31505815 http://dx.doi.org/10.3390/s19183882 |
work_keys_str_mv | AT lumeiyi asmartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT kaoweichen asmartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT belkinshimshon asmartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT chengjiyen asmartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT lumeiyi smartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT kaoweichen smartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT belkinshimshon smartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk AT chengjiyen smartphonebasedwholecellarraysensorfordetectionofantibioticsinmilk |