Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting
BACKGROUND: Infrared (IR) analysis is an emerging technology that may be a useful tool for milk banks to manage the nutrient variability in donor human milk. OBJECTIVE: To evaluate the accuracy, reliability, and comparability of commercial infrared analyzers for measuring human milk macronutrients i...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6838652/ https://www.ncbi.nlm.nih.gov/pubmed/31723725 http://dx.doi.org/10.1093/cdn/nzz116 |
_version_ | 1783467260734603264 |
---|---|
author | Perrin, Maryanne T Festival, Jannette Starks, Shaina Mondeaux, Lesley Brownell, Elizabeth A Vickers, Amy |
author_facet | Perrin, Maryanne T Festival, Jannette Starks, Shaina Mondeaux, Lesley Brownell, Elizabeth A Vickers, Amy |
author_sort | Perrin, Maryanne T |
collection | PubMed |
description | BACKGROUND: Infrared (IR) analysis is an emerging technology that may be a useful tool for milk banks to manage the nutrient variability in donor human milk. OBJECTIVE: To evaluate the accuracy, reliability, and comparability of commercial infrared analyzers for measuring human milk macronutrients in a milk bank setting. METHODS: Three nonprofit milk banks received blinded test kits of human milk that had been assessed using reference methods. Four infrared instruments were used to measure macronutrients as follows: 1 filtered mid-IR, 2 Fourier-transformed full-spectra mid-IR, and 1 near-IR. Twenty-five unique samples were read concurrently for the accuracy arm. An identical sample was read daily for 1 mo for the reliability arm. RESULTS: Values for R(2) describing relationships with reference methods for total fat, crude protein, and lactose, were as follows: filtered mid-IR, 0.98, 0.94, and 0.48; Fourier-transformed full-spectra mid-IR, 0.97, 0.93, and 0.36 for instrument 1 and 0.98, 0.98, and 0.31 for instrument 2; and near-IR 0.93, 0.93, and 0.12. There was no significant difference between instruments for crude protein and total fat measurements. There were significant differences in carbohydrate measurements between instruments. For 1 mo of daily measurements in the reliability arm, CVs for filtered mid-IR were ≤4.6%, for Fourier-transformed full spectra mid-IR were ≤1.7%, and for near-IR were ≤5.1%. CONCLUSIONS: Infrared analysis is an accurate and reliable method for measuring crude protein and total fat in a milk bank setting. Carbohydrate measurements are less accurate and are significantly different between instruments, which will likely lead to differences in derived calorie values. |
format | Online Article Text |
id | pubmed-6838652 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-68386522019-11-13 Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting Perrin, Maryanne T Festival, Jannette Starks, Shaina Mondeaux, Lesley Brownell, Elizabeth A Vickers, Amy Curr Dev Nutr Original Research BACKGROUND: Infrared (IR) analysis is an emerging technology that may be a useful tool for milk banks to manage the nutrient variability in donor human milk. OBJECTIVE: To evaluate the accuracy, reliability, and comparability of commercial infrared analyzers for measuring human milk macronutrients in a milk bank setting. METHODS: Three nonprofit milk banks received blinded test kits of human milk that had been assessed using reference methods. Four infrared instruments were used to measure macronutrients as follows: 1 filtered mid-IR, 2 Fourier-transformed full-spectra mid-IR, and 1 near-IR. Twenty-five unique samples were read concurrently for the accuracy arm. An identical sample was read daily for 1 mo for the reliability arm. RESULTS: Values for R(2) describing relationships with reference methods for total fat, crude protein, and lactose, were as follows: filtered mid-IR, 0.98, 0.94, and 0.48; Fourier-transformed full-spectra mid-IR, 0.97, 0.93, and 0.36 for instrument 1 and 0.98, 0.98, and 0.31 for instrument 2; and near-IR 0.93, 0.93, and 0.12. There was no significant difference between instruments for crude protein and total fat measurements. There were significant differences in carbohydrate measurements between instruments. For 1 mo of daily measurements in the reliability arm, CVs for filtered mid-IR were ≤4.6%, for Fourier-transformed full spectra mid-IR were ≤1.7%, and for near-IR were ≤5.1%. CONCLUSIONS: Infrared analysis is an accurate and reliable method for measuring crude protein and total fat in a milk bank setting. Carbohydrate measurements are less accurate and are significantly different between instruments, which will likely lead to differences in derived calorie values. Oxford University Press 2019-10-21 /pmc/articles/PMC6838652/ /pubmed/31723725 http://dx.doi.org/10.1093/cdn/nzz116 Text en Copyright © The Author(s) 2019. http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | Original Research Perrin, Maryanne T Festival, Jannette Starks, Shaina Mondeaux, Lesley Brownell, Elizabeth A Vickers, Amy Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting |
title | Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting |
title_full | Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting |
title_fullStr | Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting |
title_full_unstemmed | Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting |
title_short | Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting |
title_sort | accuracy and reliability of infrared analyzers for measuring human milk macronutrients in a milk bank setting |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6838652/ https://www.ncbi.nlm.nih.gov/pubmed/31723725 http://dx.doi.org/10.1093/cdn/nzz116 |
work_keys_str_mv | AT perrinmaryannet accuracyandreliabilityofinfraredanalyzersformeasuringhumanmilkmacronutrientsinamilkbanksetting AT festivaljannette accuracyandreliabilityofinfraredanalyzersformeasuringhumanmilkmacronutrientsinamilkbanksetting AT starksshaina accuracyandreliabilityofinfraredanalyzersformeasuringhumanmilkmacronutrientsinamilkbanksetting AT mondeauxlesley accuracyandreliabilityofinfraredanalyzersformeasuringhumanmilkmacronutrientsinamilkbanksetting AT brownellelizabetha accuracyandreliabilityofinfraredanalyzersformeasuringhumanmilkmacronutrientsinamilkbanksetting AT vickersamy accuracyandreliabilityofinfraredanalyzersformeasuringhumanmilkmacronutrientsinamilkbanksetting |