Cargando…

Physicochemical parameters - hydration performance relationship of the new endodontic cement MTA Repair HP

BACKGROUND: To characterize the chemical composition and textural parameters of the MTA Repair HP precursor powder and their influence to hydration performance. MATERIAL AND METHODS: Un-hydrated precursor material was characterized by X-ray diffraction (XRD), Fourier transform infrared spectroscopy...

Descripción completa

Detalles Bibliográficos
Autores principales: Jiménez-Sánchez, María del Carmen, Segura-Egea, Juan-José, Díaz-Cuenca, Aránzazu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medicina Oral S.L. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6776399/
https://www.ncbi.nlm.nih.gov/pubmed/31598203
http://dx.doi.org/10.4317/jced.56013
Descripción
Sumario:BACKGROUND: To characterize the chemical composition and textural parameters of the MTA Repair HP precursor powder and their influence to hydration performance. MATERIAL AND METHODS: Un-hydrated precursor material was characterized by X-ray diffraction (XRD), Fourier transform infrared spectroscopy (FT-IR), X-ray fluorescence (XRF), laser diffraction (LD), N2 physisorption and field emission gun scanning electron microscopy (FEG-SEM). Setting time was assessed according to ASTM specification C 266. Hydrated material was analysed by XRD, FT-IR, energy dispersive X-ray (EDX) analysis and FEG-SEM. RESULTS: Ca3SiO5 and Ca2SiO4, in addition to CaWO4 as radiopacifier are the main compositional phases. Other measured parameter indicates high specific surface area of 4.8 m2 g-1, high aluminium content of 1.7 wt.% and low initial and final setting times of 12 and 199 min, respectively. Singular microstructural features consisting of high aspect ratio nanoparticles are main constituents of un-hydrated precursor. Besides, FEM-SEM observation shows notably growth of hexagonal shaped plate-like morphologies homogeneously distributed along the sample during hydration process. CONCLUSIONS: The short setting time measured for HP Repair, is correlated with high surface area of precursor powder, high Al content and the absence of compositional sulphate phases. Key words:Bioactive endodontic cements, hydration performance, MTA HP Repair, physicochemical parameters.