Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000

BACKGROUND: To compare the Mindray BC‐5180 and Sysmex XN‐1000 instruments by analyzing the results of complete blood count in the external quality assessment in Shandong Province in 2018. METHODS: In the external quality assessment, 10 batches of quality control materials were issued throughout the...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Jing, Zhao, Shengmei, Su, Zhenguo, Liu, Xiangdong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6805265/
https://www.ncbi.nlm.nih.gov/pubmed/31218736
http://dx.doi.org/10.1002/jcla.22955
_version_ 1783461341962436608
author Wang, Jing
Zhao, Shengmei
Su, Zhenguo
Liu, Xiangdong
author_facet Wang, Jing
Zhao, Shengmei
Su, Zhenguo
Liu, Xiangdong
author_sort Wang, Jing
collection PubMed
description BACKGROUND: To compare the Mindray BC‐5180 and Sysmex XN‐1000 instruments by analyzing the results of complete blood count in the external quality assessment in Shandong Province in 2018. METHODS: In the external quality assessment, 10 batches of quality control materials were issued throughout the year. The test items were WBC, RBC, Hb, PLT, and HCT. The laboratories using Mindray BC‐5180 and Sysmex XN‐1000 were screened, and the results were analyzed by t test, Passing‐Bablok regression analysis, and Bland‐Altman analysis. RESULTS: Thirty‐six laboratories using Mindray BC‐5180 instruments and thirty‐six laboratories using Sysmex XN‐1000 instruments were screened, and the average difference between the two instruments results is not significant (P > 0.05, t test). Passing‐Bablok regression analysis showed that the 95% confidence interval of the regression equation interception of each test item included 0, and the 95% confidence interval of the slope contained 1, r > 0.98, which showed that the correlation is good. The Bland‐Altman analysis showed that both instruments had more than 95% of the points within the 95% consistency limit (WBC97.2%, RBC95.6%, PLT97.2%, Hb96.7%, HCT97.5%). Within the consistency limit, the absolute value of the difference between the Mindray BC‐5180 instrument and the Sysmex XN‐1000 instrument is WBC 0.14%, RBC 0.26%, PLT 2.7%, and Hb 1.9%. HCT is 0.69%, and the difference is clinically acceptable. CONCLUSION: It can be considered that the two instruments have good correlation and consistency, and the two instruments can replace each other.
format Online
Article
Text
id pubmed-6805265
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-68052652019-11-12 Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000 Wang, Jing Zhao, Shengmei Su, Zhenguo Liu, Xiangdong J Clin Lab Anal Research Articles BACKGROUND: To compare the Mindray BC‐5180 and Sysmex XN‐1000 instruments by analyzing the results of complete blood count in the external quality assessment in Shandong Province in 2018. METHODS: In the external quality assessment, 10 batches of quality control materials were issued throughout the year. The test items were WBC, RBC, Hb, PLT, and HCT. The laboratories using Mindray BC‐5180 and Sysmex XN‐1000 were screened, and the results were analyzed by t test, Passing‐Bablok regression analysis, and Bland‐Altman analysis. RESULTS: Thirty‐six laboratories using Mindray BC‐5180 instruments and thirty‐six laboratories using Sysmex XN‐1000 instruments were screened, and the average difference between the two instruments results is not significant (P > 0.05, t test). Passing‐Bablok regression analysis showed that the 95% confidence interval of the regression equation interception of each test item included 0, and the 95% confidence interval of the slope contained 1, r > 0.98, which showed that the correlation is good. The Bland‐Altman analysis showed that both instruments had more than 95% of the points within the 95% consistency limit (WBC97.2%, RBC95.6%, PLT97.2%, Hb96.7%, HCT97.5%). Within the consistency limit, the absolute value of the difference between the Mindray BC‐5180 instrument and the Sysmex XN‐1000 instrument is WBC 0.14%, RBC 0.26%, PLT 2.7%, and Hb 1.9%. HCT is 0.69%, and the difference is clinically acceptable. CONCLUSION: It can be considered that the two instruments have good correlation and consistency, and the two instruments can replace each other. John Wiley and Sons Inc. 2019-06-20 /pmc/articles/PMC6805265/ /pubmed/31218736 http://dx.doi.org/10.1002/jcla.22955 Text en © 2019 The Authors Journal of Clinical Laboratory Analysis Published by Wiley Periodicals, Inc. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Wang, Jing
Zhao, Shengmei
Su, Zhenguo
Liu, Xiangdong
Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000
title Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000
title_full Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000
title_fullStr Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000
title_full_unstemmed Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000
title_short Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000
title_sort analytical comparison between two hematological analyzer systems: mindray bc‐5180 vs sysmex xn‐1000
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6805265/
https://www.ncbi.nlm.nih.gov/pubmed/31218736
http://dx.doi.org/10.1002/jcla.22955
work_keys_str_mv AT wangjing analyticalcomparisonbetweentwohematologicalanalyzersystemsmindraybc5180vssysmexxn1000
AT zhaoshengmei analyticalcomparisonbetweentwohematologicalanalyzersystemsmindraybc5180vssysmexxn1000
AT suzhenguo analyticalcomparisonbetweentwohematologicalanalyzersystemsmindraybc5180vssysmexxn1000
AT liuxiangdong analyticalcomparisonbetweentwohematologicalanalyzersystemsmindraybc5180vssysmexxn1000