A randomised prospective comparison of equilibration point and changing gas composition during low-flow anaesthesia with sevoflurane vs desflurane

BACKGROUND AND AIMS: Safety and economy are the main concerns while using low- flow anaesthesia with newer inhalational agents. The main objective of this study was to use and compare 'equilibration time' of sevoflurane and desflurane after change-over from high-to-low flow anaesthesia. ME...

Descripción completa

Detalles Bibliográficos
Autores principales: Toal, Pratibha, Chiplonkar, Sheetal, Panchal, Pratik
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6798620/
https://www.ncbi.nlm.nih.gov/pubmed/31649393
http://dx.doi.org/10.4103/ija.IJA_31_19
Descripción
Sumario:BACKGROUND AND AIMS: Safety and economy are the main concerns while using low- flow anaesthesia with newer inhalational agents. The main objective of this study was to use and compare 'equilibration time' of sevoflurane and desflurane after change-over from high-to-low flow anaesthesia. METHODS: This prospective study included consenting adult patients between 18–70 years undergoing elective surgery under general anaesthesia. Patients were randomised initially to receive high-flow anaesthesia with 1.3 MAC of either desflurane or sevoflurane with nitrous oxide. After equilibration point, low-flow anaesthesia was initiated. Heart rate, non-invasive blood pressure, pulse oximeter, 5 electrode ECG and gas monitoring was done. Statistical analysis was done with the help of Med CalC version 12.5.0.0 (student version) and IBM SPSS Version 20.0. RESULTS: Mean equilibration time in sevoflurane group was higher (4.59 ± 0.77 minutes) than desflurane group (3.78 ± 0.56 minutes, P < 0.001). Inspired concentrations of both inhalational agents varied from their vaporiser settings over 2 hours, more so with sevoflurane than desflurane. Inspired oxygen concentration (FiO2) remained above 30% during anaesthesia in both groups with stable haemodynamics. CONCLUSION: Change-over from high-to-low flow anaesthesia is faster in desflurane. With fresh gas flow (FGF) of 1 L with 50% oxygen and dial concentration of 1–1.5% of sevoflurane and 3.8–4.4% of desflurane, the risk of hypoxia is minimal. The disparity between the set and delivered concentrations is more (20%) in sevoflurane than desflurane (12%).