A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model

We have developed an automated device for the measurement of oxygen consumption rate (OCR) called Chip-sensing Embryo Respiratory Measurement system (CERMs). To verify the safety and the significance of the OCR measurement by CERMs, we conducted comprehensive tests using a mouse model prior to clini...

Descripción completa

Detalles Bibliográficos
Autores principales: Kuno, Takashi, Tachibana, Masahito, Fujimine-Sato, Ayako, Fue, Misaki, Higashi, Keiko, Takahashi, Aiko, Kurosawa, Hiroki, Nishio, Keisuke, Shiga, Naomi, Watanabe, Zen, Yaegashi, Nobuo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6888687/
https://www.ncbi.nlm.nih.gov/pubmed/31726651
http://dx.doi.org/10.3390/ijms20225650
_version_ 1783475287918379008
author Kuno, Takashi
Tachibana, Masahito
Fujimine-Sato, Ayako
Fue, Misaki
Higashi, Keiko
Takahashi, Aiko
Kurosawa, Hiroki
Nishio, Keisuke
Shiga, Naomi
Watanabe, Zen
Yaegashi, Nobuo
author_facet Kuno, Takashi
Tachibana, Masahito
Fujimine-Sato, Ayako
Fue, Misaki
Higashi, Keiko
Takahashi, Aiko
Kurosawa, Hiroki
Nishio, Keisuke
Shiga, Naomi
Watanabe, Zen
Yaegashi, Nobuo
author_sort Kuno, Takashi
collection PubMed
description We have developed an automated device for the measurement of oxygen consumption rate (OCR) called Chip-sensing Embryo Respiratory Measurement system (CERMs). To verify the safety and the significance of the OCR measurement by CERMs, we conducted comprehensive tests using a mouse model prior to clinical trials in a human in vitro fertilization (IVF) program. Embryo transfer revealed that the OCR measured by CERMs did not compromise the full-term development of mice or their future fertility, and was positively correlated with adenosine triphosphate (ATP) production and the mitochondrial membrane potential (ΔΨm), thereby indirectly reflecting mitochondrial oxidative phosphorylation (OXPHOS) activity. We demonstrated that the OCR is independent of embryo morphology (the size) and number of mitochondria (mitochondrial DNA copy number). The OCR correlated with the total cell numbers, whereas the inner cell mass (ICM) cell numbers and the fetal developmental rate were not. Thus, the OCR may serve as an indicator of the numbers of trophectoderm (TE) cells, rather than number or quality of ICM cells. However, implantation ability was neither correlated with the OCR, nor the embryo size in this model. This can probably be attributed to the limitation that chimeric embryos contain non-physiological high TE cells counts that are beneficial for implantation. CERMs can be safely employed in clinical IVF owing to it being a safe, highly effective, non-invasive, accurate, and quantitative tool for OCR measurement. Utilization of CERMs for clinical testing of human embryos would provide further insights into the nature of oxidative metabolism and embryonic viability.
format Online
Article
Text
id pubmed-6888687
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-68886872019-12-09 A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model Kuno, Takashi Tachibana, Masahito Fujimine-Sato, Ayako Fue, Misaki Higashi, Keiko Takahashi, Aiko Kurosawa, Hiroki Nishio, Keisuke Shiga, Naomi Watanabe, Zen Yaegashi, Nobuo Int J Mol Sci Article We have developed an automated device for the measurement of oxygen consumption rate (OCR) called Chip-sensing Embryo Respiratory Measurement system (CERMs). To verify the safety and the significance of the OCR measurement by CERMs, we conducted comprehensive tests using a mouse model prior to clinical trials in a human in vitro fertilization (IVF) program. Embryo transfer revealed that the OCR measured by CERMs did not compromise the full-term development of mice or their future fertility, and was positively correlated with adenosine triphosphate (ATP) production and the mitochondrial membrane potential (ΔΨm), thereby indirectly reflecting mitochondrial oxidative phosphorylation (OXPHOS) activity. We demonstrated that the OCR is independent of embryo morphology (the size) and number of mitochondria (mitochondrial DNA copy number). The OCR correlated with the total cell numbers, whereas the inner cell mass (ICM) cell numbers and the fetal developmental rate were not. Thus, the OCR may serve as an indicator of the numbers of trophectoderm (TE) cells, rather than number or quality of ICM cells. However, implantation ability was neither correlated with the OCR, nor the embryo size in this model. This can probably be attributed to the limitation that chimeric embryos contain non-physiological high TE cells counts that are beneficial for implantation. CERMs can be safely employed in clinical IVF owing to it being a safe, highly effective, non-invasive, accurate, and quantitative tool for OCR measurement. Utilization of CERMs for clinical testing of human embryos would provide further insights into the nature of oxidative metabolism and embryonic viability. MDPI 2019-11-12 /pmc/articles/PMC6888687/ /pubmed/31726651 http://dx.doi.org/10.3390/ijms20225650 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Kuno, Takashi
Tachibana, Masahito
Fujimine-Sato, Ayako
Fue, Misaki
Higashi, Keiko
Takahashi, Aiko
Kurosawa, Hiroki
Nishio, Keisuke
Shiga, Naomi
Watanabe, Zen
Yaegashi, Nobuo
A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
title A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
title_full A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
title_fullStr A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
title_full_unstemmed A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
title_short A Preclinical Evaluation towards the Clinical Application of Oxygen Consumption Measurement by CERMs by a Mouse Chimera Model
title_sort preclinical evaluation towards the clinical application of oxygen consumption measurement by cerms by a mouse chimera model
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6888687/
https://www.ncbi.nlm.nih.gov/pubmed/31726651
http://dx.doi.org/10.3390/ijms20225650
work_keys_str_mv AT kunotakashi apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT tachibanamasahito apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT fujiminesatoayako apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT fuemisaki apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT higashikeiko apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT takahashiaiko apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT kurosawahiroki apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT nishiokeisuke apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT shiganaomi apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT watanabezen apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT yaegashinobuo apreclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT kunotakashi preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT tachibanamasahito preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT fujiminesatoayako preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT fuemisaki preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT higashikeiko preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT takahashiaiko preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT kurosawahiroki preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT nishiokeisuke preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT shiganaomi preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT watanabezen preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel
AT yaegashinobuo preclinicalevaluationtowardstheclinicalapplicationofoxygenconsumptionmeasurementbycermsbyamousechimeramodel