Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST

PURPOSE: To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST (CST) in healthy subjects. METHODS: In a prospective, observational study, IOP measurements with GAT (GAT-IOPc), ORA (...

Descripción completa

Detalles Bibliográficos
Autores principales: Ramm, Lisa, Herber, Robert, Spoerl, Eberhard, Raiskup, Frederik, Pillunat, Lutz E., Terai, Naim
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6815996/
https://www.ncbi.nlm.nih.gov/pubmed/31737355
http://dx.doi.org/10.1155/2019/3879651
Descripción
Sumario:PURPOSE: To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST (CST) in healthy subjects. METHODS: In a prospective, observational study, IOP measurements with GAT (GAT-IOPc), ORA (IOPcc), DCT (DCT-IOP), and CST (bIOP) were performed and analyzed in 94 healthy subjects. RESULTS: Mean age of the participants was 45.6 ± 17.2 years (range 18 to 81 years). Mean GAT-IOPc was 12.9 ± 2.4 mmHg, mean DCT-IOP was 16.1 ± 2.6 mmHg, and mean IOPcc was 15.6 ± 3.3 mmHg. DCT-IOP and IOPcc were significantly higher than GAT-IOPc (P < 0.001). Mean bIOP was 13.5 ± 2.4 mmHg that was slightly higher but not significantly different from GAT-IOPc (P=0.146). Correlation analysis of IOP values and central corneal thickness (CCT) revealed a negative correlation between GAT-IOPc and CCT (r = −0.347; P=0.001). However, IOPcc, DCT-IOP, and bIOP showed no significant correlation to CCT. Only bIOP revealed a weak but significant age dependency (r = 0.321, P=0.002). CONCLUSION: All tonometry devices showed a good agreement of biomechanical corrected IOP values with GAT-IOPc. As no influence of CCT on IOPcc, DCT-IOP, and bIOP was detectable, the used correction algorithms appear to be appropriate in these tonometers in the clinical setting. The highest agreement was found between GAT-IOPc and bIOP. However, bIOP weakly correlated with participants' age. Further studies are needed to elucidate the role of bIOP for IOP measurement.