A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
We have developed an automated device for the measurement of oxygen consumption rate (OCR) called Chip-sensing Embryo Respiratory Measurement system (CERMs). To verify the safety and the significance of the OCR measurement by CERMs, we conducted comprehensive tests using a mouse model prior to clini...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6888687/ https://www.ncbi.nlm.nih.gov/pubmed/31726651 http://dx.doi.org/10.3390/ijms20225650 |
_version_ | 1783475287918379008 |
---|---|
author | Kuno, Takashi Tachibana, Masahito Fujimine-Sato, Ayako Fue, Misaki Higashi, Keiko Takahashi, Aiko Kurosawa, Hiroki Nishio, Keisuke Shiga, Naomi Watanabe, Zen Yaegashi, Nobuo |
author_facet | Kuno, Takashi Tachibana, Masahito Fujimine-Sato, Ayako Fue, Misaki Higashi, Keiko Takahashi, Aiko Kurosawa, Hiroki Nishio, Keisuke Shiga, Naomi Watanabe, Zen Yaegashi, Nobuo |
author_sort | Kuno, Takashi |
collection | PubMed |
description | We have developed an automated device for the measurement of oxygen consumption rate (OCR) called Chip-sensing Embryo Respiratory Measurement system (CERMs). To verify the safety and the significance of the OCR measurement by CERMs, we conducted comprehensive tests using a mouse model prior to clinical trials in a human in vitro fertilization (IVF) program. Embryo transfer revealed that the OCR measured by CERMs did not compromise the full-term development of mice or their future fertility, and was positively correlated with adenosine triphosphate (ATP) production and the mitochondrial membrane potential (ΔΨm), thereby indirectly reflecting mitochondrial oxidative phosphorylation (OXPHOS) activity. We demonstrated that the OCR is independent of embryo morphology (the size) and number of mitochondria (mitochondrial DNA copy number). The OCR correlated with the total cell numbers, whereas the inner cell mass (ICM) cell numbers and the fetal developmental rate were not. Thus, the OCR may serve as an indicator of the numbers of trophectoderm (TE) cells, rather than number or quality of ICM cells. However, implantation ability was neither correlated with the OCR, nor the embryo size in this model. This can probably be attributed to the limitation that chimeric embryos contain non-physiological high TE cells counts that are beneficial for implantation. CERMs can be safely employed in clinical IVF owing to it being a safe, highly effective, non-invasive, accurate, and quantitative tool for OCR measurement. Utilization of CERMs for clinical testing of human embryos would provide further insights into the nature of oxidative metabolism and embryonic viability. |
format | Online Article Text |
id | pubmed-6888687 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-68886872019-12-09 A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model Kuno, Takashi Tachibana, Masahito Fujimine-Sato, Ayako Fue, Misaki Higashi, Keiko Takahashi, Aiko Kurosawa, Hiroki Nishio, Keisuke Shiga, Naomi Watanabe, Zen Yaegashi, Nobuo Int J Mol Sci Article We have developed an automated device for the measurement of oxygen consumption rate (OCR) called Chip-sensing Embryo Respiratory Measurement system (CERMs). To verify the safety and the significance of the OCR measurement by CERMs, we conducted comprehensive tests using a mouse model prior to clinical trials in a human in vitro fertilization (IVF) program. Embryo transfer revealed that the OCR measured by CERMs did not compromise the full-term development of mice or their future fertility, and was positively correlated with adenosine triphosphate (ATP) production and the mitochondrial membrane potential (ΔΨm), thereby indirectly reflecting mitochondrial oxidative phosphorylation (OXPHOS) activity. We demonstrated that the OCR is independent of embryo morphology (the size) and number of mitochondria (mitochondrial DNA copy number). The OCR correlated with the total cell numbers, whereas the inner cell mass (ICM) cell numbers and the fetal developmental rate were not. Thus, the OCR may serve as an indicator of the numbers of trophectoderm (TE) cells, rather than number or quality of ICM cells. However, implantation ability was neither correlated with the OCR, nor the embryo size in this model. This can probably be attributed to the limitation that chimeric embryos contain non-physiological high TE cells counts that are beneficial for implantation. CERMs can be safely employed in clinical IVF owing to it being a safe, highly effective, non-invasive, accurate, and quantitative tool for OCR measurement. Utilization of CERMs for clinical testing of human embryos would provide further insights into the nature of oxidative metabolism and embryonic viability. MDPI 2019-11-12 /pmc/articles/PMC6888687/ /pubmed/31726651 http://dx.doi.org/10.3390/ijms20225650 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Kuno, Takashi Tachibana, Masahito Fujimine-Sato, Ayako Fue, Misaki Higashi, Keiko Takahashi, Aiko Kurosawa, Hiroki Nishio, Keisuke Shiga, Naomi Watanabe, Zen Yaegashi, Nobuo A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model |
title | A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model |
title_full | A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model |
title_fullStr | A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model |
title_full_unstemmed | A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model |
title_short | A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model |
title_sort | preclinical evaluation towards the clinical application of oxygen consumption measurement by cerms by a mouse chimera model |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6888687/ https://www.ncbi.nlm.nih.gov/pubmed/31726651 http://dx.doi.org/10.3390/ijms20225650 |
work_keys_str_mv | AT kunotakashi apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT tachibanamasahito apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT fujiminesatoayako apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT fuemisaki apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT higashikeiko apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT takahashiaiko apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT kurosawahiroki apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT nishiokeisuke apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT shiganaomi apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT watanabezen apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT yaegashinobuo apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT kunotakashi preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT tachibanamasahito preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT fujiminesatoayako preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT fuemisaki preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT higashikeiko preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT takahashiaiko preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT kurosawahiroki preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT nishiokeisuke preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT shiganaomi preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT watanabezen preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel AT yaegashinobuo preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel |