Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study

BACKGROUND: Blood pressure (BP) is an important modifiable cardiovascular risk factor, yet its long-term monitoring remains problematic. Wearable cuffless devices enable the capture of multiple BP measures during everyday activities and could improve BP monitoring, but little is known about their va...

Descripción completa

Detalles Bibliográficos
Autores principales: Islam, Sheikh Mohammed Shariful, Cartledge, Susie, Karmakar, Chandan, Rawstorn, Jonathan Charles, Fraser, Steve F, Chow, Clara, Maddison, Ralph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6827985/
https://www.ncbi.nlm.nih.gov/pubmed/31628788
http://dx.doi.org/10.2196/14706
_version_ 1783465366237741056
author Islam, Sheikh Mohammed Shariful
Cartledge, Susie
Karmakar, Chandan
Rawstorn, Jonathan Charles
Fraser, Steve F
Chow, Clara
Maddison, Ralph
author_facet Islam, Sheikh Mohammed Shariful
Cartledge, Susie
Karmakar, Chandan
Rawstorn, Jonathan Charles
Fraser, Steve F
Chow, Clara
Maddison, Ralph
author_sort Islam, Sheikh Mohammed Shariful
collection PubMed
description BACKGROUND: Blood pressure (BP) is an important modifiable cardiovascular risk factor, yet its long-term monitoring remains problematic. Wearable cuffless devices enable the capture of multiple BP measures during everyday activities and could improve BP monitoring, but little is known about their validity or acceptability. OBJECTIVE: This study aimed to validate a wrist-worn cuffless wearable BP device (Model T2; TMART Technologies Limited) and assess its acceptability among users and health care professionals. METHODS: A mixed methods study was conducted to examine the validity and comparability of a wearable cuffless BP device against ambulatory and home devices. BP was measured simultaneously over 24 hours using wearable and ambulatory devices and over 7 days using wearable and home devices. Pearson correlation coefficients compared the degree of association between the measures, and limits of agreement (LOA; Bland-Altman plots) were generated to assess measurement bias. Semistructured interviews were conducted with users and 10 health care professionals to assess acceptability, facilitators, and barriers to using the wearable device. Interviews were audio recorded, transcribed, and analyzed. RESULTS: A total of 9090 BP measurements were collected from 20 healthy volunteers (mean 20.3 years, SD 5.4; N=10 females). Mean (SD) systolic BP (SBP)/diastolic BP (DBP) measured using the ambulatory (24 hours), home (7 days), and wearable (7 days) devices were 126 (SD 10)/75 (SD 6) mm Hg, 112 (SD 10)/71 (SD 9) mm Hg and 125 (SD 4)/77 (SD 3) mm Hg, respectively. Mean (LOA) biases and precision between the wearable and ambulatory devices over 24 hours were 0.5 (−10.1 to 11.1) mm Hg for SBP and 2.24 (−17.6 to 13.1) mm Hg for DBP. The mean biases (LOA) and precision between the wearable and home device over 7 days were −12.7 (−28.7 to 3.4) mm Hg for SBP and −5.6 (−20.5 to 9.2) mm Hg for DBP. The wearable BP device was well accepted by participants who found the device easy to wear and use. Both participants and health care providers agreed that the wearable cuffless devices were easy to use and that they could be used to improve BP monitoring. CONCLUSIONS: Wearable BP measures compared well against a gold-standard ambulatory device, indicating potential for this user-friendly method to augment BP management, particularly by enabling long-term monitoring that could improve treatment titration and increase understanding of users’ BP response during daily activity and stressors.
format Online
Article
Text
id pubmed-6827985
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-68279852019-11-14 Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study Islam, Sheikh Mohammed Shariful Cartledge, Susie Karmakar, Chandan Rawstorn, Jonathan Charles Fraser, Steve F Chow, Clara Maddison, Ralph JMIR Mhealth Uhealth Original Paper BACKGROUND: Blood pressure (BP) is an important modifiable cardiovascular risk factor, yet its long-term monitoring remains problematic. Wearable cuffless devices enable the capture of multiple BP measures during everyday activities and could improve BP monitoring, but little is known about their validity or acceptability. OBJECTIVE: This study aimed to validate a wrist-worn cuffless wearable BP device (Model T2; TMART Technologies Limited) and assess its acceptability among users and health care professionals. METHODS: A mixed methods study was conducted to examine the validity and comparability of a wearable cuffless BP device against ambulatory and home devices. BP was measured simultaneously over 24 hours using wearable and ambulatory devices and over 7 days using wearable and home devices. Pearson correlation coefficients compared the degree of association between the measures, and limits of agreement (LOA; Bland-Altman plots) were generated to assess measurement bias. Semistructured interviews were conducted with users and 10 health care professionals to assess acceptability, facilitators, and barriers to using the wearable device. Interviews were audio recorded, transcribed, and analyzed. RESULTS: A total of 9090 BP measurements were collected from 20 healthy volunteers (mean 20.3 years, SD 5.4; N=10 females). Mean (SD) systolic BP (SBP)/diastolic BP (DBP) measured using the ambulatory (24 hours), home (7 days), and wearable (7 days) devices were 126 (SD 10)/75 (SD 6) mm Hg, 112 (SD 10)/71 (SD 9) mm Hg and 125 (SD 4)/77 (SD 3) mm Hg, respectively. Mean (LOA) biases and precision between the wearable and ambulatory devices over 24 hours were 0.5 (−10.1 to 11.1) mm Hg for SBP and 2.24 (−17.6 to 13.1) mm Hg for DBP. The mean biases (LOA) and precision between the wearable and home device over 7 days were −12.7 (−28.7 to 3.4) mm Hg for SBP and −5.6 (−20.5 to 9.2) mm Hg for DBP. The wearable BP device was well accepted by participants who found the device easy to wear and use. Both participants and health care providers agreed that the wearable cuffless devices were easy to use and that they could be used to improve BP monitoring. CONCLUSIONS: Wearable BP measures compared well against a gold-standard ambulatory device, indicating potential for this user-friendly method to augment BP management, particularly by enabling long-term monitoring that could improve treatment titration and increase understanding of users’ BP response during daily activity and stressors. JMIR Publications 2019-09-14 /pmc/articles/PMC6827985/ /pubmed/31628788 http://dx.doi.org/10.2196/14706 Text en ©Sheikh Mohammed Shariful Islam, Susie Cartledge, Chandan Karmakar, Jonathan Charles Rawstorn, Steve F Fraser, Clara Chow, Ralph Maddison. Originally published in JMIR Mhealth and Uhealth (http://mhealth.jmir.org), 04.10.2019 https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mhealth and uhealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included.
spellingShingle Original Paper
Islam, Sheikh Mohammed Shariful
Cartledge, Susie
Karmakar, Chandan
Rawstorn, Jonathan Charles
Fraser, Steve F
Chow, Clara
Maddison, Ralph
Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
title Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
title_full Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
title_fullStr Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
title_full_unstemmed Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
title_short Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
title_sort validation and acceptability of a cuffless wrist-worn wearable blood pressure monitoring device among users and health care professionals: mixed methods study
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6827985/
https://www.ncbi.nlm.nih.gov/pubmed/31628788
http://dx.doi.org/10.2196/14706
work_keys_str_mv AT islamsheikhmohammedshariful validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy
AT cartledgesusie validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy
AT karmakarchandan validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy
AT rawstornjonathancharles validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy
AT fraserstevef validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy
AT chowclara validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy
AT maddisonralph validationandacceptabilityofacufflesswristwornwearablebloodpressuremonitoringdeviceamongusersandhealthcareprofessionalsmixedmethodsstudy