A Reinforced Endotracheal Tube Completely Severed by a Patient Bite and Lodged in the Right Main Bronchus

Reinforced endotracheal tubes (ETTs) are designed to resist kinking or compression. However, these have a potential risk of being obstructed or severed by a patient’s bite. We report a case in which a reinforced ETT was severed by tube-bite while the patient was in the prone position during an inten...

Descripción completa

Detalles Bibliográficos
Autores principales: Yoon, Susie, Choo, Hyunjung, Kim, Se Eun, Kwon, Heeyeon, Lee, Hannah
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society of Critical Care Medicine 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6786742/
https://www.ncbi.nlm.nih.gov/pubmed/31723618
http://dx.doi.org/10.4266/kjccm.2016.00437
Descripción
Sumario:Reinforced endotracheal tubes (ETTs) are designed to resist kinking or compression. However, these have a potential risk of being obstructed or severed by a patient’s bite. We report a case in which a reinforced ETT was severed by tube-bite while the patient was in the prone position during an intensive care unit stay. Bronchoscopic evaluation showed that the severed distal part of the tube had lodged in the patient’s right main bronchus, and it had to be surgically removed. The patency of reinforced ETTs should be carefully monitored in patients intubated in the prone position.