Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine

BACKGROUND: We aimed to evaluate the basic performance and basic parameters of a new type of blood vacuum tube cap opening and closing machine for further application. METHODS: From July 2013 to March 2018, 110, 000 cases involved in the clinical trials in Third People’s Hospital of Dongguan City we...

Descripción completa

Detalles Bibliográficos
Autores principales: PU, Rong, PU, Meicen, XIN, Qingsong, CHEN, Feng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Tehran University of Medical Sciences 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6825677/
https://www.ncbi.nlm.nih.gov/pubmed/31700815
_version_ 1783464931457236992
author PU, Rong
PU, Meicen
XIN, Qingsong
CHEN, Feng
author_facet PU, Rong
PU, Meicen
XIN, Qingsong
CHEN, Feng
author_sort PU, Rong
collection PubMed
description BACKGROUND: We aimed to evaluate the basic performance and basic parameters of a new type of blood vacuum tube cap opening and closing machine for further application. METHODS: From July 2013 to March 2018, 110, 000 cases involved in the clinical trials in Third People’s Hospital of Dongguan City were selected as the instrument group, and 10,000 cases were selected as the manual operation group. The application demonstration and instrument performance assessment were performed by five units. Unified evaluation indicatory system and the standard of assessment were set up. The instrument assessments and demonstration tests were carried out by these 5 units. Finally, the basic parameters of the instrument were analyzed and compared. RESULTS: The new instrument had excellent performance, and thirty parameters were excellent. Five patents had been granted already. There were no differences between the instrument groups in different units, and among different units. However, there were some differences in the manual operation group between different units. The average cap opening time was 21 “04 in the instrument group, and 152” 48 in the manual operation group (P<0.05); the cap closing time was 18 “56 in the instrument group, and 104”24 in the manual operation group (P<0.05). The instrument group outperformed the manual operation group in cap closing time, pollution rate, spill rate and failure rate (P<0.01). The design of the instrument was original. CONCLUSION: This blood vacuum tube cap opening and closing machine can perform a wide variety of functions, and it is stable, intelligent and superior to manual operation. More efforts need to be made for the industrialization of this instrument.
format Online
Article
Text
id pubmed-6825677
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Tehran University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-68256772019-11-07 Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine PU, Rong PU, Meicen XIN, Qingsong CHEN, Feng Iran J Public Health Original Article BACKGROUND: We aimed to evaluate the basic performance and basic parameters of a new type of blood vacuum tube cap opening and closing machine for further application. METHODS: From July 2013 to March 2018, 110, 000 cases involved in the clinical trials in Third People’s Hospital of Dongguan City were selected as the instrument group, and 10,000 cases were selected as the manual operation group. The application demonstration and instrument performance assessment were performed by five units. Unified evaluation indicatory system and the standard of assessment were set up. The instrument assessments and demonstration tests were carried out by these 5 units. Finally, the basic parameters of the instrument were analyzed and compared. RESULTS: The new instrument had excellent performance, and thirty parameters were excellent. Five patents had been granted already. There were no differences between the instrument groups in different units, and among different units. However, there were some differences in the manual operation group between different units. The average cap opening time was 21 “04 in the instrument group, and 152” 48 in the manual operation group (P<0.05); the cap closing time was 18 “56 in the instrument group, and 104”24 in the manual operation group (P<0.05). The instrument group outperformed the manual operation group in cap closing time, pollution rate, spill rate and failure rate (P<0.01). The design of the instrument was original. CONCLUSION: This blood vacuum tube cap opening and closing machine can perform a wide variety of functions, and it is stable, intelligent and superior to manual operation. More efforts need to be made for the industrialization of this instrument. Tehran University of Medical Sciences 2019-09 /pmc/articles/PMC6825677/ /pubmed/31700815 Text en Copyright© Iranian Public Health Association & Tehran University of Medical Sciences http://creativecommons.org/licenses/by/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
PU, Rong
PU, Meicen
XIN, Qingsong
CHEN, Feng
Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine
title Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine
title_full Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine
title_fullStr Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine
title_full_unstemmed Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine
title_short Parameter Evaluation and Application Demonstration of a New Type of Blood Vacuum Tube Cap Opening and Closing Machine
title_sort parameter evaluation and application demonstration of a new type of blood vacuum tube cap opening and closing machine
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6825677/
https://www.ncbi.nlm.nih.gov/pubmed/31700815
work_keys_str_mv AT purong parameterevaluationandapplicationdemonstrationofanewtypeofbloodvacuumtubecapopeningandclosingmachine
AT pumeicen parameterevaluationandapplicationdemonstrationofanewtypeofbloodvacuumtubecapopeningandclosingmachine
AT xinqingsong parameterevaluationandapplicationdemonstrationofanewtypeofbloodvacuumtubecapopeningandclosingmachine
AT chenfeng parameterevaluationandapplicationdemonstrationofanewtypeofbloodvacuumtubecapopeningandclosingmachine