Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers

BACKGROUND: Automated cellular analyzers are expected to improve the analytical performance in body fluid (BF) analysis. We evaluated the analytical performance of three automated cellular analyzers and established optimum reflex analysis guidelines. METHODS: A total of 542 BF samples (88 cerebrospi...

Descripción completa

Detalles Bibliográficos
Autores principales: Cho, Jooyoung, Oh, Joowon, Lee, Sang-Guk, Lee, You-Hee, Song, Jaewoo, Kim, Jeong-Ho
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society for Laboratory Medicine 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6822009/
https://www.ncbi.nlm.nih.gov/pubmed/31650728
http://dx.doi.org/10.3343/alm.2020.40.2.122
_version_ 1783464248101306368
author Cho, Jooyoung
Oh, Joowon
Lee, Sang-Guk
Lee, You-Hee
Song, Jaewoo
Kim, Jeong-Ho
author_facet Cho, Jooyoung
Oh, Joowon
Lee, Sang-Guk
Lee, You-Hee
Song, Jaewoo
Kim, Jeong-Ho
author_sort Cho, Jooyoung
collection PubMed
description BACKGROUND: Automated cellular analyzers are expected to improve the analytical performance in body fluid (BF) analysis. We evaluated the analytical performance of three automated cellular analyzers and established optimum reflex analysis guidelines. METHODS: A total of 542 BF samples (88 cerebrospinal fluid [CSF] samples and 454 non-CSF samples) were examined using manual counting and three automated cellular analyzers: UniCel DxH 800 (Beckman Coulter), XN-350 (Sysmex), and UF-5000 (Sysmex). Additionally, 2,779 BF analysis results were retrospectively reviewed. For malignant cell analysis, the receiver operating characteristic (ROC) curve was used, and the detection of high fluorescence-BF cells (HF-BFs) using the XN-350 analyzer was compared with cytology results. RESULTS: All three analyzers showed good agreement for total nucleated cell (TNC) and red blood cell (RBC) counts, except for the RBC count in CSF samples using the UniCel DxH 800. However, variable degrees of differences were observed during differential cell counting. For malignant cell analysis, the area under the curve was 0.63 for the XN-350 analyzer and 0.76 for manual counting. We established our own reflex analysis guidelines as follows: HF-BFs <0.7/100 white blood cells (WBCs) is the criterion for quick scans with 100× magnification microscopic examination as a rule-out cut-off, while HF-BFs >83.4/100 WBCs or eosinophils >3.8% are the criteria for mandatory double check confirmation with 1,000× magnification examination. CONCLUSIONS: The three automated analyzers showed good analytical performances. Application of reflex analysis guidelines is recommended for eosinophils and HF-BFs, and manual confirmation is warranted.
format Online
Article
Text
id pubmed-6822009
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher The Korean Society for Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-68220092020-03-01 Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers Cho, Jooyoung Oh, Joowon Lee, Sang-Guk Lee, You-Hee Song, Jaewoo Kim, Jeong-Ho Ann Lab Med Original Article BACKGROUND: Automated cellular analyzers are expected to improve the analytical performance in body fluid (BF) analysis. We evaluated the analytical performance of three automated cellular analyzers and established optimum reflex analysis guidelines. METHODS: A total of 542 BF samples (88 cerebrospinal fluid [CSF] samples and 454 non-CSF samples) were examined using manual counting and three automated cellular analyzers: UniCel DxH 800 (Beckman Coulter), XN-350 (Sysmex), and UF-5000 (Sysmex). Additionally, 2,779 BF analysis results were retrospectively reviewed. For malignant cell analysis, the receiver operating characteristic (ROC) curve was used, and the detection of high fluorescence-BF cells (HF-BFs) using the XN-350 analyzer was compared with cytology results. RESULTS: All three analyzers showed good agreement for total nucleated cell (TNC) and red blood cell (RBC) counts, except for the RBC count in CSF samples using the UniCel DxH 800. However, variable degrees of differences were observed during differential cell counting. For malignant cell analysis, the area under the curve was 0.63 for the XN-350 analyzer and 0.76 for manual counting. We established our own reflex analysis guidelines as follows: HF-BFs <0.7/100 white blood cells (WBCs) is the criterion for quick scans with 100× magnification microscopic examination as a rule-out cut-off, while HF-BFs >83.4/100 WBCs or eosinophils >3.8% are the criteria for mandatory double check confirmation with 1,000× magnification examination. CONCLUSIONS: The three automated analyzers showed good analytical performances. Application of reflex analysis guidelines is recommended for eosinophils and HF-BFs, and manual confirmation is warranted. The Korean Society for Laboratory Medicine 2020-03 2019-10-23 /pmc/articles/PMC6822009/ /pubmed/31650728 http://dx.doi.org/10.3343/alm.2020.40.2.122 Text en © The Korean Society for Laboratory Medicine http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Cho, Jooyoung
Oh, Joowon
Lee, Sang-Guk
Lee, You-Hee
Song, Jaewoo
Kim, Jeong-Ho
Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers
title Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers
title_full Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers
title_fullStr Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers
title_full_unstemmed Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers
title_short Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers
title_sort performance evaluation of body fluid cellular analysis using the beckman coulter unicel dxh 800, sysmex xn-350, and uf-5000 automated cellular analyzers
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6822009/
https://www.ncbi.nlm.nih.gov/pubmed/31650728
http://dx.doi.org/10.3343/alm.2020.40.2.122
work_keys_str_mv AT chojooyoung performanceevaluationofbodyfluidcellularanalysisusingthebeckmancoulteruniceldxh800sysmexxn350anduf5000automatedcellularanalyzers
AT ohjoowon performanceevaluationofbodyfluidcellularanalysisusingthebeckmancoulteruniceldxh800sysmexxn350anduf5000automatedcellularanalyzers
AT leesangguk performanceevaluationofbodyfluidcellularanalysisusingthebeckmancoulteruniceldxh800sysmexxn350anduf5000automatedcellularanalyzers
AT leeyouhee performanceevaluationofbodyfluidcellularanalysisusingthebeckmancoulteruniceldxh800sysmexxn350anduf5000automatedcellularanalyzers
AT songjaewoo performanceevaluationofbodyfluidcellularanalysisusingthebeckmancoulteruniceldxh800sysmexxn350anduf5000automatedcellularanalyzers
AT kimjeongho performanceevaluationofbodyfluidcellularanalysisusingthebeckmancoulteruniceldxh800sysmexxn350anduf5000automatedcellularanalyzers