Feasibility of a GATE Monte Carlo platform in a clinical pretreatment QA system for VMAT treatment plans using TrueBeam with an HD120 multileaf collimator
PURPOSE: To evaluate the quality of patient‐specific complicated treatment plans, including commercialized treatment planning systems (TPS) and commissioned beam data, we developed a process of quality assurance (QA) using a Monte Carlo (MC) platform. Specifically, we constructed an interface system...
Autores principales: | Lee, Boram, Jeong, Seonghoon, Chung, Kwangzoo, Yoon, Myonggeun, Park, Hee Chul, Han, Youngyih, Jung, Sang Hoon |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806485/ https://www.ncbi.nlm.nih.gov/pubmed/31544350 http://dx.doi.org/10.1002/acm2.12718 |
Ejemplares similares
-
Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans
por: Bergman, Alanah M., et al.
Publicado: (2014) -
Monte Carlo modeling of a Novalis TX Varian 6 MV with HD‐120 multileaf collimator
por: Vazquez‐Quino, Luis Alberto, et al.
Publicado: (2012) -
Monitor unit verification for Varian TrueBeam VMAT plans using Monte Carlo calculations and phase space data
por: Pant, Ankit, et al.
Publicado: (2023) -
Use of TrueBeam developer mode for imaging QA
por: Valdes, Gilmer, et al.
Publicado: (2015) -
Monte Carlo Study of Unflattened Photon Beams Shaped by Multileaf Collimator
por: Kajaria, A., et al.
Publicado: (2019)