Cargando…

Detection of 8-Hydroxy-2′-Deoxyguanosine Biomarker with a Screen-Printed Electrode Modified with Graphene

In this work we present the preparation of graphene material by exfoliation of graphite rods via pulses of current in electrolyte, containing a mixture of boric acid (0.05 M) and sodium chloride (0.05 M). The material was morphologically and structurally characterized by SEM/TEM/HR-TEM, XRD and FTIR...

Descripción completa

Detalles Bibliográficos
Autores principales: Varodi, Codruta, Pogacean, Florina, Coros, Maria, Rosu, Marcela-Corina, Stefan-van Staden, Raluca-Ioana, Gal, Emese, Tudoran, Lucian-Barbu, Pruneanu, Stela, Mirel, Simona
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806156/
https://www.ncbi.nlm.nih.gov/pubmed/31590208
http://dx.doi.org/10.3390/s19194297
Descripción
Sumario:In this work we present the preparation of graphene material by exfoliation of graphite rods via pulses of current in electrolyte, containing a mixture of boric acid (0.05 M) and sodium chloride (0.05 M). The material was morphologically and structurally characterized by SEM/TEM/HR-TEM, XRD and FTIR techniques. TEM investigation of graphene flakes deposited onto carbon-coated grids allowed the visualization of thin and transparent regions, attributed to few-layer graphene (FLG), as well as thick and dark regions attributed to multi-layer graphene (MLG). The mixed composition of the material was additionally confirmed by XRD, which further indicated that the amount of FLG within the sample was around 83%, while MLG was around 17%. The performance of a screen-printed electrode (SPE) modified with graphene (SPE-Gr) was tested for 8-hydroxy-2′-deoxyguanosine detection. The graphene-modified electrode had a higher sensitivity in comparison with that of SPE, both in standard laboratory solutions (phosphate buffered saline—PBS) and in human saliva.