Technical assessment of a mobile CT scanner for image‐guided brachytherapy
PURPOSE: The imaging performance and dose of a mobile CT scanner (Brainlab Airo®, Munich, Germany) is evaluated, with particular consideration to assessment of technique protocols for image‐guided brachytherapy. METHOD: Dose measurements were performed using a 100‐mm‐length pencil chamber at the cen...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806478/ https://www.ncbi.nlm.nih.gov/pubmed/31578811 http://dx.doi.org/10.1002/acm2.12738 |
_version_ | 1783461636130996224 |
---|---|
author | Chernavsky, Nicole E. Morcos, Marc Wu, Pengwei Viswanathan, Akila N. Siewerdsen, Jeffrey H. |
author_facet | Chernavsky, Nicole E. Morcos, Marc Wu, Pengwei Viswanathan, Akila N. Siewerdsen, Jeffrey H. |
author_sort | Chernavsky, Nicole E. |
collection | PubMed |
description | PURPOSE: The imaging performance and dose of a mobile CT scanner (Brainlab Airo®, Munich, Germany) is evaluated, with particular consideration to assessment of technique protocols for image‐guided brachytherapy. METHOD: Dose measurements were performed using a 100‐mm‐length pencil chamber at the center and periphery of 16‐ and 32‐cm‐diameter CTDI phantoms. Hounsfield unit (HU) accuracy and linearity were assessed using materials of specified electron density (Gammex RMI, Madison, WI), and image uniformity, noise, and noise‐power spectrum (NPS) were evaluated in a 20‐cm‐diameter water phantom as well as an American College of Radiology (ACR) CT accreditation phantom (Model 464, Sun Nuclear, Melbourne, FL). Spatial resolution (modulation transfer function, MTF) was assessed with an edge‐spread phantom and visually assessed with respect to line‐pair patterns in the ACR phantom and in structures of interest in anthropomorphic phantoms. Images were also obtained on a diagnostic CT scanner (Big Bore CT simulator, Philips, Amsterdam, Netherlands) for qualitative and quantitative comparison. The manufacturer’s metal artifact reduction (MAR) algorithm was assessed in an anthropomorphic body phantom containing surgical instrumentation. Performance in application to brachytherapy was assessed with a set of anthropomorphic brachytherapy phantoms — for example, a vaginal cylinder and interstitial ring and tandem. RESULT: Nominal dose for helical and axial modes, respectively, was 56.4 and 78.9 mGy for the head protocol and 17.8 and 24.9 mGy for the body protocol. A high degree of HU accuracy and linearity was observed for both axial and helical scan modes. Image nonuniformity (e.g., cupping artifact) in the transverse (x,y) plane was less than 5 HU, but stitching artifacts (~5 HU) in the longitudinal (z) direction were observed in axial scan mode. Helical and axial modes demonstrated comparable spatial resolution of ~5 lp/cm, with the MTF reduced to 10% at ~0.38 mm(−1). Contrast‐to‐noise ratio was suitable to soft‐tissue visualization (e.g., fat and muscle), but windmill artifacts were observed in helical mode in relation to high‐frequency bone and metal. The MAR algorithm provided modest improvement to image quality. Overall, image quality appeared suitable to relevant clinical tasks in intracavitary and interstitial (e.g., gynecological) brachytherapy, including visualization of soft‐tissue structures in proximity to the applicators. CONCLUSION: The technical assessment highlighted key characteristics of dose and imaging performance pertinent to incorporation of the mobile CT scanner in clinical procedures, helping to inform clinical deployment and technique protocol selection in brachytherapy. For this and other possible applications, the work helps to identify protocols that could reduce radiation dose and/or improve image quality. The work also identified areas for future improvement, including reduction of stitching, windmill, and metal artifacts. |
format | Online Article Text |
id | pubmed-6806478 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-68064782019-10-28 Technical assessment of a mobile CT scanner for image‐guided brachytherapy Chernavsky, Nicole E. Morcos, Marc Wu, Pengwei Viswanathan, Akila N. Siewerdsen, Jeffrey H. J Appl Clin Med Phys Medical Imaging PURPOSE: The imaging performance and dose of a mobile CT scanner (Brainlab Airo®, Munich, Germany) is evaluated, with particular consideration to assessment of technique protocols for image‐guided brachytherapy. METHOD: Dose measurements were performed using a 100‐mm‐length pencil chamber at the center and periphery of 16‐ and 32‐cm‐diameter CTDI phantoms. Hounsfield unit (HU) accuracy and linearity were assessed using materials of specified electron density (Gammex RMI, Madison, WI), and image uniformity, noise, and noise‐power spectrum (NPS) were evaluated in a 20‐cm‐diameter water phantom as well as an American College of Radiology (ACR) CT accreditation phantom (Model 464, Sun Nuclear, Melbourne, FL). Spatial resolution (modulation transfer function, MTF) was assessed with an edge‐spread phantom and visually assessed with respect to line‐pair patterns in the ACR phantom and in structures of interest in anthropomorphic phantoms. Images were also obtained on a diagnostic CT scanner (Big Bore CT simulator, Philips, Amsterdam, Netherlands) for qualitative and quantitative comparison. The manufacturer’s metal artifact reduction (MAR) algorithm was assessed in an anthropomorphic body phantom containing surgical instrumentation. Performance in application to brachytherapy was assessed with a set of anthropomorphic brachytherapy phantoms — for example, a vaginal cylinder and interstitial ring and tandem. RESULT: Nominal dose for helical and axial modes, respectively, was 56.4 and 78.9 mGy for the head protocol and 17.8 and 24.9 mGy for the body protocol. A high degree of HU accuracy and linearity was observed for both axial and helical scan modes. Image nonuniformity (e.g., cupping artifact) in the transverse (x,y) plane was less than 5 HU, but stitching artifacts (~5 HU) in the longitudinal (z) direction were observed in axial scan mode. Helical and axial modes demonstrated comparable spatial resolution of ~5 lp/cm, with the MTF reduced to 10% at ~0.38 mm(−1). Contrast‐to‐noise ratio was suitable to soft‐tissue visualization (e.g., fat and muscle), but windmill artifacts were observed in helical mode in relation to high‐frequency bone and metal. The MAR algorithm provided modest improvement to image quality. Overall, image quality appeared suitable to relevant clinical tasks in intracavitary and interstitial (e.g., gynecological) brachytherapy, including visualization of soft‐tissue structures in proximity to the applicators. CONCLUSION: The technical assessment highlighted key characteristics of dose and imaging performance pertinent to incorporation of the mobile CT scanner in clinical procedures, helping to inform clinical deployment and technique protocol selection in brachytherapy. For this and other possible applications, the work helps to identify protocols that could reduce radiation dose and/or improve image quality. The work also identified areas for future improvement, including reduction of stitching, windmill, and metal artifacts. John Wiley and Sons Inc. 2019-10-02 /pmc/articles/PMC6806478/ /pubmed/31578811 http://dx.doi.org/10.1002/acm2.12738 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Medical Imaging Chernavsky, Nicole E. Morcos, Marc Wu, Pengwei Viswanathan, Akila N. Siewerdsen, Jeffrey H. Technical assessment of a mobile CT scanner for image‐guided brachytherapy |
title | Technical assessment of a mobile CT scanner for image‐guided brachytherapy |
title_full | Technical assessment of a mobile CT scanner for image‐guided brachytherapy |
title_fullStr | Technical assessment of a mobile CT scanner for image‐guided brachytherapy |
title_full_unstemmed | Technical assessment of a mobile CT scanner for image‐guided brachytherapy |
title_short | Technical assessment of a mobile CT scanner for image‐guided brachytherapy |
title_sort | technical assessment of a mobile ct scanner for image‐guided brachytherapy |
topic | Medical Imaging |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806478/ https://www.ncbi.nlm.nih.gov/pubmed/31578811 http://dx.doi.org/10.1002/acm2.12738 |
work_keys_str_mv | AT chernavskynicolee technicalassessmentofamobilectscannerforimageguidedbrachytherapy AT morcosmarc technicalassessmentofamobilectscannerforimageguidedbrachytherapy AT wupengwei technicalassessmentofamobilectscannerforimageguidedbrachytherapy AT viswanathanakilan technicalassessmentofamobilectscannerforimageguidedbrachytherapy AT siewerdsenjeffreyh technicalassessmentofamobilectscannerforimageguidedbrachytherapy |