Development and Validation of a New High-Performance Liquid Chromatography Method for the Simultaneous Quantification of Coenzyme Q10, Phosphatidylserine, and Vitamin C from a Cutting-Edge Liposomal Vehiculization

[Image: see text] A high-performance liquid chromatography (HPLC) method was developed to simultaneously quantify three widely used active substances such as coenzyme Q10, phosphatidylserine, and vitamin C. This new method optimizes current timing and costs in the analyses of these three active subs...

Descripción completa

Detalles Bibliográficos
Autores principales: Ruiz-Garcia, Marta, Pérez-Lozano, Pilar, Mercadé-Frutós, Débora, Nardi-Ricart, Anna, Suñé-Pou, Marc, Cano-Sarabia, Mary, Garcia-Jimeno, Sonia, Suñé-Negre, Josep M., Maspoch, Daniel, García-Montoya, Encarna
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2019
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6881841/
https://www.ncbi.nlm.nih.gov/pubmed/31788602
http://dx.doi.org/10.1021/acsomega.9b02456
Descripción
Sumario:[Image: see text] A high-performance liquid chromatography (HPLC) method was developed to simultaneously quantify three widely used active substances such as coenzyme Q10, phosphatidylserine, and vitamin C. This new method optimizes current timing and costs in the analyses of these three active substances. Additionally, since the analyzed compounds were encapsulated on a cutting-edge liposomal formulation, further processing was necessary to be developed prior to HPLC analyses. The technique was studied and adequately validated in accordance with the guidelines of the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) regarding selectivity, linearity, accuracy, precision, and robustness. After data treatment of results, linear regressions for all active substances showed an optimal linearity with a correlation coefficient of >0.999 in the concentration range between 70 to 130% of the liposomal formulation and less than a 3% relative standard deviation (RSD) in accuracy and precision.