Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing

[Image: see text] We created a glucose oxidase (GOx) working electrode on a silicon-on-insulator (SOI) wafer for glucose sensing. The SOI wafer was electrically connected to a copper wire, and the GOx was immobilized onto the hydrophilized SOI surface via silanization with aminopropyltriethoxysilane...

Descripción completa

Detalles Bibliográficos
Autores principales: Soto, Jacqueline, Hughes, Tyler, Li, Yize Stephanie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2019
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6844104/
https://www.ncbi.nlm.nih.gov/pubmed/31720532
http://dx.doi.org/10.1021/acsomega.9b02384
_version_ 1783468370304172032
author Soto, Jacqueline
Hughes, Tyler
Li, Yize Stephanie
author_facet Soto, Jacqueline
Hughes, Tyler
Li, Yize Stephanie
author_sort Soto, Jacqueline
collection PubMed
description [Image: see text] We created a glucose oxidase (GOx) working electrode on a silicon-on-insulator (SOI) wafer for glucose sensing. The SOI wafer was electrically connected to a copper wire, and the GOx was immobilized onto the hydrophilized SOI surface via silanization with aminopropyltriethoxysilane and glutaraldehyde. Electrochemical analysis (i.e., cyclic voltammetry) was employed to identify the sensing mechanism and to evaluate the performance of these SOI–GOx glucose sensors. The response of the SOI–GOx working electrode was significantly higher in the presence of oxygen than that without oxygen, indicating that a hydrogen peroxide pathway dominated in our SOI–GOx electrode. The height of cathodic peaks increased linearly with the increase of glucose concentrations up to 15 mM. The SOI–GOx working electrode displayed good stability after more than 30 cycles. On the 133(rd) day after the electrode was made, although the response of the SOI–GOx electrode dropped to about one-half of its original response, it was still capable of distinguishing different glucose concentrations. This work suggests that the SOI–GOx working electrode that we developed might be a promising candidate for implantable glucose sensors.
format Online
Article
Text
id pubmed-6844104
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher American Chemical Society
record_format MEDLINE/PubMed
spelling pubmed-68441042019-11-12 Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing Soto, Jacqueline Hughes, Tyler Li, Yize Stephanie ACS Omega [Image: see text] We created a glucose oxidase (GOx) working electrode on a silicon-on-insulator (SOI) wafer for glucose sensing. The SOI wafer was electrically connected to a copper wire, and the GOx was immobilized onto the hydrophilized SOI surface via silanization with aminopropyltriethoxysilane and glutaraldehyde. Electrochemical analysis (i.e., cyclic voltammetry) was employed to identify the sensing mechanism and to evaluate the performance of these SOI–GOx glucose sensors. The response of the SOI–GOx working electrode was significantly higher in the presence of oxygen than that without oxygen, indicating that a hydrogen peroxide pathway dominated in our SOI–GOx electrode. The height of cathodic peaks increased linearly with the increase of glucose concentrations up to 15 mM. The SOI–GOx working electrode displayed good stability after more than 30 cycles. On the 133(rd) day after the electrode was made, although the response of the SOI–GOx electrode dropped to about one-half of its original response, it was still capable of distinguishing different glucose concentrations. This work suggests that the SOI–GOx working electrode that we developed might be a promising candidate for implantable glucose sensors. American Chemical Society 2019-10-21 /pmc/articles/PMC6844104/ /pubmed/31720532 http://dx.doi.org/10.1021/acsomega.9b02384 Text en Copyright © 2019 American Chemical Society This is an open access article published under an ACS AuthorChoice License (http://pubs.acs.org/page/policy/authorchoice_termsofuse.html) , which permits copying and redistribution of the article or any adaptations for non-commercial purposes.
spellingShingle Soto, Jacqueline
Hughes, Tyler
Li, Yize Stephanie
Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing
title Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing
title_full Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing
title_fullStr Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing
title_full_unstemmed Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing
title_short Silicon-Based Glucose Oxidase Working Electrode for Glucose Sensing
title_sort silicon-based glucose oxidase working electrode for glucose sensing
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6844104/
https://www.ncbi.nlm.nih.gov/pubmed/31720532
http://dx.doi.org/10.1021/acsomega.9b02384
work_keys_str_mv AT sotojacqueline siliconbasedglucoseoxidaseworkingelectrodeforglucosesensing
AT hughestyler siliconbasedglucoseoxidaseworkingelectrodeforglucosesensing
AT liyizestephanie siliconbasedglucoseoxidaseworkingelectrodeforglucosesensing