Cargando…

Feasibility of a circulation model for the assessment of endovascular recanalization procedures and periprocedural thromboembolism in-vitro

Aim of this study was to establish a simple and highly reproducible physiological circulation model to investigate endovascular device performance. The developed circulation model included a pneumatically driven pulsatile pump to generate a flow rate of 2.7 L/min at 70 beats per minute. Sections fro...

Descripción completa

Detalles Bibliográficos
Autores principales: Rusch, René, Trentmann, Jens, Hummitzsch, Lars, Rusch, Melanie, Aludin, Schekeb, Haneya, Assad, Albrecht, Martin, Schäfer, Jost Philipp, Puehler, Thomas, Cremer, Jochen, Berndt, Rouven
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6874641/
https://www.ncbi.nlm.nih.gov/pubmed/31757980
http://dx.doi.org/10.1038/s41598-019-53607-2
Descripción
Sumario:Aim of this study was to establish a simple and highly reproducible physiological circulation model to investigate endovascular device performance. The developed circulation model included a pneumatically driven pulsatile pump to generate a flow rate of 2.7 L/min at 70 beats per minute. Sections from the superficial femoral arteries were used in order to simulate device/tissue interaction and a filter was integrated to analyze periinterventional thromboembolism of white, red and mixed thrombi. The working fluid (3 L) was a crystalloid solution constantly tempered at 36.5 °C. To evaluate the model, aspiration thrombectomy, stent-implantation and thrombectomy with the Fogarty catheter were performed. Usability of the model was measured by the System Usability Scale (SUS) – Score. Histological specimens were prepared and analyzed postinterventional to quantify tissue/device interaction. Moreover, micro- and macroembolism were evaluated for each thrombus entity and each device. Results were tested for normality using the D’Agostino-Pearson test. Statistical comparisons of two groups were performed using the Student’s t-test. All devices were able to remove the occlusions after a maximum of 2 attempts. First-pass-recanalization was not fully achieved for aspiration thrombectomy of mixed thrombi (90.6%), aspiration thrombectomy of red thrombi (84.4%) and stent-implantation in occlusions of red thrombi (92.2%). Most micro- and macroembolism were observed using the Fogarty catheter and after stent-implantation in occlusions of white thrombi. Histological examinations revealed a significant reduction of the vascular layers suggesting vascular damage after use of the Fogarty catheter (327.3 ± 3.5 μm vs. 440.6 ± 3.9 μm; p = 0.026). Analysis of SUS rendered a mean SUS-Score of 80.4 which corresponds to an excellent user acceptability of the model. In conclusion, we describe a stable, easy to handle and reproducible physiological circulation model for the simulation of endovascular thrombectomy including device performance and thromboembolism.