Is fat suppression in T1 and T2 FSE with mDixon superior to the frequency selection-based SPAIR technique in musculoskeletal tumor imaging?

OBJECTIVE: To determine the image quality of fast spin echo (FSE) with mDixon relative to spectral attenuated inversion recovery (SPAIR) FSE sequences in musculoskeletal tumor imaging on a 1.5-T MRI system. MATERIALS AND METHODS: In a HIPAA-compliant prospective study, 265 patients requiring musculo...

Descripción completa

Detalles Bibliográficos
Autores principales: Huijgen, Willemijn H. F., van Rijswijk, Catherina S. P., Bloem, Johan L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6813285/
https://www.ncbi.nlm.nih.gov/pubmed/31154494
http://dx.doi.org/10.1007/s00256-019-03227-8
Descripción
Sumario:OBJECTIVE: To determine the image quality of fast spin echo (FSE) with mDixon relative to spectral attenuated inversion recovery (SPAIR) FSE sequences in musculoskeletal tumor imaging on a 1.5-T MRI system. MATERIALS AND METHODS: In a HIPAA-compliant prospective study, 265 patients requiring musculoskeletal tumor MRI scans were included. Patient consent was waived by the medical ethical committee. Two radiologists compared SPAIR and mDixon FSE water-only images in both T2- and T1-weighted gadolinium-enhanced (T1-Gd) sequences using a five-point scale (paired samples t test and visual grading characteristics curves (VGC)). Homogeneity of fat suppression, noise, contrast, several artifacts (motion, phase, edge blurring and water–fat swap) and subjective preference were evaluated. RESULTS: Readers did not have subjective preference for either sequence in 71% and 55% (reader 1 and 2, respectively). Scores for homogeneous fat suppression were significantly (p < 0.01) higher for mDixon (4.88 in T2 and 4.87 in T1-Gd) than for SPAIR (4.31 for T2 and 4.21 for T1-Gd). All VGC curves for homogeneity demonstrated preference for mDixon. In 57 individual mDixon cases, fat-suppression homogeneity was strikingly better (≥ 2 points higher), namely in areas with field heterogeneity. Average noise and contrast scores were slightly higher for mDixon, as were motion artifact scores for SPAIR (< 0.5 points difference). CONCLUSIONS: mDixon fat suppression was significantly more homogeneous than SPAIR on both T2 and T1-Gd FSE images in musculoskeletal tumor protocols. In areas of field inhomogeneity, mDixon outperforms SPAIR. SPAIR had slightly less motion artifacts than mDixon.