Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease

BACKGROUND: Ongoing efforts within the Alzheimer’s disease (AD) field have focused on improving the intra- and inter-laboratory variability for cerebrospinal fluid (CSF) biomarkers. Fully automated assays offer the possibility to eliminate sample manipulation steps and are expected to contribute to...

Descripción completa

Detalles Bibliográficos
Autores principales: Leitão, Maria João, Silva-Spínola, Anuschka, Santana, Isabel, Olmedo, Veronica, Nadal, Alicia, Le Bastard, Nathalie, Baldeiras, Inês
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6875031/
https://www.ncbi.nlm.nih.gov/pubmed/31759396
http://dx.doi.org/10.1186/s13195-019-0550-8
_version_ 1783472939205656576
author Leitão, Maria João
Silva-Spínola, Anuschka
Santana, Isabel
Olmedo, Veronica
Nadal, Alicia
Le Bastard, Nathalie
Baldeiras, Inês
author_facet Leitão, Maria João
Silva-Spínola, Anuschka
Santana, Isabel
Olmedo, Veronica
Nadal, Alicia
Le Bastard, Nathalie
Baldeiras, Inês
author_sort Leitão, Maria João
collection PubMed
description BACKGROUND: Ongoing efforts within the Alzheimer’s disease (AD) field have focused on improving the intra- and inter-laboratory variability for cerebrospinal fluid (CSF) biomarkers. Fully automated assays offer the possibility to eliminate sample manipulation steps and are expected to contribute to this improvement. Recently, fully automated chemiluminescence enzyme immunoassays for the quantification of all four AD biomarkers in CSF became available. The aims of this study were to (i) evaluate the analytical performance of the Lumipulse G β-Amyloid 1-42 (restandardized to Certified Reference Materials), β-Amyloid 1-40, total Tau, and pTau 181 assays on the fully automated LUMIPULSE G600II; (ii) compare CSF biomarker results of the Lumipulse G assays with the established manual ELISA assays (INNOTEST®) from the same company (Fujirebio); and (iii) establish cut-off values and the clinical performance of the Lumipulse G assays for AD diagnosis. METHODS: Intra- and inter-assay variation was assessed in CSF samples with low, medium, and high concentrations of each parameter. Method comparison and clinical evaluation were performed on 40 neurological controls (NC) and 80 patients with a diagnosis of probable AD supported by a follow-up ≥ 3 years and/or positive amyloid PET imaging. A small validation cohort of 10 NC and 20 AD patients was also included to validate the cut-off values obtained on the training cohort. RESULTS: The maximal observed intra-assay and inter-assay coefficients of variation (CVs) were 3.25% and 5.50%, respectively. Method comparisons revealed correlation coefficients ranging from 0.89 (for Aβ40) to 0.98 (for t-Tau), with those for Aβ42 (0.93) and p-Tau (0.94) in-between. ROC curve analysis showed area under the curve values consistently above 0.85 for individual biomarkers other than Aβ40, and with the Aβ42/40, Aβ42/t-Tau, and Aβ42/p-Tau ratios outperforming Aβ42. Validation of the cut-off values in the independent cohort showed a sensitivity ranging from 75 to 95% and a specificity of 100%. The overall percentage of agreement between Lumipulse and INNOTEST was very high (> 87.5%). CONCLUSIONS: The Lumipulse G assays show a very good analytical performance that makes them well-suited for CSF clinical routine measurements. The good clinical concordance between the Lumipulse G and INNOTEST assays facilitates the implementation of the new method in routine practice.
format Online
Article
Text
id pubmed-6875031
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-68750312019-11-29 Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease Leitão, Maria João Silva-Spínola, Anuschka Santana, Isabel Olmedo, Veronica Nadal, Alicia Le Bastard, Nathalie Baldeiras, Inês Alzheimers Res Ther Research BACKGROUND: Ongoing efforts within the Alzheimer’s disease (AD) field have focused on improving the intra- and inter-laboratory variability for cerebrospinal fluid (CSF) biomarkers. Fully automated assays offer the possibility to eliminate sample manipulation steps and are expected to contribute to this improvement. Recently, fully automated chemiluminescence enzyme immunoassays for the quantification of all four AD biomarkers in CSF became available. The aims of this study were to (i) evaluate the analytical performance of the Lumipulse G β-Amyloid 1-42 (restandardized to Certified Reference Materials), β-Amyloid 1-40, total Tau, and pTau 181 assays on the fully automated LUMIPULSE G600II; (ii) compare CSF biomarker results of the Lumipulse G assays with the established manual ELISA assays (INNOTEST®) from the same company (Fujirebio); and (iii) establish cut-off values and the clinical performance of the Lumipulse G assays for AD diagnosis. METHODS: Intra- and inter-assay variation was assessed in CSF samples with low, medium, and high concentrations of each parameter. Method comparison and clinical evaluation were performed on 40 neurological controls (NC) and 80 patients with a diagnosis of probable AD supported by a follow-up ≥ 3 years and/or positive amyloid PET imaging. A small validation cohort of 10 NC and 20 AD patients was also included to validate the cut-off values obtained on the training cohort. RESULTS: The maximal observed intra-assay and inter-assay coefficients of variation (CVs) were 3.25% and 5.50%, respectively. Method comparisons revealed correlation coefficients ranging from 0.89 (for Aβ40) to 0.98 (for t-Tau), with those for Aβ42 (0.93) and p-Tau (0.94) in-between. ROC curve analysis showed area under the curve values consistently above 0.85 for individual biomarkers other than Aβ40, and with the Aβ42/40, Aβ42/t-Tau, and Aβ42/p-Tau ratios outperforming Aβ42. Validation of the cut-off values in the independent cohort showed a sensitivity ranging from 75 to 95% and a specificity of 100%. The overall percentage of agreement between Lumipulse and INNOTEST was very high (> 87.5%). CONCLUSIONS: The Lumipulse G assays show a very good analytical performance that makes them well-suited for CSF clinical routine measurements. The good clinical concordance between the Lumipulse G and INNOTEST assays facilitates the implementation of the new method in routine practice. BioMed Central 2019-11-23 /pmc/articles/PMC6875031/ /pubmed/31759396 http://dx.doi.org/10.1186/s13195-019-0550-8 Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Leitão, Maria João
Silva-Spínola, Anuschka
Santana, Isabel
Olmedo, Veronica
Nadal, Alicia
Le Bastard, Nathalie
Baldeiras, Inês
Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease
title Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease
title_full Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease
title_fullStr Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease
title_full_unstemmed Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease
title_short Clinical validation of the Lumipulse G cerebrospinal fluid assays for routine diagnosis of Alzheimer’s disease
title_sort clinical validation of the lumipulse g cerebrospinal fluid assays for routine diagnosis of alzheimer’s disease
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6875031/
https://www.ncbi.nlm.nih.gov/pubmed/31759396
http://dx.doi.org/10.1186/s13195-019-0550-8
work_keys_str_mv AT leitaomariajoao clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease
AT silvaspinolaanuschka clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease
AT santanaisabel clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease
AT olmedoveronica clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease
AT nadalalicia clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease
AT lebastardnathalie clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease
AT baldeirasines clinicalvalidationofthelumipulsegcerebrospinalfluidassaysforroutinediagnosisofalzheimersdisease