Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer

The vitamin E-bonded polysulfone membrane hemodialyzer (ViE™-21) was evaluated in a clinical study for regulatory submission. Seventeen patients on hemodialysis were treated with conventional high-flux hemodialyzers for 2 weeks (Pre-ViE phase) and switched to the ViE-21 for 36 sessions (ViE phase) f...

Descripción completa

Detalles Bibliográficos
Autores principales: Kiaii, Mercedeh, Aritomi, Masaharu, Nagase, Mitsuyo, Farah, Myriam, Jung, Beverly
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Japan 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6861196/
https://www.ncbi.nlm.nih.gov/pubmed/31254226
http://dx.doi.org/10.1007/s10047-019-01110-w
_version_ 1783471299787489280
author Kiaii, Mercedeh
Aritomi, Masaharu
Nagase, Mitsuyo
Farah, Myriam
Jung, Beverly
author_facet Kiaii, Mercedeh
Aritomi, Masaharu
Nagase, Mitsuyo
Farah, Myriam
Jung, Beverly
author_sort Kiaii, Mercedeh
collection PubMed
description The vitamin E-bonded polysulfone membrane hemodialyzer (ViE™-21) was evaluated in a clinical study for regulatory submission. Seventeen patients on hemodialysis were treated with conventional high-flux hemodialyzers for 2 weeks (Pre-ViE phase) and switched to the ViE-21 for 36 sessions (ViE phase) followed by an additional 2 weeks on conventional hemodialyzers (Post-ViE phase). Reduction ratios of urea, creatinine, beta-2-microglobulin, albumin, and ultrafiltration coefficients (KUF) were measured once during the Pre-ViE phase and twice during the ViE phase. Moreover, biocompatibility markers [leucocyte count, platelet count, and activated complement factor (C3a) levels] were evaluated pre-dialysis, 15 min after initiation, and post-dialysis. During the study, type and number of adverse events (AEs), and device malfunctions were recorded. ViE-21 reduction ratios and KUF were not noticeably different than those of conventional hemodialyzers. Fluctuations of leucocyte counts and C3a concentrations were similar using ViE-21 and conventional hemodialyzers; however, the platelet count fluctuation was lower in ViE-21 sessions. The frequency of episodes of hypotension occurring during the ViE phase was lower than that occurring during the Pre- and Post-ViE phases. In conclusion, this study provided performance and safety data of the ViE-21 for regulatory application. The data suggest that vitamin E-bonded hemodialyzers are beneficial in lowering platelet activation and frequency of intradialytic hypotension. Larger randomized controlled trials are needed to confirm these findings.
format Online
Article
Text
id pubmed-6861196
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Springer Japan
record_format MEDLINE/PubMed
spelling pubmed-68611962019-12-03 Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer Kiaii, Mercedeh Aritomi, Masaharu Nagase, Mitsuyo Farah, Myriam Jung, Beverly J Artif Organs Original Article The vitamin E-bonded polysulfone membrane hemodialyzer (ViE™-21) was evaluated in a clinical study for regulatory submission. Seventeen patients on hemodialysis were treated with conventional high-flux hemodialyzers for 2 weeks (Pre-ViE phase) and switched to the ViE-21 for 36 sessions (ViE phase) followed by an additional 2 weeks on conventional hemodialyzers (Post-ViE phase). Reduction ratios of urea, creatinine, beta-2-microglobulin, albumin, and ultrafiltration coefficients (KUF) were measured once during the Pre-ViE phase and twice during the ViE phase. Moreover, biocompatibility markers [leucocyte count, platelet count, and activated complement factor (C3a) levels] were evaluated pre-dialysis, 15 min after initiation, and post-dialysis. During the study, type and number of adverse events (AEs), and device malfunctions were recorded. ViE-21 reduction ratios and KUF were not noticeably different than those of conventional hemodialyzers. Fluctuations of leucocyte counts and C3a concentrations were similar using ViE-21 and conventional hemodialyzers; however, the platelet count fluctuation was lower in ViE-21 sessions. The frequency of episodes of hypotension occurring during the ViE phase was lower than that occurring during the Pre- and Post-ViE phases. In conclusion, this study provided performance and safety data of the ViE-21 for regulatory application. The data suggest that vitamin E-bonded hemodialyzers are beneficial in lowering platelet activation and frequency of intradialytic hypotension. Larger randomized controlled trials are needed to confirm these findings. Springer Japan 2019-06-26 2019 /pmc/articles/PMC6861196/ /pubmed/31254226 http://dx.doi.org/10.1007/s10047-019-01110-w Text en © The Author(s) 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Article
Kiaii, Mercedeh
Aritomi, Masaharu
Nagase, Mitsuyo
Farah, Myriam
Jung, Beverly
Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
title Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
title_full Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
title_fullStr Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
title_full_unstemmed Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
title_short Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer
title_sort clinical evaluation of performance, biocompatibility, and safety of vitamin e-bonded polysulfone membrane hemodialyzer compared to non-vitamin e-bonded hemodialyzer
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6861196/
https://www.ncbi.nlm.nih.gov/pubmed/31254226
http://dx.doi.org/10.1007/s10047-019-01110-w
work_keys_str_mv AT kiaiimercedeh clinicalevaluationofperformancebiocompatibilityandsafetyofvitaminebondedpolysulfonemembranehemodialyzercomparedtononvitaminebondedhemodialyzer
AT aritomimasaharu clinicalevaluationofperformancebiocompatibilityandsafetyofvitaminebondedpolysulfonemembranehemodialyzercomparedtononvitaminebondedhemodialyzer
AT nagasemitsuyo clinicalevaluationofperformancebiocompatibilityandsafetyofvitaminebondedpolysulfonemembranehemodialyzercomparedtononvitaminebondedhemodialyzer
AT farahmyriam clinicalevaluationofperformancebiocompatibilityandsafetyofvitaminebondedpolysulfonemembranehemodialyzercomparedtononvitaminebondedhemodialyzer
AT jungbeverly clinicalevaluationofperformancebiocompatibilityandsafetyofvitaminebondedpolysulfonemembranehemodialyzercomparedtononvitaminebondedhemodialyzer