Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study
BACKGROUND: Atrial fibrillation (AF) is often paroxysmal and minimally symptomatic, hindering its diagnosis. Smartwatches may enhance AF care by facilitating long-term, noninvasive monitoring. OBJECTIVE: This study aimed to examine the accuracy and usability of arrhythmia discrimination using a smar...
Autores principales: | , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6834225/ https://www.ncbi.nlm.nih.gov/pubmed/31758787 http://dx.doi.org/10.2196/13850 |
_version_ | 1783466447514632192 |
---|---|
author | Ding, Eric Y Han, Dong Whitcomb, Cody Bashar, Syed Khairul Adaramola, Oluwaseun Soni, Apurv Saczynski, Jane Fitzgibbons, Timothy P Moonis, Majaz Lubitz, Steven A Lessard, Darleen Hills, Mellanie True Barton, Bruce Chon, Ki McManus, David D |
author_facet | Ding, Eric Y Han, Dong Whitcomb, Cody Bashar, Syed Khairul Adaramola, Oluwaseun Soni, Apurv Saczynski, Jane Fitzgibbons, Timothy P Moonis, Majaz Lubitz, Steven A Lessard, Darleen Hills, Mellanie True Barton, Bruce Chon, Ki McManus, David D |
author_sort | Ding, Eric Y |
collection | PubMed |
description | BACKGROUND: Atrial fibrillation (AF) is often paroxysmal and minimally symptomatic, hindering its diagnosis. Smartwatches may enhance AF care by facilitating long-term, noninvasive monitoring. OBJECTIVE: This study aimed to examine the accuracy and usability of arrhythmia discrimination using a smartwatch. METHODS: A total of 40 adults presenting to a cardiology clinic wore a smartwatch and Holter monitor and performed scripted movements to simulate activities of daily living (ADLs). Participants’ clinical and sociodemographic characteristics were abstracted from medical records. Participants completed a questionnaire assessing different domains of the device’s usability. Pulse recordings were analyzed blindly using a real-time realizable algorithm and compared with gold-standard Holter monitoring. RESULTS: The average age of participants was 71 (SD 8) years; most participants had AF risk factors and 23% (9/39) were in AF. About half of the participants owned smartphones, but none owned smartwatches. Participants wore the smartwatch for 42 (SD 14) min while generating motion noise to simulate ADLs. The algorithm determined 53 of the 314 30-second noise-free pulse segments as consistent with AF. Compared with the gold standard, the algorithm demonstrated excellent sensitivity (98.2%), specificity (98.1%), and accuracy (98.1%) for identifying irregular pulse. Two-thirds of participants considered the smartwatch highly usable. Younger age and prior cardioversion were associated with greater overall comfort and comfort with data privacy with using a smartwatch for rhythm monitoring, respectively. CONCLUSIONS: A real-time realizable algorithm analyzing smartwatch pulse recordings demonstrated high accuracy for identifying pulse irregularities among older participants. Despite advanced age, lack of smartwatch familiarity, and high burden of comorbidities, participants found the smartwatch to be highly acceptable. |
format | Online Article Text |
id | pubmed-6834225 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-68342252019-11-21 Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study Ding, Eric Y Han, Dong Whitcomb, Cody Bashar, Syed Khairul Adaramola, Oluwaseun Soni, Apurv Saczynski, Jane Fitzgibbons, Timothy P Moonis, Majaz Lubitz, Steven A Lessard, Darleen Hills, Mellanie True Barton, Bruce Chon, Ki McManus, David D JMIR Cardio Original Paper BACKGROUND: Atrial fibrillation (AF) is often paroxysmal and minimally symptomatic, hindering its diagnosis. Smartwatches may enhance AF care by facilitating long-term, noninvasive monitoring. OBJECTIVE: This study aimed to examine the accuracy and usability of arrhythmia discrimination using a smartwatch. METHODS: A total of 40 adults presenting to a cardiology clinic wore a smartwatch and Holter monitor and performed scripted movements to simulate activities of daily living (ADLs). Participants’ clinical and sociodemographic characteristics were abstracted from medical records. Participants completed a questionnaire assessing different domains of the device’s usability. Pulse recordings were analyzed blindly using a real-time realizable algorithm and compared with gold-standard Holter monitoring. RESULTS: The average age of participants was 71 (SD 8) years; most participants had AF risk factors and 23% (9/39) were in AF. About half of the participants owned smartphones, but none owned smartwatches. Participants wore the smartwatch for 42 (SD 14) min while generating motion noise to simulate ADLs. The algorithm determined 53 of the 314 30-second noise-free pulse segments as consistent with AF. Compared with the gold standard, the algorithm demonstrated excellent sensitivity (98.2%), specificity (98.1%), and accuracy (98.1%) for identifying irregular pulse. Two-thirds of participants considered the smartwatch highly usable. Younger age and prior cardioversion were associated with greater overall comfort and comfort with data privacy with using a smartwatch for rhythm monitoring, respectively. CONCLUSIONS: A real-time realizable algorithm analyzing smartwatch pulse recordings demonstrated high accuracy for identifying pulse irregularities among older participants. Despite advanced age, lack of smartwatch familiarity, and high burden of comorbidities, participants found the smartwatch to be highly acceptable. JMIR Publications 2019-05-15 /pmc/articles/PMC6834225/ /pubmed/31758787 http://dx.doi.org/10.2196/13850 Text en ©Eric Y Ding, Dong Han, Cody Whitcomb, Syed Khairul Bashar, Oluwaseun Adaramola, Apurv Soni, Jane Saczynski, Timothy P Fitzgibbons, Majaz Moonis, Steven A Lubitz, Darleen Lessard, Mellanie True Hills, Bruce Barton, Ki Chon, David D McManus. Originally published in JMIR Cardio (http://cardio.jmir.org), 15.05.2019. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Cardio, is properly cited. The complete bibliographic information, a link to the original publication on http://cardio.jmir.org, as well as this copyright and license information must be included. |
spellingShingle | Original Paper Ding, Eric Y Han, Dong Whitcomb, Cody Bashar, Syed Khairul Adaramola, Oluwaseun Soni, Apurv Saczynski, Jane Fitzgibbons, Timothy P Moonis, Majaz Lubitz, Steven A Lessard, Darleen Hills, Mellanie True Barton, Bruce Chon, Ki McManus, David D Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study |
title | Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study |
title_full | Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study |
title_fullStr | Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study |
title_full_unstemmed | Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study |
title_short | Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults: Observational Study |
title_sort | accuracy and usability of a novel algorithm for detection of irregular pulse using a smartwatch among older adults: observational study |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6834225/ https://www.ncbi.nlm.nih.gov/pubmed/31758787 http://dx.doi.org/10.2196/13850 |
work_keys_str_mv | AT dingericy accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT handong accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT whitcombcody accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT basharsyedkhairul accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT adaramolaoluwaseun accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT soniapurv accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT saczynskijane accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT fitzgibbonstimothyp accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT moonismajaz accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT lubitzstevena accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT lessarddarleen accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT hillsmellanietrue accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT bartonbruce accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT chonki accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy AT mcmanusdavidd accuracyandusabilityofanovelalgorithmfordetectionofirregularpulseusingasmartwatchamongolderadultsobservationalstudy |