Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system

We tested whether an ionization chamber array (ICA) and a one‐dimensional water scanner (1DS) could be used instead of a three‐dimensional water scanning system (3DWS) for acceptance testing and commissioning verification of the Varian Halcyon–Eclipse Treatment Planning System (TPS). The Halcyon lin...

Descripción completa

Detalles Bibliográficos
Autores principales: Gao, Song, Netherton, Tucker, Chetvertkov, Mikhail A., Li, Yuting, Court, Laurence E., Simon, William E., Shi, Jie, Balter, Peter A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806699/
https://www.ncbi.nlm.nih.gov/pubmed/31553525
http://dx.doi.org/10.1002/acm2.12719
_version_ 1783461684504952832
author Gao, Song
Netherton, Tucker
Chetvertkov, Mikhail A.
Li, Yuting
Court, Laurence E.
Simon, William E.
Shi, Jie
Balter, Peter A.
author_facet Gao, Song
Netherton, Tucker
Chetvertkov, Mikhail A.
Li, Yuting
Court, Laurence E.
Simon, William E.
Shi, Jie
Balter, Peter A.
author_sort Gao, Song
collection PubMed
description We tested whether an ionization chamber array (ICA) and a one‐dimensional water scanner (1DS) could be used instead of a three‐dimensional water scanning system (3DWS) for acceptance testing and commissioning verification of the Varian Halcyon–Eclipse Treatment Planning System (TPS). The Halcyon linear accelerator has a single 6‐MV flattening‐filter‐free beam and a nonadjustable beam model for the TPS. Beam data were measured with a 1DS, ICA, ionization chambers, and electrometer. Acceptance testing and commissioning were done simultaneously by comparing the measured data with TPS‐calculated percent‐depth‐dose (PDD) and profiles. The ICA was used to measure profiles of various field sizes (10‐, 20‐, and 28 cm(2)) at depths of d(max) (1.3 cm), 5‐, 10‐, and 20 cm. The 1DS was used for output factors (OFs) and PDDs. OFs were measured with 1DS for various fields (2–28 cm(2)) at a source‐to‐surface distance of 90 cm. All measured data were compared with TPS‐calculations. Profiles, off‐axis ratios (OAR), PDDs and OFs were also measured with a 3DWS as a secondary check. Profiles between the ICA and TPS (ICA and 3DWS) at various depths across the fields indicated that the maximum discrepancies in high‐dose and low‐dose tail were within 2% and 3%, respectively, and the maximum distance‐to‐agreement in the penumbra region was <3 mm. The largest OAR differences between ICA and TPS (ICA and 3DWS) values were 0.23% (−0.25%) for a 28 × 28 cm(2) field, and the largest point‐by‐point PDD differences between 1DS and TPS (1DS and 3DWS) were −0.41% ± 0.12% (−0.32% ± 0.17%) across the fields. Both OAR and PDD showed the beam energy is well matched to the TPS model. The average ratios of 1DS‐measured OFs to the TPS (1DS to 3DWS) values were 1.000 ± 0.002 (0.999 ± 0.003). The Halcyon–Eclipse system can be accepted and commissioned without the need for a 3DWS.
format Online
Article
Text
id pubmed-6806699
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-68066992019-10-28 Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system Gao, Song Netherton, Tucker Chetvertkov, Mikhail A. Li, Yuting Court, Laurence E. Simon, William E. Shi, Jie Balter, Peter A. J Appl Clin Med Phys Radiation Oncology Physics We tested whether an ionization chamber array (ICA) and a one‐dimensional water scanner (1DS) could be used instead of a three‐dimensional water scanning system (3DWS) for acceptance testing and commissioning verification of the Varian Halcyon–Eclipse Treatment Planning System (TPS). The Halcyon linear accelerator has a single 6‐MV flattening‐filter‐free beam and a nonadjustable beam model for the TPS. Beam data were measured with a 1DS, ICA, ionization chambers, and electrometer. Acceptance testing and commissioning were done simultaneously by comparing the measured data with TPS‐calculated percent‐depth‐dose (PDD) and profiles. The ICA was used to measure profiles of various field sizes (10‐, 20‐, and 28 cm(2)) at depths of d(max) (1.3 cm), 5‐, 10‐, and 20 cm. The 1DS was used for output factors (OFs) and PDDs. OFs were measured with 1DS for various fields (2–28 cm(2)) at a source‐to‐surface distance of 90 cm. All measured data were compared with TPS‐calculations. Profiles, off‐axis ratios (OAR), PDDs and OFs were also measured with a 3DWS as a secondary check. Profiles between the ICA and TPS (ICA and 3DWS) at various depths across the fields indicated that the maximum discrepancies in high‐dose and low‐dose tail were within 2% and 3%, respectively, and the maximum distance‐to‐agreement in the penumbra region was <3 mm. The largest OAR differences between ICA and TPS (ICA and 3DWS) values were 0.23% (−0.25%) for a 28 × 28 cm(2) field, and the largest point‐by‐point PDD differences between 1DS and TPS (1DS and 3DWS) were −0.41% ± 0.12% (−0.32% ± 0.17%) across the fields. Both OAR and PDD showed the beam energy is well matched to the TPS model. The average ratios of 1DS‐measured OFs to the TPS (1DS to 3DWS) values were 1.000 ± 0.002 (0.999 ± 0.003). The Halcyon–Eclipse system can be accepted and commissioned without the need for a 3DWS. John Wiley and Sons Inc. 2019-09-25 /pmc/articles/PMC6806699/ /pubmed/31553525 http://dx.doi.org/10.1002/acm2.12719 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Gao, Song
Netherton, Tucker
Chetvertkov, Mikhail A.
Li, Yuting
Court, Laurence E.
Simon, William E.
Shi, Jie
Balter, Peter A.
Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
title Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
title_full Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
title_fullStr Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
title_full_unstemmed Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
title_short Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
title_sort acceptance and verification of the halcyon‐eclipse linear accelerator‐treatment planning system without 3d water scanning system
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6806699/
https://www.ncbi.nlm.nih.gov/pubmed/31553525
http://dx.doi.org/10.1002/acm2.12719
work_keys_str_mv AT gaosong acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT nethertontucker acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT chetvertkovmikhaila acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT liyuting acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT courtlaurencee acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT simonwilliame acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT shijie acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem
AT balterpetera acceptanceandverificationofthehalcyoneclipselinearacceleratortreatmentplanningsystemwithout3dwaterscanningsystem